E.A.R.® Plug Care
The E.A.R.® foam earplugs can be washed with soap and water and reused or they can
simply be thrown away and replaced. (Remove the E.A.R.® plug from the blue ear mold
prior to washing.)
If the black tube separates from the yellow foam or if the plug does not slide
properly over the temperature sensor, discard the plug and try another.
Operating
1.
Insert the ear sensor assembly as described in Section IV and allow 5 minutes for
the temperature to stabilize. [It may be beneficial to run the sensor cable under
clothing or fasten it to keep it from being snagged.]
2.
Turn the QUESTEMP
II ON. RESET to clear old data if desired. Put the unit in the
RUN mode.
3.
Check the battery strength by pressing the BATTERY/TEST key. Replace the
ALKALINE battery if 'Lo' appears in the display.
4.
Calibrate the unit following the guidelines in CALIBRATION section, page 12.
5.
If your situation requires it, attach the front cover.
6.
Clip the QUESTEMP
II housing to your belt or pocket and wear it throughout the
work day or as needed.
7.
After use, remove the earplug, print out the data if desired, and turn the unit OFF.
Indicators and Action
ALARM - If the alarm activates, indicating that the body temperature has exceeded the
set alarm level, take appropriate measures to allow the body to cool itself and properly
regulate its temperature. Lowering the work load, moving into a cooler environment, and
drinking cool liquids are all methods to help cool the body and avoid heat related injuries.
LOW BATTery - If this indicator comes on, replace the 9-volt ALKALINE battery as soon
as possible.
To preserve battery life, be sure to turn the unit OFF when not in use.