    A-dec 200 i nstructions for u se dental chair, delivery and support systems, cuspidor, dental light.

  • Page 2

    Copyright © 2012 a-dec inc. All rights reserved. A-dec inc. Makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. A-dec inc. Shall not be held liable for any errors contained herein o...

  • Page 3: Content Map

    86.0024.00 rev c 1 content map 23 6 15 17 20 6 8 specifications/ warranty 28, 29 quick start 2 15.

  • Page 4: Quick Start

    2 a-dec 200 instructions for use quick start turn the system on. Check the water bottle and fill if near empty. Lift to activate the handpiece. Flip the foot control toggle to turn the water coolant on or off. Step on the disc to operate. Turn the knob counterclockwise to disengage the flexarm brake...

  • Page 5

    86.0024.00 rev c 3 use the manual or programmable buttons to position the chair. For complete programming instructions, see page 8. Use the touchpad to activate the dental light and cuspidor functions. For complete programming instructions, see pages 20 and 23. Manual buttons bowl rinse cupfill manu...

  • Page 6

    4 a-dec 200 instructions for use quick start flip the dental light toggle switch or press the dental light bu tton on the touchpad to turn the light on high or low intensity*. Control cupfill and bowl rinse functions from the cuspidor tower. Push in the syringe tip retainer lock to insert or remove ...

  • Page 7: A-Dec 200 Reference Section

    86.0024.00 rev c 5 a-dec 200 reference section chair and delivery system ........ 6 assistant’s instrumentation ......15 cuspidor ............................20 dental light ........................23 specifications ......................28 warranty ............................29.

  • Page 8

    6 a-dec 200 instructions for use dental chair and delivery system footswitch 8 headrest 9 doctor’s touchpad 8 instrument holder 11 delivery system handle 11 water/air coolant flow controls 10 master toggle 7 tubing flush 12 utilities 13 foot control 10 upholstery care 13 oil collector 10 flexarm 12 ...

  • Page 9

    86.0024.00 rev c 7 power and system status safety features the chair and system power is controlled by the master toggle on the delivery system. To save energy, turn the power off at the end of the work day and during longer periods of non-use. The power should always be turned off before service. W...

  • Page 10

    8 a-dec 200 instructions for use programmable chair positions you can operate the chair manually or by using programmed presets from the footswitch or optional touchpad. See page 3 for manual controls. The programmable buttons vary depending on which device you use to control the chair: chair preset...

  • Page 11

    86.0024.00 rev c 9 x-ray/rinse chair position the x-ray/rinse preset button moves the chair and patient into an upright position for x-rays or cuspidor access. If you’d like to change this function to an additional programmable chair position, contact your authorized a-dec dealer. X-ray/ rinse posit...

  • Page 12

    10 use the two adjustment knobs to customize water and air coolant flow to your handpieces. The blue knob adjusts water coolant flow to all handpieces. The other knob adjusts air coolant flow to all handpieces. 1. Lift the handpiece from the holder. 2. Flip the wet/dry toggle on the foot control to ...

  • Page 13

    86.0024.00 rev c 11 instrument holder adjustment delivery system handle adjustment the instrument holders rotate independently to allow customized positioning for each handpiece. Pull the holder slightly away from the adjacent one, rotate to the desired position, and then release. * if your assistan...

  • Page 14

    12 a-dec 200 instructions for use flexarm rotation if the control head flexarm begins to drift right or left, use a 3/32" hex key to adjust the tension setscrew. Turn the screw clockwise to tighten or counterclockwise to loosen the tension. Tension adjustments dental chair and delivery system tensio...

  • Page 15

    86.0024.00 rev c 13 touch and transfer surfaces touch surfaces are those areas that come into contact with hands and become potential cross-contamination points during dental procedures. Transfer surfaces are those surfaces that are contaminated by contact with instruments and other inanimate object...

  • Page 16

    14 a-dec 200 instructions for use dental chair and delivery system to check for a clogged air or water filter: 1. Turn on the master toggle. 2. While watching the air pressure gauge, press the syringe air button. If the system air pressure drops by more than 15 psi, replace the air filter. 3. While ...

  • Page 17

    86.0024.00 rev c 15 assistant’s instrumentation and support center optional assistant’s syringe 18 self-contained water bottle 16 saliva ejector 19 hve (high-volume evacuator) 19 solids collector 17 optional assistant’s touchpad 21 doctor’s syringe 18 support center a-dec icx waterline treatment tab...

  • Page 18

    16 a-dec 200 instructions for use the 2-liter self-contained water bottle supplies water to handpieces, syringes, and the cuspidor cupfill. To remove the bottle, turn off the master toggle, and rotate the bottle to the left. For proper installation: 1. Place an icx waterline treatment tablet in the ...

  • Page 19

    86.0024.00 rev c 17 waterline maintenance protocol a consistent waterline maintenance protocol is required to maintain good water quality in your system. A-dec’s recommendation for waterline maintenance is based on continuous use of the a-dec icx waterline treatment tablet. Icx is intended for use w...

  • Page 20

    18 a-dec 200 instructions for use assistant’s instrumentation and support center all syringe tips are considered critical items and must be heat sterilized or disposed of after each use. Precleaning always preclean the syringe tips before sterilization. 1. Remove the tip from the syringe. 2. Immerse...

  • Page 21

    86.0024.00 rev c 19 hve and saliva ejector asepsis if you use barrier film to protect the main body of the hve and saliva ejector, clean and sterilize these instruments weekly, or more frequently as needed. If you do not use barrier film, or if you perform oral surgery, clean and sterilize these ins...

  • Page 22: Cuspidor

    20 a-dec 200 instructions for use cuspidor support center bowl screen 22 bowl rinse spout cupfill button bowl rinse button the cuspidor cupfill and bowl rinse functions are dependent on your specific configuration. Standard cuspidor press and hold the cupfill button on the cuspidor for the desired a...

  • Page 23

    86.0024.00 rev c 21 bowl rinse programming (only systems with doctor’s touchpad) to reprogram the timed cupfill or bowl rinse functions: 1. Press on the touchpad or press and hold both the cupfill and bowl rinse buttons on the cuspidor. Release them when you hear one beep. 2. Press and hold the cupf...

  • Page 24

    22 a-dec 200 instructions for use cuspidor cleaning and maintenance cuspidor bowl the contoured spouts and smooth bowl of the cuspidor provide for quick and easy cleaning. Remember to empty and clean the bowl screen every time you clean the cuspidor. Gravity drain cleaning at the end of each day, fl...

  • Page 25: Dental Light

    86.0024.00 rev c 23 dental light pivot housing/ diagonal tension adjustment 25 light shield tab (hidden) 26 light shield 26 the a-dec 200 dental light can be operated from the manual 3-position switch or the optional touchpad. The dental light is always off when the manual switch is in the center lo...

  • Page 26

    24 a-dec 200 instructions for use auto on/off the optional auto on/off feature turns the light on when the chair back reaches a treatment position. Press or , and the dental light turns off. To disable the auto on/off feature, press and hold and simulta- neously for three seconds. One beep confirms ...

  • Page 27

    86.0024.00 rev c 25 rotation adjustments if the light head is difficult to position, moves too easily, or tends to slip out of position, you can adjust the diagonal and vertical rotation tension. Diagonal adjustment for diagonal rotation tension, adjust the setscrew on the pivot housing using a 5/64...

  • Page 28

    26 a-dec 200 instructions for use dental light bulb replacement follow these steps when replacing the dental light bulb: 1. Turn the light off and allow it to cool. 2. Hold the light shield and gently squeeze one side while pulling the shield away from the dental light. Set the light shield aside. 3...

  • Page 29

    86.0024.00 rev c 27 common dental light replacement parts contact your authorized a-dec dealer for available dental light maintenance parts. Bulb 041.709.00 light shield 97.0411.00 description part number dental light shield maintenance for optimal performance and maximum life of your dental light s...

  • Page 30: Specifications

    28 a-dec 200 instructions for use specifications dental light operating specifications focal range: 305 mm - 686 mm (12" - 27") bulb: quartz halogen xenon, single-end lamp rating: 17v/ 95w color temperature: 4800 kelvin (nominal; at high intensity) light pattern: 230 mm x 110 mm at 700 mm (9" x 4.3"...

  • Page 31: Warranty

    86.0024.00 rev c 29 a-dec warrants all products against defects in materials or workman- ship for one year from time of delivery. A-dec’s sole obligation under the warranty is to provide parts for the repair, or at its option, to provide the replacement product (excluding labor). The buyer shall hav...

  • Page 32

    Ifulsbk2 86.0024.00 rev c copyright 2012 a-dec inc. All rights reserved. A-dec australia unit 8 5-9 ricketty street mascot, nsw 2020 australia tel: 1.800.225.010 within australia tel: +61 (0)2 8332 4000 outside australia www.A-dec.Com.Au a-dec headquarters 2601 crestview drive newberg, or 97132 usa ...