JENCO 3020M Operation manual

Manual is about: Hand-held Conductivity/Salinity/TDS/ Temperature Meter

Summary of 3020M

  • Page 1

    Operation manual hand-held conductivity/salinity/tds/ temperature meter 3020m.

  • Page 2: Contents

    Contents general introduction ……………… ....................2 initial inspection …… ... ……………… ....................2 water proof …… . …… ... ……………… ....................2 installing the batteries.. ………… ....................3 display & keys functions … .. …………… ... ………… 4 a. Display ........................

  • Page 3

    General introduction thank you for selecting the 3020m meter. The 3020m is a precision tool that measure conductivity, salinity, tds and temperature. A built-in microprocessor stores, calculates and compensates for all parameters related to conductivity and temperature determinations. This unit has ...

  • Page 4

    May affect probe connections. 2. Wait for the unit and probe to dry completely before resuming operation. 3. If the unit does not function correctly after steps1 and 2, call jenco for possible repair or replacement (see warranty). Installing the batteries the 3020m meter is packaged with one 9v batt...

  • Page 5

    Display & keys functions a. Display figure 2: active lcd screen 1. Bat- 8. Delete- low battery indicator. 2. Full- this will indicate that all 50 data storage sites are used up. 3. Cell- to delete stored data. 9. Each- to delete a single set of data from the data storage. 10. ℃ /%- displays during c...

  • Page 6

    15. Temperature display and unit 18. Conductivity- displays when measuring conductivity. 19. Recall- 16. Data storage site number. To recall data from the data storage. 20. Uncompensated- 17. Save- distinguish between to save a reading into the data temperature compensated storage. And non- temperat...

  • Page 7

    Enter- in "calibration" mode, press this key to save the current parameter to memory. In "measurement" mode, press this key to save reading into the next available data storage site. At the recall interface, press this key to display the last set of saved data. At the delete interface, press this ke...

  • Page 8

    4. Salinity mode: the unit will display salinity reading. 5. Recall interface: press "enter" key to go into recall mode. 6. Recall mode: in this mode, user can recall data saved in memory. 7. Delete interface: press "enter" key to go into delete mode. 8. Delete mode: in this mode, user can erase eac...

  • Page 9

    Operational procedures a. Preparing standard solutions suitable conductivity standards are available commercially or the user can prepare them using research grade reagents. Here are some standard solutions the user can prepare to calibrate the probe of the model 3020m. 1. 2. 3. 4. 5. Standard solut...

  • Page 10

    2. Allow temperature reading to stabilize, press "cal" key to enter the calibration mode. The "cal" icon appears on the lcd. Press "mode" key to sequentially view previous calibration settings. Tds tds is determined by multiplying conductivity (ms) by a tds factor. The default factor value is 0.65. ...

  • Page 11

    Changes made will be cancelled and the previous calibration settings will be retained. Conductivity calibration (a) immerse the probe in a standard of known conductivity, preferably a standard in the middle range of the solutions to be measured. Immerse the probe (at least 2" to 3" or 5~7cm from the...

  • Page 12

    B. 1. 2. 3. C. 1. 2. 3. 4. 5. Recalling readings from memory. To recall saved data, press "enter" key at the recall interface to go into "recall" mode. Press the "up" or "down" keys to select the storage site number. Press "mode" key to exit "recall" mode. Deleting data. Press the "enter" key at the...

  • Page 13

    Error displays and troubleshooting main display secondary display possible cause(s) corrective action(s) ● conductivity is > 200.0ms (k=0.475). ● ensure the probe is immersed at least 2"~3" or 5~7cm from the tip, below the "over" during / ● conductivity surface of the measurements is > 200.0us (k=0....

  • Page 14

    Specifications display range resolution accuracy 0.0 to 475.0us/cm 0.1us/cm conductivity 475 to 4750us/cm 1us/cm ±0.5% full scale k=0.475 4.75 to 47.50ms/cm 0.01ms/cm 47.5 to 200.0ms/cm 0.1ms/cm conductivity 0.00 to 99.99us/cm 0.01us/cm ±0.5% full scale k=0.1 100.0 to 200.0us/cm 0.1us/cm salinity te...

  • Page 15

    Warranty jenco warrants this product to be free from significant deviations in material and workmanship for a period of 1 year from date of purchase. If repair or adjustment is necessary and has not been the result of abuse or misuse, within the year period, please return-freight-prepaid and the cor...

  • Page 16

    Jenco instruments, inc. 7968 arjons drive, suite c san diego, ca 92126 usa tel: 858-578-2828 fax: 858-578-2886 e-mail: jencoinfo@jencoi.Com; sales@jencoi.Com website: www.Jencoi.Com jenco electronics inc. 1f., no. 11, lane 370, sec. 6, zhongxiao e. Rd. Nangang dist., taipei, taiwan tel: 886-2-2782-3...

  • Page 17

    16.