A&D Density Determination Kit AD-1653 BM Series Instruction Manual - page 13
11
5.4. Assembling the Kit: HA Series
Note: Do not use the front door to open or close the balance.
Assembled view
1. Remove the weighing pan, balance weight, pan
support, breeze break ring and dust plate from
the balance.
2. Place the density pan stand on the balance.
3. (Only for HA-202M)
Place one balance weight (small) on the density
pan stand.
4. Place the beaker stand so that it does not touch
the density pan stand.
5. Attach the thermometer clamp to the beaker
and insert the thermometer into the
thermometer clamp.
6. Pour a liquid with a known density (such as
distilled water) into the beaker and place the
beaker on the beaker stand.
7. Place the density pan on the density pan stand.
8. Adjust the amount of the liquid so that the sample,
when placed on the lower pan (in the liquid), is about
10 mm below the surface of the liquid.
9. When the value displayed on the balance is stable,
press the RE-ZERO key to set the displayed value
to zero.
Now the balance is ready for density measurement.
Weighing pan
Breeze break ring
Balance weight
Pan support
Dust plate
HA-202M only
Thermometer
Thermometer clamp
Beaker
Beaker stand
Density pan
Density pan stand
Balance weights (1 small, only for HA-202M)