NCI 6702 User Manual - page 10
Configuration Mode- Continued
Follow these steps to access and configure the items in the ConF menu. Refer to
figure 3.
1. From the dIAG display press the ZERO key
OR
From normal weighing mode, move switch 1 to
Menu Mode or OPEN position, then press
the ZERO key. . .
ConF is displayed
2. Press the TEST key. . .
FiLt is displayed. This stands for
filtering.
3. Press the TEST key. . .
The current setting, FAST or SLO,
is displayed.
4. Use the ZERO key to toggle
between the two choices. Press
the TEST key when the choice
you want is displayed. . .
Your choice is accepted and the
display shows FILt.
5. Press the ZERO key once. . .
BAUD is displayed.
6. Press the TEST key. . .
The current baud and parity choice
is displayed.
7. Use the ZERO key to scroll
the choices found in Table 1.
When the choice you want is
displayed, press the TEST key. . .
Your choice is accepted and the
display FILt.
8. Press the ZERO key twice. . .
PROT is displayed. This stands for
protocol.
9. Press the TEST key. . .
The current RS-232 communication
protocol is displayed.
10. Press the ZERO key to scroll
through the choices. When the
choice you want is displayed,
press of the TEST key. . .
Your choice is accepted and the
display shows FILt.
11. When you are finished with
configuring your scale press
the ZERO key three times,
which displays donE, then
press the TEST key. . .
dIAG is displayed
Step-by-Step
Instructions for
ConF Mode
8