A&D FX-1200iWP Instruction Manual

Manual is about: Dust and waterproof Precision Balances FZ-iWP Series/FX-iWP Series

Summary of FX-1200iWP

  • Page 1

    Fz-120iwp / fz-200iwp / fz-300iwp fz-1200iwp / fz-2000iwp / fz-3000iwp fx-120iwp / fx-200iwp / fx-300iwp fx-1200iwp / fx-2000iwp / fx-3000iwp 1wmpd4001591a.

  • Page 2

    This is a caution mark. © 2007 a&d company ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of a&d company ltd. The contents of this manual and the specifications...

  • Page 3

    1 contents basic operation 1. Introduction ...................................................................................................................................... 3 1-1 about this manual .....................................................................................................

  • Page 4

    2 9-3 details of the function table ............................................................................................................ 37 9-4 description of the class “environment, display” .............................................................................. 40 9-5 description of...

  • Page 5

    3 1. Introduction this manual describes how the fz-i wpseries and the fx-i wpseries balance works and how to get the most out of it in terms of performance. Read this manual thoroughly before using the balance and keep it at hand for future reference. 1-1 about this manual this manual consists of th...

  • Page 6

    4 1-3 compliance compliance with fcc rules please note that this device generates, uses and can radiate radio frequency energy. This device has been tested and has been found to comply with the limits of a class a computing device pursuant to subpart j of part 15 of fcc rules. These rules are design...

  • Page 7

    5.

  • Page 8

    6.

  • Page 9

    7 2. Unpacking the balance 2-1 unpacking the balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be used for transporting the balance in the future. The packing contents depend on the balance model. See the illustrations to confirm that everything is contain...

  • Page 10

    8 2-2 installing the balance install the balance as follows: 1. Place the balance on a solid weighing table. Refer to “3. Precautions” for installing the balance. 2. In case of fz-i wpseries and fx-120i wp / fx-200i wp / fx-300i wp assemble the pan support, weighing pan and breeze break, on the bala...

  • Page 11

    9 3. Precautions to get the optimum performance from the balance and acquire accurate weight data, note the following: 3-1 before use the meaning of ip65 is "no ingress of dust. Projected against water jets". If a powerful water jet is used or the balance is immersed in water, the ingress of water m...

  • Page 12

    10 3-2 during use the fz-i wpseries and the fx-i wp series balance has a casing with high sealing performance due to its dust-tight and waterproof design. Therefore, a very subtle change in the atmospheric pressure in the installation site, for example, generated by opening or closing the door, affe...

  • Page 13

    11 prevent foreign materials, such as powder, liquid and metal, from invading the area around the weighing pan. The breeze break (fz-i wpseries and fx-120i wp / fx-200i wp/ fx-300i wp ) is provided as an accessory. An anti-static treatment has been applied to the breeze break components, but they ma...

  • Page 14

    12 3-5 display symbols and key operation key operation key operation affects how the balance functions. The basic key operations are: “press and release the key immediately” or “press the key” = normal key operation during measurement. “press and hold the key”. Display symbols each key, when pressed...

  • Page 15

    13 4. Weighing units 4-1 units with the fz-i wpseries and the fx-i wp series balance, the following weighing units and weighing modes are available: a unit or mode can be selected and stored in the function table as described on page 15. If a weighing mode (or unit of weight) has been turned off, th...

  • Page 16

    14 the tables below indicate the weighing capacity and the minimum display for each unit, depending on the balance model. Capacity unit fx-120 i wp fx-200 i wp fx-300 i wp minimum display gram 122.000 220.000 320.000 0.001 ounce (avoir) 4.30340 7.76025 11.28765 0.00005 pound 0.268965 0.485015 0.7054...

  • Page 17

    15 4-2 storing units the units or modes can be selected and stored in the function table. The sequence of displaying the units or modes can be arranged in the function table so as to fit the frequency of use. Select a unit or mode and arrange the sequence of display as follows: 1 press and hold the ...

  • Page 18

    16 5. Weighing 5-1 basic operation (gram mode) 1 place a container on the weighing pan, if necessary. Press the re-zero key to cancel the weight (tare). The balance displays 0.00 g. (the decimal point position depends on the balance model.) 2 place a sample on the pan or in the container. 3 wait for...

  • Page 19

    17 5-2 counting mode (pcs) this is the mode to determine the number of objects in a sample based on the standard sample unit mass. Unit mass means the mass of one sample. The smaller the variables in each sample unit mass are, the more accurate the counting will be. The fz-i wpseries and the fx-i wp...

  • Page 20

    18 if the balance judges that the mass of the samples is too light to aquire accurate weighing, it displays an error requiring the addition of more samples to the specified number. In the example above, 1 50- pcs appears, requiring 25 more samples. Add 25 samples and press the print key. When the un...

  • Page 21

    19 5-3 percent mode (%) this is the mode to display the weight value in percentage compared to a 100% reference mass and is used for target weighing or checking the sample variable. Selecting the percent mode 1 press the mode key to select % (percent mode). Storing the 100% reference mass 2 press th...

  • Page 22

    20 5-4 statistical calculation mode the statistical calculation mode statistically calculates the weight data, and displays or outputs the results. To use the statistical calculation mode, set the "application function (apf)" parameter of “application (ap fnc)” in the function table to “2”, as descr...

  • Page 23

    21 selecting the statistical items to output 5 press the sample key to display 5taf 0 . 6 press the re-zero key to select the output items. In the example, 5taf 1 is selected to output the number of data, sum, maximum, minimum, range (maximum-minimum) and average. Parameter description 0 number of d...

  • Page 24

    22 5-4-2 using the statistical calculation mode entering data for statistical calculation use the following keys to operate the statistical calculation mode. Mode key............ When the data is entered, moves between the displaying items (weighing mode, statistical results and data operation) each...

  • Page 25

    23 outputting the statistical results 5 each time the mode key is pressed, the display changes: the results as selected in “statistical function mode output items (5taf)”, cancel and clear . Notes when the number of data is 1, the standard deviation and coefficient of variation are displayed as - - ...

  • Page 26

    24 deleting the latest data when the wrong data is entered, it can be deleted and excluded from statistical calculation. Only the latest data can be deleted. 1 in the weighing mode, press the mode key, and then press the sample key several times to display cancel . 2 press the print key to display c...

  • Page 27

    25 5-5 statistical calculation mode (example of use) here, as an example of use of the statistical calculation mode, mixing of the multiple formulae such as medicine is described. The mixing process is recorded using the balance and the printer. In the example, the fx-3000i wp and the ad-8121b (dump...

  • Page 28

    26 5-5-2 using the statistical calculation mode 1 press the re-zero key to set the display to zero. 2 place a container on the weighing pan. Press the print key to cancel the weight (tare). The balance displays 0.00 g. (storing the tare value) the tare value data is output when the peripheral output...

  • Page 29

    27 6. Response adjustment this function stabilizes the weight value, reducing the influence on weighing that is caused by drafts and vibration at the place where the balance is installed. The function has three stages as follows and can be changed by simple key operation. Indicator description fast ...

  • Page 30

    28 7. Calibration 7-1 calibration mode the fz-i wpseries and the fx-i wpseries balance has the following two modes. Calibration using the internal mass (one-touch calibration, only for the fz-i wpseries) calibration using an external weight calibration test using an external weight (calibration test...

  • Page 31

    29 7-2 calibration using the internal mass (one-touch calibration, only for fz-iwp series) this function calibrates the balance using the internal mass. The only operation required is to press the cal key operation 1 connect the ac adapter and warm up the balance for at least 30 minutes with nothing...

  • Page 32

    30 7-3 calibration using an external weight this function calibrates the balance using an external weight. Operation 1 connect the ac adapter and warm up the balance for at least 30 minutes with nothing on the pan. 2 press and hold the cal key until calout is displayed, and then release the key. 3 t...

  • Page 33

    31 5 confirm that there is nothing on the pan and press the print key. The balance measures the zero point. Do not allow vibration or drafts to affect the balance. The balance displays the calibration weight value. 6 place the displayed calibration weight on the pan and press the print key. The bala...

  • Page 34

    32 7-4 calibration test using an external weight this function tests the balance weighing accuracy using an external weight and outputs the result. This is available only when the “glp output (info)” parameter is set to “1” or “2”. (calibration test does not perform calibration.) operation 1 connect...

  • Page 35

    33 5 confirm that there is nothing on the pan and press the print key. The balance measures the zero point and displays the measured value. Do not allow vibration or drafts to affect the balance. The balance displays the target weight value. 6 place the displayed target weight on the pan and press t...

  • Page 36

    34 8. Function switch and initialization 8-1 permit or inhibit the balance stores parameters that must not be changed unintentionally. There are two switches for the purpose of protecting these parameters. Each switch can select either “permit” or “inhibit”. The “inhibit” protects parameters against...

  • Page 37

    35 operation 1 press the on:off key to turn off the display. 2 while pressing and holding the print key and the sample key, press the on:off key. The balance displays p5 . 3 press the print key. Then the balance displays the function switches. 4 set the switches using the following keys. Sample key ...

  • Page 38

    36 9. Function table the function table reads or rewrites the parameters that are stored in the balance. These parameters stored, even if the ac adapter is removed, are maintained in non-volatile memory. 9-1 structure and sequence of the function table the function table menu consists of two layers....

  • Page 39

    37 9-3 details of the function table class item and parameter description 0 1 cond condition 2 fast response, sensitive value fast mid. Slow response, stable value slow with “hold 1”, sets the averaging time. 0 1 5t-b stability band width 2 stable range is ±1 digit stable range is ±3 digits the stab...

  • Page 40

    38 class item and parameter description 0 key mode accepts the print key only when the display is stable. 1 auto print mode a (reference = zero) 2 auto print mode b (reference = last stable value) outputs data when the display is stable and conditions of ap-p, ap-b and the reference value are met. 3...

  • Page 41

    39 class item and parameter description 0 600 bps 1 1200 bps 2 2400 bps 3 4800 bps 4 9600 bps bp5 baud rate 5 19200 bps 0 7 bits, even 1 7 bits, odd btpr data bit, parity bit 2 8 bits, none 0 cr lf crlf terminator 1 cr cr: ascii code 0dh lf: ascii code 0ah 0 a&d standard format 1 dp format 2 kf form...

  • Page 42

    40 9-4 description of the class “environment, display” condition ( cond ) cond 0 this parameter is for sensitive response to the fluctuation of a weight value. Used for powder target weighing, weighing a very light sample or when quick response weighing is required. After setting, the balance displa...

  • Page 43

    41 zero tracking ( trc ) this function tracks zero point drift caused by changes in the environment and stabilizes the zero point. When the weight data is only a few digits, turn the function off for accurate weighing. Note digit, when used for the fx-i wp series balance, indicates a unit of minimum...

  • Page 44

    42 9-5 description of the item “data output mode” the parameter setting of the “data output mode ( prt )” applies to the performance when the data is transmitted using the rs-232c serial interface. Key mode when the print key is pressed with the stabilization indictor turned on, the balance outputs ...

  • Page 45

    43 interval output mode the weight data is periodically output. When the print key is pressed, the balance starts to output the weight data at a preset interval time. When the print key is pressed again, the balance stops outputting the weight data. Required setting dout prt 6 interval output mode d...

  • Page 46

    44 kf format 5if type 2 this is the karl-fischer moisture meter format and is used when the peripheral equipment can only communicate using this format. This format consists of fourteen characters excluding the terminator. This format has no header characters. The polarity sign is placed before the ...

  • Page 47

    45 9-7 data format examples.

  • Page 48

    46 note when “pound ounce” is selected, the data is output with the unit of ounce (oz). 9-8 description of the item “application function” capacity indicator ( apf 1) in the weighing mode, the indicator displays the weight data relative to the weighing capacity in percentage. (zero = 0%, maximum cap...

  • Page 49

    47 9-9 comparator function the results of the comparison are indicated by hi ok lo on the display. Operating conditions: no comparison comparison when the weight data is stable or overloaded, excluding “near zero” comparison when the weight data is stable or overloaded, including “near zero” continu...

  • Page 50

    48 9-10 clock and calendar function the balance is equipped with a clock and calendar function. When the “glp output (info)” parameter is set to “1” or “2” and the “time/date output (5-td)” parameter is set to “1”, “2” or “3”, the time and date are added to the output data. Set or confirm the time a...

  • Page 51

    49 quitting the operation 8 the balance displays the next menu of the function table. Press the cal key to exit the clock and calendar function and return to the weighing mode. Notes do not enter invalid values such as a non-existing date when setting the time and date. When the clock backup battery...

  • Page 52

    50 10. Id number and glp report the id number is used to identify the balance when good laboratory practice (glp) is used. The id number is maintained in non-volatile memory even if the ac adapter is removed. The glp output format is selected at “glp output (info)” of the function table and can be o...

  • Page 53

    51 calibration report using an external weight when the setting is “info 1”: when the setting is “info 2”: calibration test report using an external weight (calibration test does not perform calibration.) when the setting is “info 1”: when the setting is “info 2”: ad-8121 format general data format ...

  • Page 54

    52 title block and end block when a weight value is recorded as the glp data, “title block” and “end block” are inserted at the beginning and at the end of a group of weight values, in the glp report. Note to output the report to an ad-8121b , use mode 3 of the ad-8121b. Operation 1 with the weight ...

  • Page 55

    53 11. Underhook the underhook can be used for magnetic materials or density measurement. The built-in underhook is revealed by removing the plastic cap on the bottom of the balance. Use the underhook as shown below. Caution do not apply excessive force to the underhook. When the underhook is not in...

  • Page 56

    54 12. Programmable-unit this is a programmable unit conversion function. It multiplies the weight data in grams by an arbitrary coefficient set in the function table and displays the result. The coefficient must be within the range between the minimum and maximum shown below. If the coefficient set...

  • Page 57

    55 13. Rs-232c serial interface the balance is a data communication equipment (dce) device. Connect the balance to a personal computer (dte) using a straight through cable. Transmission system : eia rs-232c (d-sub 9-pin, female connector) transmission form : asynchronous, bi-directional, half duplex...

  • Page 58

    56 14. Connection to peripheral equipment 14-1 connection to the ad-8121b printer set the following parameters to use the ad-8121b printer. Example of use ad-8121b mode setting to print a&d standard format weight data, using the fx-i wp print key or fx-i wpauto print mode. (the time and date can be ...

  • Page 59

    57 14-2 connection to a computer the fx-iwpseries balance can be connected to a personal computer using the rs-232c serial interface. As an option, the fxi–02 usb interface is available to transmit the balance data to a personal computer. The fx-iwpseries balance is a dce. Use a straight through cab...

  • Page 60

    58 14-3 using windows communication tools (winct) when windows is used as an operating system in a personal computer, the winct software, that is downloaded from the a&d website, can be used to transmit the weight data to the personal computer. The winct has two communication methods: “rscom” and “r...

  • Page 61

    59 15. Commands 15-1 command list note a command has a terminator added, that is specified using the "terminator (crlf)" parameter of “serial interface ( 5if )” in the function table, and is sent to the balance. Commands to query weight data c cancels the s or sir command. Q requests the weight data...

  • Page 62

    60 15-2 acknowledge code and error codes when the “ak, error code (ercd)” parameter of “serial interface ( 5if )” is set to “1”, the balance outputs code or an error code for each command as follows: (06h) acknowledge in ascii code. When the balance receives a command to request data and can not pro...

  • Page 63

    61 t command example pt command example when a communication error has occurred due to external noise, or a parity error has occurred due to transmission error, the balance transmits an error code. In this case, send the command again. 15-3 settings related to the rs-232c concerning the rs-232c, the...

  • Page 64

    62 16. Maintenance in normal use, the balance can be cleaned with water. But, keep the following precautions so that dust and water do not invade the balance. Do not direct water pressure at the bottom of the balance. Do not use powerful water jets. Do not immerse the balance in water. Do not use or...

  • Page 65

    63 17. Troubleshooting 17-1 checking the balance performance and environment the balance is a precision instrument. When the operating environment or the operating method is inadequate, correct weighing can not be performed. Place a sample on the pan and remove it, and repeat this several times. If ...

  • Page 66

    64 17-2 error codes display error code description ec, e11 stability error the balance can not stabilize due to an environmental problem. Prevent vibration, drafts, temperature changes, static electricity and magnetic fields. Confirm that the weighing pan is properly installed. Confirm that the wate...

  • Page 67

    65 display error code description ec, e00 communications error a protocol error occurred in communications. Confirm the format, baud rate and parity. Ec, e01 undefined command error an undefined command was received. Confirm the command. Ec, e02 not ready a received command can not be processed. E.G...

  • Page 68

    66 18. Options notes the fxi-02, fxi-08 and fxi-09 can not be used at the same time. The fx-i wp series balance does not comply with ip65 when the fxi-02, fxi-08 or fxi-09 and ax-usb-9p-ex is used with the balance. Fxi-02 usb interface (installed in the balance, applicable os: windows 98 osr2 or lat...

  • Page 69

    67 fxi-wp-10 small breeze break provided for the fz-i wp series and fx-120i wp/ fx-200i wp / fx-300i wpas standard. Can be used for fx-1200i wp/ fx-2000i wp / fx-3000i wp. Fxi-wp-11 large breeze break breeze break with greater height. Can be used for all models. Convenient when a large beaker or mea...

  • Page 70

    68 ad-1683 dc static eliminator used to minimize weighing errors due to static electricity on the material. The ad-1683 is direct-current static eliminator. The ions generated produce no breeze and are effective over a long distance. Therefore, the balance can accurately weight powders, etc. By usin...

  • Page 71

    69 19. Specifications fz-120 i wp fz-200 i wp fz-300 i wp fz-1200 i wp fz-2000 i wp fz-3000 i wp weighing capacity 122 g 220 g 320 g 1220 g 2200 g 3200 g maximum display 122.084 g 220.084 g 320.084 g 1220.84 g 2200.84 g 3200.84 g minimum weighing value (1 digit) 0.001 g 0.01 g repeatability (standar...

  • Page 72

    70 fx-120 i wp fx-200 i wp fx-300 i wp fx-1200 i wp fx-2000 i wp fx-3000 i wp weighing capacity 122 g 220 g 320 g 1220 g 2200 g 3200 g maximum display 122.084 g 220.084 g 320.084 g 1220.84 g 2200.84 g 3200.84 g minimum weighing value (1 digit) 0.001 g 0.01 g repeatability (standard deviation) 0.001 ...

  • Page 73

    71 20. External dimensions fz-120i wp/fz-200i wp/ fz-300i wp fx-120i wp/fx-200i wp/ fx-300i wp * 1: width of the opening when a clear plate is removed * 2: inside dimension * 3: weighing pan diameter * 4: height from the weighing pan up to the cap of the breeze break (inside dimension) * 5: height o...

  • Page 74

    72 fxi-wp-11 large breeze break unit: mm 19 9. 5 112 (i nt er nal s iz e ab ov e th e p an) (opening) 294 6 156 (when fully opened) (o pen ing) 270 105 31 198 107 (when fully opened) (when fully opened) (when fully opened).

  • Page 75

    73 21. Terms / index terms stable value the weight data when the stabilization indicator appears. Environment ambient conditions such as vibration, drafts, temperature, static electricity and magnetic fields which affect the weighing operation. Calibration adjustment of the balance so that it can we...

  • Page 76

    74 index keys and symbols on/off key ............................ 12, 35 sample key ........................... 12, 15 mode key ..........................12, 13, 15 cal key................................... 12, 32 print key ............................... 12, 15 re-zero key.....................10, ...

  • Page 77

    75 display at start................................................... 37 display refresh rate ..................................... 37, 41 dout data output ..................................... 38 dp format .................................................. 39, 43 dte .................................

  • Page 78

    76 permit ........................................................ 34 pnt decimal point ............................ 37, 41 poff auto display-off ...................... 37, 41 p-on auto display-on........................ 37, 41 print key .................................................. 12, 15 prin...

  • Page 80

    3-23-14 higashi-ikebukuro, toshima-ku, tokyo 170-0013 japan telephone: [81] (3) 5391-6132 fax: [81] (3) 5391-6148 a&d engineering, inc. 1756 automation parkway, san jose, california 95131 u.S.A. Telephone: [1] (408) 263-5333 fax: [1] (408)263-0119 a&d instruments limited unit 24/26 blacklands way, a...