A&D AD-8923-BCD Instruction Manual

Manual is about: Remote Controller (BCD)

Summary of AD-8923-BCD

  • Page 1

    Ad-8923-bcd remote controller (bcd) instruction manual 1wmpd4002137.

  • Page 2

    © 2010 a&d company, limited. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of a&d company, limited. The contents of this manual and the specifications of the instru...

  • Page 3: Contents

    1 contents 1. Introduction ....................................................................................... 2 1.1. Features ............................................................................................... 2 2. Description of each part ..............................................

  • Page 4: 1.   Introduction

    2 1. Introduction this manual describes how the ad-8923-bcd remote controller works and how to get the most out of it in terms of performance. Read this manual thoroughly before using the ad-8923-bcd and keep it at hand for future reference. 1.1. Features connecting the ad-8923-bcd remote controller...

  • Page 5

    3 accessories display ac adapter input jack rear view keys dc input terminal bcd connector rs-232c connector (d-sub 9-pin) fg terminal bcd plug 1 pc . . Ac adapter 1 pc. Ad adapter id label 2. Description of each part see the note below. Note when the ad-8923-bcd is to be built into a weighing syste...

  • Page 6: 2.1. Display

    4 2.1. Display z displays the weight data received. When the unit is “g” (gram), the unit indicator turns on. Z when the weight value is stable (the header of the weight data received is “st”), the stable indicator turns on. Z if the ad-8923-bcd does not receive the weight data for two seconds or mo...

  • Page 7: 3.   Connection

    5 3. Connection 3.1. Setting the weighing instrument and the ad-8923-bcd set the following items so that the weighing instrument and the ad-8923-bcd have the same value for each item. Item weighing instrument ad-8923-bcd baud rate 600, 1200, 2400*, 4800, 9600, 19200 bps data bits, parity 7 bits even...

  • Page 8: 3.3. Turning The Power On

    6 3.3. Turning the power on as a power supply, an external 24-vdc power supply (24 vdc±10% / 700ma) or a 12-vdc ac adapter can be used. Note when the ad-4212c is used, connecting power to either the ad-8923-bcd or the ad-4212c will supply power to both devices. So, instead of the ad-8923-bcd, using ...

  • Page 9

    7 z be sure to add a noise filter on the front end of the switching power supply and ground the fg terminal. • this will increase the resistance against noises. Z be sure to ground the fg terminal of the ad-8923-bcd and the ad-4212c. • this will increase the resistance against noises. Cable connecti...

  • Page 10: 3.4. Operation

    8 3.4. Operation z displays the data transmitted by the weighing instrument connected. Z the ad-8923-bcd key functions when connected to a weighing instrument are listed below: (e.G. When the ad-4212c is connected) keys of the ad-8923-bcd model on:off cal sample print mode re-zero ad-4212c switches ...

  • Page 11: Calibration Procedure

    9 calibration procedure calibrates the ad-4212c using the calibration weight. Operation 1. Warm up the ad-4212c for 30 minutes or more with nothing on the pan. 2. Press the cal key. Cal 0 is displayed. Z if you want to cancel calibration, press the cal key. The display will return to the weighing mo...

  • Page 12: 4.   Function Setting

    10 4. Function setting function setting specifies the ad-8923-bcd performance. The parameters are stored in non-volatile memory, and are maintained even if the power line or ac adapter is removed. The function setting menu consists of two layers. The first layer is the “class” and the second layer i...

  • Page 13: 4.1. Display And Keys

    11 4.1. Display and keys the stable indicator turns on to indicate that the parameter displayed is in effect. Selects a class or item. Changes the parameter. When a class is displayed, moves to an item in the class. When an item is displayed, stores the new parameter and displays the next class. Whe...

  • Page 14

    12 4.3. Initializing the ad-8923-bcd initialization restores the function settings of the ad-8923-bcd to factory settings. Operation 1 turn the power on. - - - - - or weighing mode display appears. 2 while holding down the on:off key, press the print key. Clr is displayed. 3 press the print key. To ...

  • Page 15: 5.  Rs-232C Connector

    13 5. Rs-232c connector the rs-232c cable provided with the ad-4212c can be connected directly. 5.1. Rs-232c serial interface specifications rs-232c transmission system : eia rs-232c transmission form : asynchronous, bi-directional, half duplex data format : baud rate : 600, 1200, 2400, 4800, 9600, ...

  • Page 16: 6.   Bcd Output Connector

    14 6. Bcd output connector outputs the weighing data received from the weighing instrument in bcd format, along with the polarity (+/-) and the data status (stable/unstable and over (e display)). Using the strobe signal, the data can be read easily. Busy input enables the data to be held or prevents...

  • Page 17

    15 input logic busy data will be held during on (when connected to input signal gnd). Switch input switch will be performed with on (when connected to input signal gnd). - all input, no voltage contact or open collector (connected to 5 v internally) plug (provided) part name product number manufactu...

  • Page 18

    16 6.2. Fixing of the decimal point position the ad-8923-bcd can set the display digit and the bcd output digit by setting dpp of the function. When fixing the decimal point position, the bcd output digit does not change if changing the minimum display digit by pressing the sample key. Example 1)whe...

  • Page 19: 7.   Troubleshooting

    17 7. Troubleshooting symptom description e error 10 appears. Communication settings of the ad-8923-bcd do not match with those of the weighing instrument. Check the settings such as baud rate and parity and correct them as necessary. For details, refer to “3.1. Setting the weighing instrument and t...

  • Page 20

    18 144 72 (1 16 ) 11 0 10 10 0 137 138 67 68 +1 0 +0.7 0 panel cutout dimensions ad-8923 remote controller on:off cal sample print mode re-zero stable fast mid. Response slow bcd out 8. Specifications power supply : external 24-vdc power supply (24 vdc±10% / 700ma) or ac adapter (output: 12 vdc / 30...