A&D FG-150KAL Instruction Manual

Other manuals for FG-150KAL: Instruction Manual
Manual is about: FG Series Digital Platform Scales

Summary of FG-150KAL

  • Page 1

    F f g g s s e e r r i i e e s s d d i i g g i i t t a a l l p p l l a a t t f f o o r r m m s s c c a a l l e e s s fg-60kal / fg-150kal fg-30kam / fg-60kam / fg-150kam fg-30kbm / fg-60kbm / fg-150kbm wm+pd4000966.

  • Page 2

    This manual and marks all safety messages are identified by the following, “warning” or “caution”, of ansi z535.4 (american national standard institute: product safety signs and labels). The meanings are as follows: warning a potentially hazardous situation which, if not avoided, could result in dea...

  • Page 3

    1 contents 1. Introduction .............................................................................................. 2 2. Unpacking .................................................................................................... 2 3. Names and functions ........................................

  • Page 4: 1. Introduction

    2 1. Introduction this manual describes how this scale works and how to get the most out of it in terms of performance. Fg series platform scales have the following features: ? The fg series has three kinds of weight display resolution, 1/3,000, 1/6,000 (1/7,500) and 1/12,000 (1/15,000). ? There are...

  • Page 5: 3. Names and Functions

    3 3. Names and functions metric models weighing unit lcd display stable indicator net indicator zero indicator hi ok lo indicator print switch on/off switch mode switch re-zero switch spirit level fg-kbm series display column column support foot (fg-kam only) spirit level weighing pan display pod le...

  • Page 6

    4 u.S.A. Models indicators stable ? Indicates when the reading is stable. Net ? Indicates when net weight is displayed. (tare function is used.) zero ? Indicates when the scale zero is correct. Hi ok lo indicates when the scale zero is correct. Weighing units “kg” and “pcs” for metric models “lb”, “...

  • Page 7

    5 ? The re-zero , zero and tare switches work when the weight value is stable. ? The re-zero and zero switches will zero the scale if the weight is within ±2% of the weighing capacity (kg) around the power-on zero point. The zero indicator ? Turns on. ? If the weight exceeds +2% of the weighing capa...

  • Page 8: 4. Setting Up

    6 screws 4. Setting up 4-1. Attaching a display pod to the base (fg-kal and fg-kam) 1. First, remove the 4 screws from the bottom of the display column. 2. Set the display column to the base by pulling the cable into the base. ? Take care not pinch the cable between the column and the base. 3. Tight...

  • Page 9

    7 cautions for installing the scale consider the following conditions to get the most out of your scale. ? Install the scale where the temperature and relative humidity is stable. There is no draft and a stable power source is available. ? Install the scale on a solid and level surface. ? Do not ins...

  • Page 10: 5. Basic Operation

    8 5. Basic operation 5-1. Turning the power on and off 1. Press the on/off switch to turn the power on. All the display symbols are displayed and the scale waits for the weighing data to become stable. (only the units available illuminate.) after the weighing value internally becomes stable, the dis...

  • Page 11

    9 6. Remove the object from the weighing pan. ? The re-zero switch will zero the scale if the weight is within ±2% of the weighing capacity (kg) around the power-on zero point. The zero indicator ? Turns on. When the weight exceeds +2% of the weighing capacity (kg), it will be subtracted to zero as ...

  • Page 12: 6. Counting Mode

    10 6. Counting mode determines a unit weight (the weight of one piece) from sample pieces, and calculates how many pieces on the weighing pan using the unit weight. 1. Press the mode switch to select “pcs”. (“pcs” = pieces) 2. Press and hold the mode switch to enter the sample unit weight storing mo...

  • Page 13: 7. Comparator

    11 7. Comparator the results of the comparison are indicated by hi, ok or lo on the display. The formula to compare is as follows: lo operating conditions (see the “function list f6”): f6-0: no comparison (comparator function disabled). F6-1: to compare all data. F6-2: to compare more than +4d or le...

  • Page 14: 8. Calibration

    12 8. Calibration this function adjusts the scale for accurate weighing. Calibrate the scale in the following cases. ? When the scale is first used. ? When the scale has bee moved. ? When the ambient environment has changed. ? For regular calibration. Loose the lock screws on the rear side of the di...

  • Page 15

    13 6. Place the calibration weight on the pan with the same value as displayed, and wait until the stable indicator turns on. 7. Press the mode switch. The scale calibrate span and end will appear. Remove the weight from the pan, and turns the power off. ? Note the value set in step 5 is cleared aft...

  • Page 16: 9. Functions

    14 9. Functions the scale has function settings to expand your applications. The parameters set in the function settings are maintained even if the power switched off. 9-1. The procedure for setting parameters 1. Turn the power off. 2. Press and hold the zero switch and turn the power on via the on/...

  • Page 17

    15 9-2. Function list item description f 1- 0 auto power-off disabled ? F 1- 1 auto power-off enabled for battery use only auto power-off function f 1- 2 auto power-off enabled for battery and ac adapter automatically power off f 2- 0 normal (1/3,000 class) ? F 2- 1 high (1/6,000~1/7,500 class) disp...

  • Page 18: 10. Options

    16 10. Options the following options are available for the fg series: ? Op-23 (fg-23) rs-232c serial interface ? Op-24 (fg-24) rs-232c serial interface and comparator relay output ? Op-23 and op-24 cannot be used together. 10-1. Installation of op-23/op-24 the op-23/op-24 has an interface board, an ...

  • Page 19

    17 ax-ko1786-200 din 7p d-sub 9s 1 1 2 2 3 3 4 4 5 5 6 6 7 7 8 9 10-2. Op-23 rs-232c serial interface this interface allows fg series to be connected with a multi-function printer or a personal computer. ? The rs-232c interface has the following four modes. Stream mode outputs data continuously. Com...

  • Page 20

    18 pin connections data format s t , + 0 0 0 0 0 . 0 0 _ k g c r l f header data unit terminator ? There are 4 types of headers: st : stable weighing data qt : stable counting data us : unstable weighing data (including counting data) ol : out of weighing range (over) ? The data is normally 9 digits...

  • Page 21

    19 data output mode ? Stream mode function setting “f5-0” the scale outputs the current display data. The data-update rate is approximately 10 times per second. This rate is the same as the display-update. The scale does not output data while it is in setting mode. ? Print switch mode function setti...

  • Page 22

    20 reply to the command when the “f8-0” is selected, the scale reacts to the received command as follows. ? For the “q” command, the scale will send the data. ? For the “z” and “t” commands, the scale will send the same code as a reply after executing the command. Reply z c r l f reply t c r l f whe...

  • Page 23: 11. Maintenance

    21 11. Maintenance 11-1. Notes on maintenance ? Do not disassemble the scale. Contact your local a&d dealer if your scale needs service or repair. ? Please use the original packaging for transportation. ? Do not use organic solvents to clean the scale. Use a warm lint free cloth dampened with a mild...

  • Page 24: 12. Specifications

    22 12. Specifications 12-1. Specifications model fg-30kam fg-30kbm fg-60kam fg-60kbm fg-150kam fg-150kbm fg-60kal fg-150kal weight capacity (kg) 30 60 150 60 150 0.01 0.02 0.05 0.02 0.05 0.005 * 0.01 * 0.02 * 0.01 * 0.02 * min. Display (kg) 0.002 0.005 0.01 0.005 0.01 weight capacity (lb) 60 150 300...

  • Page 25

    23 12-2. Dimensions fg-60kal fg-150kal fg-30kam fg-60kam fg-150kam fg-30kbm fg-60kbm fg-150kbm 530 390 289 300 440 99 128 781 380 220 82 300 300 202 118 781 89 230 380 300 300 220 145 89 118 mm.

  • Page 26: Gravity Acceleration Map

    24 gravity acceleration map values of gravity at various locations amsterdam 9.813 m/s 2 manila 9.784 m/s 2 athens 9.807 m/s 2 melbourne 9.800 m/s 2 auckland nz 9.799 m/s 2 mexico city 9.779 m/s 2 bangkok 9.783 m/s 2 milan 9.806 m/s 2 birmingham 9.813 m/s 2 new york 9.802 m/s 2 brussels 9.811 m/s 2 ...

  • Page 27

    25 world map.

  • Page 28

    26 memo.

  • Page 29

    27 memo.

  • Page 30

    28 memo.

  • Page 31: A&d Instruments Ltd.

    3-23-14 higashi-ikebukuro, toshima-ku, tokyo 170-0013 japan telephone: [81] (3) 5391-6132 fax: [81] (3) 5391-6148 a&d engineering, inc . 1555, mccandless drive, milpitas, ca. 95035 u.S.A. Telephone: [1] (408) 263-5333 fax: [1] (408) 263-0119 a&d instruments ltd. Unit 24/26 blacklands way, abingdon b...