A&D AD-4212C-300 Instruction Manual

Manual is about: Production Weighing Unit

Summary of AD-4212C-300

  • Page 1

    Ad-4212c-300 ad-4212c-600 ad-4212c-3000 ad-4212c-6000 ad-4212c-301 ad-4212c-3100 production weighing unit instruction manual 1wmpd4001992b.

  • Page 2

    ? 2011 a&d company, limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of a&d company, limited. The contents of this manual and the specifications of the instru...

  • Page 3: Contents

    1 contents 1. Introduction..........................................................................................................................2 1.1. Features ...........................................................................................................................................

  • Page 4: 1.   Introduction

    2 1. Introduction this manual describes how the ad-4212c series balance works and how to get the most out of it in terms of performance. Read this manual thoroughly before using the balance and keep it at hand for future reference. 1.1. Features z a weighing unit, suitable for building into a produc...

  • Page 5

    3 example 1 (connecting to a personal computer or an ad-8922a) accessory rs-232c cable (10 m) personal computer usb converter (option: ax-usb-9p) ad-4212c ad-4212c usb port communication software “ winct-ad4212c” accessory rs-232c cable (10 m) ad-8922a personal computer communications cable (option:...

  • Page 6

    4 example 2 (connecting to a plc) . * specify the ad-8922a-01 connection cable when ordering the ad-8922a-01. Ax-ko3705-200 2 m ax-ko3705-500 5 m ax-ko3705-1000 10 m accessory rs-232c cable (10 m) ad-4212c ad-4212c ad-4212c bcd cable (fabricate locally) plc ad-8922a-01 connection cable* plc ad-8922a...

  • Page 7: 1.2. Compliance

    5 1.2. Compliance compliance with fcc rules please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a class a computing device pursuant to subpart j of part 15 of fcc rules. These rules are...

  • Page 8: 2. Unpacking The Balance

    6 windows communication tools (winct-ad4212c) ac adapter id label rs-232c cable (10 m) weighing pan pan support screws (flat head: m4 x 6) calibration weight 200 g oiml, class e2 or equivalent ac adapter breeze break accessories weighing unit leveling foot rs-232c cable (10 m) ac adapter jack ac ada...

  • Page 9

    7 attaching the ad-4212c series weighing unit directly on the mounting base when the ad-4212c series weighing unit is built into a system, remove the three leveling feet and use the screw holes to secure the weighing unit to the base. Attachment procedure 1. Remove the three leveling feet. Caution d...

  • Page 10: 2.2. Installing The Balance

    8 rear of the weighing unit rs-232c cable d-sub 9-pin arrow mark face up 2.2. Installing the balance install the balance as follows: 1. Refer to “3. Precautions” for installing the balance. 2. Refer to "2-1 unpacking" on page 6, to attach the pan support and the weighing pan on the weighing unit. 3....

  • Page 11

    9 mass of the locally fabricated weighing pan design the weighing pan so that the mass falls within the ranges shown in the table below: model status mass of the weighing pan (g) pan support is used (with weighing pan removed) 20 to 130 ad-4212c-300/301 pan support boss is used (with the weighing pa...

  • Page 12: 3.   Precautions

    10 3. Precautions to get the optimum performance from the balance and acquire accurate weighing data, note the following: 3.1. Before use install the weighing unit in an environment where the temperature and humidity are not excessive. The best operating temperature is about 20°c / 68°f at about 50%...

  • Page 13

    11 3.2. When building into a system the ad-4212c is a precision balance. When it is built into a system, errors such as unstable weight values may occur due to static electricity, vibration, sample temperature and magnetism of the materials weighed or used for the devices near the balance. When usin...

  • Page 14

    12 errors due to a static charge and measures to take when the ambient humidity is less than 45% rh, insulators such as plastic or glass are prone to static electricity. When charged material comes close to the balance, a pulling force is generated by static induction between the charged material an...

  • Page 15

    13 errors due to air flow and measures to take where the influence of ambient air flow is great such as: close to an air conditioner, door or passage way. Even very subtle air flow that is hard to be detected may influence the weighing operation. Avoid those areas as a weighing site. If weighing is ...

  • Page 16: 3.3. During Use

    14 3.3. During use to minimize the affect by electrical noises, earth ground the weighing unit and the display unit (option). Do not drop things upon the weighing pan, or place a sample on the pan that is beyond the balance weighing capacity. Place a sample in the center of the weighing pan to minim...

  • Page 17

    15 4. Operation of winct-ad4212c the software has the following functions: - set the weighing speed - set the minimum display value - calibration - set the baud rate • the parameters set using these functions, are stored in non-volatile memory, and maintained, even if the ac adapter is removed,. - a...

  • Page 18: 4.2. Calibration

    16 4.1. Changing the weighing speed using winct-ad4212c the weighing speed can be selected from the following three rates to minimize the influence on weighing that is caused by drafts and/or vibration at the place where the balance is installed. Indicator speed stability fast mid. Slow fast sensiti...

  • Page 19

    17 calibration procedure this function calibrates the balance using the calibration weight. (display example: ad-4212c-300) operation 1. Plug in the ac adapter and warm up the balance for 30 minutes or more with nothing on the pan. 2. Press the cal button. Cal 0 is displayed. If you want to cancel c...

  • Page 20

    18 4.3. Changing the baud rate using winct-ad4212c operation 1. Click the [settings] tab. 2. Select the baud rate to be used and click the set button. (factory setting: 2400) [settings] tab set button 3. “completed” appears when the setting is complete..

  • Page 21: 5. Smart Range Function

    19 5. Smart range function the ad-4212c-301/3100 are equipped with two ranges. The precision range has a higher resolution. The standard range has a normal resolution. The range switches automatically, depending on the value displayed. Sending the r command allows weighing in the precision range, re...

  • Page 22

    20 6. Communication specifications the ad-4212c series can communicate interactively using rs-232c. The ad-4212c series can continuously transmit the weighing data. (approx.10 times/second* to 50 times/second) the ad-4212c series is dce. The ad-4212c series can connect directly to the rs-232c termin...

  • Page 23

    21 2. Pin position pin position of the rs-232c cable (d-sub 9-pin side, female) pin no. Signal name direction description 1 (vs) - internally used * (power supply gnd terminal for external equipment) 2 txd output transmit data 3 rxd input receive data 4 - - n.C. 5 sg - signal ground 6 - - internally...

  • Page 24

    22 4. Command the balance can be controlled by the following commands. When transmitting, add the terminator to each command. Commands to query weighing data when the display is turned on, the ad-4212c series outputs the weighing data continuously. To use the q or s command, send the c command to st...

  • Page 25

    23 cal command example (pc, plc) external devices time weighing unit cal command displays the calibration weight value being weighed c cr lf a l weighing the zero point place the calibration weight calibration weight being weighed after the weight is removed, the balance returns to the weighing mode...

  • Page 26

    24 smp command example r command example when the precision range is used smp command s m p cr lf s t , + 0 0 0 0 . 0 1 2 g s t , + 0 0 0 0 0 . 0 1 g personal computer time weighing unit r cr lf the weight value exceeds 51 g. Place a sample. S t , + 0 0 0 . 0 0 0 0 g s t , + 0 0 0 . 0 0 0 0 g s t , ...

  • Page 27: 7. Maintenance

    25 7. Maintenance do not disassemble the balance. Contact the local a&d dealer if the balance needs service or repair. Use the original packing material for transportation. Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is moistened with warm water an...

  • Page 28: 8. Troubleshooting

    26 8. Troubleshooting 8.1. Checking the balance performance and environment the balance is a precision instrument. When the operating environment or the operating method is inadequate, performing accurate weighing will not be possible. If the balance seems to have a problem with repeatability or to ...

  • Page 29: 8.2. Asking For Repair

    27 sample and container has the sample absorbed or lost moisture? Has the temperature of the sample, and the container if used, been allowed to equalize to the ambient temperature? Refer to “3-2 when building into a system”. Is the sample charged with static electricity? Refer to “3-2 when building ...

  • Page 30: 9.   Specifications

    28 9. Specifications ad-4212c-300 ad-4212c-600 ad-4212c-301 weighing capacity 320 g 620 g 320 g / 51 g *1 maximum display 320.084 g 620.084 g 320.084 g / 51.0009 g minimum weighing value (1 digit) 0.001 g 0.001 g 0.001 g / 0.0001 g repeatability (standard deviation) 0.001 g 0.001 g 0.001 g / 0.0002 ...

  • Page 31

    29 ad-4212c-3000 ad-4212c-6000 ad-4212c-3100 weighing capacity 3200 g 6200 g 3200 g / 510 g *1 maximum display 3200.84 g 6200.84 g 3200.84 g / 510.009 g minimum weighing value (1 digit) 0.01 g 0.01 g 0.01 g / 0.001 g repeatability (standard deviation) 0.01 g 0.01 g 0.01 g / 0.002 g linearity ± 0.02 ...

  • Page 32: 10.  External Dimensions

    30 10. External dimensions - ad-4212c-300 / 600 / 3000 / 6000 / 301 / 3100 unit: mm 98 48 .8 30 18 .8 46 50 50 91.9 1 30 48.8 9.3 1 66 15.6 46 0.2 50 59 3 - m 6 80 .5 25 4 23 1 23 1 21 8 0. 2 50 (6 .5 ) 2- 4. 3 ho le 4 - m 4 si ze of w hen connect ed . A pprox. 30 pan support dimensions when securin...

  • Page 33: 11. Options

    31 11. Options ax-ko3590-200 rs-232c cable 2 m ax-ko3590-500 rs-232c cable 5 m ax-ko3590-1000 rs-232c cable 10 m (provided with the balance as a standard accessory.) ax-usb-9p: usb converter/cable set usb converter (d-sub 9 pin - usb) rs-232c cable (d-sub 9 pin - d-sub 9-pin) * this rs-232ccable is ...

  • Page 34

    32 ad-1684: electrostatic field meter this option measures the amount of the static charge on the sample, tare or peripheral equipment and displays the result. If those are found to be charged, discharge them using the ad-1683 dc static eliminator. Ad-8121b printer compact thermal dot-matrix printer...

  • Page 35: 12. Terms/index

    33 12. Terms/index terms stable value the weight value when the stabilization indicator appears. Environment ambient conditions such as vibration, drafts, temperature, static electricity or magnetic fields which affect the weighing operation. Calibration adjustment of the balance using the calibrati...

  • Page 36

    34 ax- ko3590-500 ..................................... 4, 31 ax- ko3590-1000 ................................... 4, 31 ax-usb-9p.............................................. 3, 31 b bcd..................................................... 2, 4, 31 breeze break ........................................

  • Page 37

    35 remote controller ........................................ 31 repeatability.............................. 26, 28, 29, 33 re-zero button ......................................... 15 rs-232c.............................................. 3, 4, 20 rs-232c cable .................................... 6,...