Nagata FAT-06 User manual - page 5
NAGATA FAT-06 Calibration Instructions
Note:
The test weight used to calibrate the scale should be no less than 1/6 the max
capacity of the scale. Example: For a 20kg capacity scale, the minimum test weight
used to calibrate the scale must be 4kg
The test weight used to calibrate the scale must not be greater than the max capacity of the
scale. Example: For a 20kg capacity scale, the maximum test weight used should be 20kg or
less.
Keypad function chart
HI
NET
Number increasing key
Confirmed key
Movement key
ZERO
LO
GROSS
+
1. Leave the scale and assemble tray.
2. Make sure that the power is off.
3. Press and hold the HI/LO key then turn on the
power while still holding the HI/LO key. Hold
the HI/LO key until a 1 appears on the screen.
Release the HI/LO key.
4. Press the ZERO+ key and the screen will
display 2.
5. Press the HI/LO key and the screen will
display 3.
6. Press ZERO+ key and the screen will display
123.
7. After 20 seconds and the screen will appear the
calibration capacity 020.000.
8. Using the NET/GROSS and ZERO+ keys,
change the weight displayed to match the test
weight you will be using. The ZERO+ key
changes the value of the flashing digit. The
NET/GROSS key changes the position of the
flashing digit.