  • Page 1

    3 ranger ™ pressure infusor model 145 operator’s manual 1 ranger pressure infusor model 145 operator's manual 2 dispositif de perfusion sous pression ranger modèle 145 manuel d’utilisation 3 ranger druckinfusor modell 145 bedienungsanleitung 4 infusore a pressione ranger modello 145 manuale dell’ope...

  • Page 3

    3 ranger ™ pressure infusor model 145 operator’s manual ةيبرعلا ةغللا english 3 français 21 deutsch 43 italiano 65 español 87 nederland 109 svenska 131 dansk 153 norsk 175 suomi 197 português 219 Ελληνικά 241 polski 263 magyar 285 Česky 307 lietuvių kalba 329 Русский 351 türkçe 373 日本語 395 英语 417 43...

  • Page 4: Revision History

    Revision history revision reason for change pages affected date a new revision - first release all may 2013 b translations added all july 2013.

  • Page 5: Table of Contents

    Table of contents section 1: technical service and order placement 3 technical service ...

  • Page 7: Usa

    Technical service and order placement 3 202041b technical service and order placement usa tel: +1-800-228-3957 proper use and maintenance 3m assumes no responsibility for the reliability, performance, or safety of the ranger pressure infusor system if any of the following events occur: • modificatio...

  • Page 8

    4 technical service and order placement.

  • Page 9: Indications For Use

    202041b introduction 5 indications for use the 3m™ ranger™ pressure infusor is intended to provide pressure to i.V. Solution bags when rapid infusion of liquids is required. Explanation of symbols the following symbols may appear on the product’s labeling or exterior packaging. Date of manufacture m...

  • Page 10

    6 introduction an equipotentiality plug (grounded) conductor other than a protective earth conductor or a neutral conductor, providing a direct connection between electrical equipment and the potential equalization busbar of the electrical installation. Please consult iec 60601-1; 2005 for requireme...

  • Page 11

    202041b introduction 7 cautions 1. To prevent tipping, do not mount this unit more than 56” (142 cm) from the floor to the base of the pressure infusor unit. 2. This product is designed for pressure infusion only. 3. Do not sterilize the pressure infusor. 4. To reduce the risks associated with envir...

  • Page 12: Product Description

    8 introduction product description the ranger pressure infusor can be used with the ranger blood/fluid warming unit, standard and high flow disposable sets, the ranger irrigation warming unit, and irrigation warming sets 1 . The infusor can accept solution bags ranging from 250 ml to 1000 ml. The pr...

  • Page 13: Pressure Infusor Panel

    202041b introduction 9 pressure infusor panel the pressure infusor panel displays the status of the pressure infusors. The pressure infusor power indicator illuminates amber (standby) when the main power switch is turned on and the pressure infusors are able to be turned on. A green led indicates th...

  • Page 14

    10 introduction high visual and audible alarm: the high amber indicator illuminates and an audible alarm notify the user when the pressure infusor bladder is above 330 mmhg. The visual indicator and alarm will continue as long as the pressure remains above 330 mmhg. If the high condition is observed...

  • Page 15

    Instructions for use 11 202041b attaching the pressure infusor to the i.V. Pole 1. Mount the model 145 pressure infusor onto a model 90068/90124 pressure infusor i.V. Pole/base as shown below. Caution: to prevent tipping, do not mount this unit more than 56 " (142 cm) from the floor to the base of t...

  • Page 16: Changing A Fluid Bag

    12 instructions for use load and pressurize the infusors 1. Plug the power cord into an appropriately grounded outlet. 2. Using the main power switch located under the pressure infusor, turn the unit on. 3. Remove excess air from i.V. Solution bags and prime. 4. Open pressure infusor door. 5. Slide ...

  • Page 17: Standby/on Mode

    Troubleshooting 13 202041b all repair, calibration, and servicing of the ranger pressure infusor requires the skill of a qualified, medical equipment service technician who is familiar with good practice for medical unit repair. All repairs and maintenance should be in accordance with manufacturer’s...

  • Page 18: Pressure Infusor

    14 troubleshooting pressure infusor condition cause solution pressure infusor is not working. Unit is not plugged into the power entry module, or power cord is not plugged into an appropriately grounded outlet. Make sure the power cord is plugged into the power entry module of the pressure infusor. ...

  • Page 19: Storage

    Maintenance and storage 15 202041b cleaning the ranger pressure infusor unit clean the ranger pressure infusor unit on an as-needed basis or per institutional protocol. To clean the outside of the unit 1. Disconnect the ranger pressure infusor unit from the power source. 2. Wipe the outside of the u...

  • Page 20

    16 maintenance and storage.

  • Page 21: Physical Characteristics

    Specifications 17 202041b physical characteristics dimensions 15.75 in (40 cm) high 20 in (51 cm) wide, 7.75 in (20 cm) deep weight 17 lb. (7.7 kg) mounting dual clamp classification classifed under iec 60601-1 guidelines as class 1, type bf, ordinary equipment. Classified by underwriters laboratori...

  • Page 22: Performance Characteristics

    18 specifications performance characteristics operating pressure 300 mmhg setpoint note: • pressure system is in range when the pressure infusor bladders are inflated to between 230 mmhg (low) and 330 mmhg (high). If pressure falls below 230 mmhg for more than approximately 30 seconds the low amber ...

  • Page 24

    Made in the usa of globally sourced material by 3m health care 3m is a trademark of 3m company, used under license in canada ranger and the bair hugger logo are trademarks of arizant healthcare inc , used under license in canada ©2013 arizant healthcare inc all rights reserved federal law (usa)...