A&D AD-6121A Instruction Manual

Manual is about: Bedside scale

Summary of AD-6121A

  • Page 1

    ふ ad-6121a 1wmpd4001528.

  • Page 2

    This manual and marks all safety messages are identified by the following, “warning” or “caution”, of ansi z535.4 (american national standard institute: product safety signs and labels). The meanings are as follows: warning a potentially hazardous situation which, if not avoided, could result in dea...

  • Page 3

    English - 1 - ad-6121a contents 1. Precautions.............................................................................................................2 1.1. Safety precautions ..................................................................................................2 1.2. Handling preca...

  • Page 4

    Ad-6121a - 2 - english 1. Precautions 1.1. Safety precautions caution do not place the scale on a soft, wet or slippery floor. Place the scale on a solid and level floor. Do not place the scale in areas where it can be tripped over or kicked. Do not step on the scale if your feet are wet. Step on th...

  • Page 5

    English - 3 - ad-6121a 2. Compliance with european directives this device conforms to the following european directives. European directives 89/336/eec en61326 emc directive european directives 73/23/eec en61010-1 low voltage directive 3. Features ad-6121a is a verified compact, lightweight scale th...

  • Page 6

    Ad-6121a - 4 - english handle lcd spirit level zero switch on/off switch leveling foot buzzer switch terminal battery compartment ac adapter input jack leveling foot zero switch on/off switch display direction switch mode selection switch weighing surface 4. Part names and accessories accessories ac...

  • Page 7

    English - 5 - ad-6121a stable mark tare mark low battery mark unit display hold mark unit switches 4.1. Display symbols and switches display symbols symbols description stable mark this symbol appears when the displayed weight is stable. Tare mark this symbol appears when tare operation is in progre...

  • Page 8

    Ad-6121a - 6 - english buzzer switch selects whether to sound the buzzer or not. On sounds the buzzer when the scale is turned on or the scale holds the weight. Off does not sound the buzzer. Note the settings of the mode selection switch, display direction switch and buzzer switch become effective ...

  • Page 9

    English - 7 - ad-6121a 5.1.2. Using the batteries caution when installing the batteries, match the + (positive) and – (negative) terminals to those indicated in the battery compartment. Use four type aa batteries. Do not mix old and new batteries. Remove the batteries if the scale is not to be used ...

  • Page 10

    Ad-6121a - 8 - english 6. Using the scale 6.1. Basic weighing 1 make sure that nothing is placed on the scale. 2 press the on/off switch to turn the scale on. All the display segments turn on and then the zero display appears. When the ac adapter is used, the display backlight turns on. 3 press the ...

  • Page 11

    English - 9 - ad-6121a 6.2. Weighing with a tare in weighing with a tare, the scale stores the tare weight, turns on the tare mark and displays the net weight. The tare range is 3 kg to 150 kg. Net weight = gross weight – tare weight (example of tare items: clothes, plate and container) for weighing...

  • Page 12

    Ad-6121a - 10 - english 7. Calibration 7.1. About calibration calibration is the function to adjust the scale so that it can weigh accurately. When the verification seal is broken, re- verification is required. Break the verification seal and locate the cal switch cover on the scale bottom. Slide th...

  • Page 13

    English - 11 - ad-6121a 6 press the zero switch to adjust the temperature setting. When the display direction switch is set to ↑, each time the zero switch is pressed, the number increases by 1. When the display direction switch is set to ↓, each time the zero switch is pressed, the number decreases...

  • Page 14

    Ad-6121a - 12 - english of thermo-sensitive resistor and moves to the next item [c-11]. When the zero switch in place of the on/off switch is pressed, the scale moves to the next item [c-11] without storing the reference count for current output of thermo-sensitive resistor. 14 with [c-11] on the di...

  • Page 15

    English - 13 - ad-6121a 7.3. Adjusting the gravitational acceleration adjust the gravitational acceleration to that of the installation site. When the scale is transported to a different area, the adjustment is required. Note the gravity values for various locations are listed on page 19. 1 set the ...

  • Page 16

    Ad-6121a - 14 - english 8. Data communication 8.1. Rs-232c serial interface the ad-6121a scale has an rs-232c serial interface installed as standard. Using the optional rs-232c cable (ax-ko2238), the scale can send the data to external devices such as a personal computer. 8.1.1. Interface specificat...

  • Page 17

    English - 15 - ad-6121a 8.2. Connecting to a personal computer 8.2.1. Connection procedure the below is an example of connecting the scale to a windows-based personal computer. 1 insert the optional rs-232c cable (ax-ko2238) into the line (external output terminal) jack located on the bottom of the ...

  • Page 18

    Ad-6121a - 16 - english 9. Maintenance 9.1. Maintaining the performance quality check the scale regularly to make sure that it weighs correctly, and perform inspection or calibration as necessary. For details on inspection or calibration, contact the local a&d dealer. 9.2. Cleaning the scale to clea...

  • Page 19

    English - 17 - ad-6121a 10. Before asking for repair before asking for repair, refer to the table below. When what to do nothing appears on the display, even when the scale is turned on. Check whether the ac adapter or batteries are installed correctly. Check whether the batteries are drained. [----...

  • Page 20

    Ad-6121a - 18 - english 11. Specifications 11.1. Specifications type ad-6121a maximum weighing capacity 150 kg minimum display 100 g minimum weighing capacity 2 kg maximum tare maximum weighing capacity display lcd with backlight (backlight: only when the ac adapter is used.) displaying direction ca...

  • Page 21

    English - 19 - ad-6121a 12. Appendix 12.1. The value of gravity at various locations amsterdam 9.813 m/s 2 havana 9.788 m/s 2 paris 9.809 m/s 2 athens 9.807 m/s 2 helsinki 9.819 m/s 2 rio de janeiro 9.788 m/s 2 auckland, nz 9.799 m/s 2 kuwait 9.793 m/s 2 rome 9.803 m/s 2 bangkok 9.783 m/s 2 lisbon 9...

  • Page 22

    3-23-14 higashi-ikebukuro, toshima-ku, tokyo 170-0013 japan telephone: [81] (3) 5391-6132 fax: [81] (3) 5391-6148 a&d engineering, inc. 1756 automation parkway, san jose, california 95131 u.S.A. Telephone: [1] (408) 263-5333 fax: [1] (408)263-0119 a&d instruments ltd. Unit 24/26 blacklands way, abin...