A&D GP-100K Instruction Manual

Manual is about: GP SERIES Precision Balance

Summary of GP-100K

  • Page 1

    Gp series gp-12k/gp-20k /gp-30k/gp-32k gp-40k/gp-60k/gp-61k gp-100k/gp-102k gp-30ks/gp-32ks gp-60ks/ gp-61ks/gp-100ks precision balance instruction manual wm+pd4000273d.

  • Page 2

    ? 2006 a&d company ltd. All rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of a&d company ltd. The contents of this manual and the specifications of the instrument cover...

  • Page 3: Contents

    1 contents basic operation 1. Introduction ............................................................................................................................ 3 1-1 about this manual ...............................................................................................................

  • Page 4

    2 9-8 data format examples .....................................................................................................................46 9-9 clock and calendar function............................................................................................................48 9-10 compar...

  • Page 5: 1. Introduction

    3 1. Introduction this manual describes how the gp series balance works and how to get the most out of it in terms of performance. Read this manual thoroughly before using the balance and keep it at hand for future reference. 1-1 about this manual this manual consists of the following five parts: ba...

  • Page 6

    4 hold function, provided for weighing a moving object such as an animal. Multiple weighing units, with most of the common units used around the world. Density mode, for calculating the density of a solid. Accumulation function, adding the weight values and outputting the sum. Adjustable display pos...

  • Page 7: 1-3 Compliance

    5 1-3 compliance compliance with fcc rules please note that this equipment generates, uses and can radiate radio frequency energy. This equipment has been tested and has been found to comply with the limits of a class a computing device pursuant to subpart j of part 15 of fcc rules. These rules are ...

  • Page 8

    6.

  • Page 9: 2.  Unpacking The Balance

    7 2. Unpacking the balance 2-1 unpacking the balance is a precision instrument. Unpack the balance carefully. Keep the packing material to be used for transporting the balance in the future. The packing contents depend on the balance model. See the illustrations to confirm that everything is contain...

  • Page 10

    8 separate display type: gp-30ks/32ks/60ks/61ks/100ks note please confirm that the ac adapter type is correct for your local voltage and receptacle type. 2-2 installing the balance install the balance as follows: 1. Refer to “3. Precautions” for installing the balance. 2. Adjust the leveling feet to...

  • Page 11: 3. Precautions

    9 3. Precautions to get the optimum performance from the balance and acquire accurate weighing data, note the following: 3-1 before use install the balance in an environment where the temperature and humidity are not excessive. The best operating temperature is about 20°c / 68°f at about 50% relativ...

  • Page 12: 3-3 After Use

    10 this balance uses a strong magnet as part of the balance assembly, so please use caution when weighing magnetic materials such as iron. If there is a problem, use the underhook on the bottom of the balance to suspend the material away from the influence of the magnet. Cancel the temperature diffe...

  • Page 13

    11 3-5 display symbols and key operation key operation key operation affects how the balance functions. The basic key operations are: “press and release the key immediately” or “press the key” = normal key operation during measurement “press and hold the key” display symbols each key, when pressed o...

  • Page 14: 3-6  Smart Range Function

    12 3-6 smart range function the gp-32k, gp-32ks and gp-102k are equipped with two ranges. The precision range has a higher resolution. The standard range has normal resolution. The range is switched automatically, depending on the value displayed. Pressing the re-zero key allows weighing in the prec...

  • Page 15: 4. Weighing Units

    13 4. Weighing units 4-1 units with the gp series balance, the following weighing units and weighing modes are available : a unit or mode can be selected and stored in the function table as described in “4-2 changing the units”. If a weighing mode (or unit of weight) has been turned off, that mode o...

  • Page 16

    14 the tables below indicate the weighing capacity and the minimum display for each unit, depending on the balance model. Gp-12k gp-20k gp-30k gp-30ks gp-61k gp-61ks gp-40k unit capacity minimum display capacity minimum display gram 12000.0 21000.0 31000.0 61000.0 0.1 41000.0 0.5 kilogram 12.0000 21...

  • Page 17

    15 gp-32k/ gp-32ks standard range precision range unit capacity minimum display capacity minimum display gram 31000 1 6100.0 0.1 kilogram 31.000 0.001 6.1000 0.0001 ounce (avoir) 1093.50 0.05 215.170 0.005 pound 68.345 0.005 13.4480 0.0005 pound/ounce 68l 5.5 0.1 13l 7.17 0.01 troy ounce 996.65 0.05...

  • Page 18: 4-2  Changing The Units

    16 4-2 changing the units the units or modes can be selected and stored in the function table. The sequence of displaying the units or modes can be arranged so as to fit the frequency of use in the function table. Select a unit or mode and arrange the sequence of display as follows: 1 press and hold...

  • Page 19: 5. Weighing

    17 5. Weighing 5-1 basic operation (kilogram mode) 1 place a container on the weighing pan, if necessary. Press the re-zero key to cancel the weight (tare). The balance displays 0.0000 kg. (the decimal point position depends on the balance model.) 2 place a sample on the pan or in the container. 3 w...

  • Page 20: 5-2  Counting Mode (Pcs)

    18 5-2 counting mode (pcs) this is the mode to determine the number of objects in a sample based on the standard sample unit mass. Unit mass means the mass of one sample. The smaller the variables in each sample unit mass is, the more accurate the counting will be. The gp series balance is equipped ...

  • Page 21

    19 counting operation 7 place the samples to be counted on the pan. Note up to 50 unit masses can be stored in memory for the multiple sample. For details, refer to “11. Data memory”. Counting mode using the acai function the acai is a function that improves the accuracy of the unit mass automatical...

  • Page 22: 5-3  Percent Mode (%)

    20 5-3 percent mode (%) this is the mode to display the weight value in percentage compared to a 100% reference mass and is used for target weighing or checking the sample variable. Selecting the percent mode 1 press the mode key to select % (percent mode). If the percent mode can not be selected, r...

  • Page 23: 5-5 Accumulation Function

    21 5-5 accumulation function the accumulation function adds the weighing data and displays the total value. To use the accumulation function, set the “accumulation function (add)” parameter of the function table as described below. Note while the accumulation function is in use, the data memory func...

  • Page 24

    22 using the accumulation function use the keys below to operate the accumulation function. Mode key: displays the weighing data and the total value alternately each time it is pressed. Will not change the unit while the accumulation function is in use. Re-zero key: sets the display to zero while th...

  • Page 25

    23 6. Response adjustment / self check function this function detects the influence on weighing that is caused by drafts and/or vibration at the place where the balance is installed and sets the response characteristic automatically. When this function is selected, the balance self-checks the perfor...

  • Page 26

    24 notes if improper performance is found in the self check, the balance displays ch no. Contact the local a&d dealer for repair. If the automatic response adjustment fails, the balance displays ch ng. Check the ambient conditions such as breeze and vibration, also check the weigning pan. Then, perf...

  • Page 27: 7. Calibration

    25 7. Calibration 7-1 calibration group the gp series balance has the following modes as a calibration group. Calibration automatic self calibration (calibration due to changes in temperature) calibration using the internal mass (one-touch calibration) calibration using an external weight calibratio...

  • Page 28

    26 display this indicator means “the balance is measuring calibration data”. Do not allow vibration or drafts to affect the balance while this indicator is displayed. 7-2 automatic self calibration (calibration due to changes in temperature) this function automatically calibrates the balance when th...

  • Page 29

    27 7-3 calibration using the internal mass (one-touch calibration) this function calibrates the balance using the internal mass. The only operation required is to press the cal key. Operation 1 plug in the ac adapter and warm up the balance for at least 30 minutes with nothing on the weighing pan. 2...

  • Page 30

    28 7-4 calibration using an external weight this function calibrates the balance using an external weight. Operation 1 plug in the ac adapter and warm up the balance for at least 30 minutes with nothing on the pan. 2 press and hold the cal key until calout is displayed, then release the key. 3 the b...

  • Page 31

    29 5 confirm that there is nothing on the pan and press the print key. The balance measures the zero point. Do not allow vibration or drafts to affect the balance. The balance displays the calibration weight value. 6 place the displayed calibration weight on the pan and press the print key. The bala...

  • Page 32

    30 7-5 calibration test using an external weight this function tests the balance weighing accuracy using an external mass and outputs the result. This is available only when the “glp output (info)” parameter is set to “1” or “2”. (calibration test does not perform calibration.) operation 1 connect t...

  • Page 33

    31 5 confirm that there is nothing on the pan and press the print key. The balance measures the zero point and displays the measured value. Do not allow vibration or drafts to affect the balance. The balance displays the target weight value. 6 place the displayed target weight on the pan and press t...

  • Page 34

    32 7-6 correcting the internal mass value the gp series balance can correct the internal mass value within the range shown below. This function corrects the internal mass value to conform to an external weight. The corrected mass value is maintained in non-volatile memory even if the ac adapter is r...

  • Page 35

    33 operation 1 calibrate the balance using the internal mass. (one-touch calibration) then, place an external weight and confirm the value to be corrected. In the example, the value is to be corrected by 0.6 gram in 20 kilograms. 2 press the on:off key to turn off the display. 3 while pressing and h...

  • Page 36: 8-1  Permit Or Inhibit

    34 8. Function switch and initialization 8-1 permit or inhibit the balance stores parameters that must not be changed unintentionally (e.G. Calibration data for accurate weighing, data for adapting to the operating environment, control data for the rs-232c interface). There are five switches for the...

  • Page 37

    35 8-2 initializing the balance this function returns the following parameters to factory settings. Calibration data function table the sample unit mass value (counting mode), 100% reference mass value (percent mode) the data that is stored in the balance using the data memory function external cali...

  • Page 38: 9. Function Table

    36 9. Function table the function table reads or rewrites the parameters that are stored in the balance. These parameters are maintained in non-volatile memory, even if the ac adapter is removed. 9-1 structure and sequence of the function table the function table menu consists of two layers. The fir...

  • Page 39

    37 9-3 details of the function table p.

  • Page 40

    38.

  • Page 41

    39 caution the balance may not transmit the data completely at the specified refresh rate, depending on the baud rate or data added to the weighing data such as time, date and id number..

  • Page 42

    40 9-4 description of the class “environment, display” condition ( cond ) cond 0 this parameter is for sensitive response to the fluctuation of a weight value. Used for powder target weighing, weighing a very light sample or when quick response weighing is required. After setting, the balance displa...

  • Page 43

    41 zero tracking ( trc ) this function tracks zero point drift caused by changes in the environment and stabilizes the zero point. When the weighing data is only a few digits, turn the function off for accurate weighing. Note digit, when used for the gp series balance, indicates the smallest display...

  • Page 44

    42 9-5 description of the item “data output mode” the parameter setting of “data output mode (prt)” applies to the performance when the “data memory (data)” parameter is set to “2” (to store the weighing data) and when the data is transmitted using the rs-232c interface. Key mode when the print key ...

  • Page 45

    43 interval memory mode the weighing data is periodically stored in memory. Required setting dout prt 3 interval memory mode dout data 2 data memory function is used. Stores weighing data. Dout int interval time optional setting dout 5-td1, 2, or 3 adds the time and date. Example “for periodical wei...

  • Page 46

    44 kf format 5if type 2 this is the karl-fischer moisture meter format and is used when the peripheral equipment can only communicate using this format. This format consists of fourteen characters excluding the terminator. This format has no header characters. The polarity sign is placed before the ...

  • Page 47: 9-7

    45 9-7 description of the data format added to the weighing data id number dout 5-id 1 the number to identify a specific balance. This format consists of seven characters excluding the terminator. Data number dout d-no 1 this format outputs the data number just before the data is transmitted using t...

  • Page 48: 9-8  Data Format Examples

    46 9-8 data format examples.

  • Page 49

    47 note when “pound ounce” is selected, the data is output with the unit of ounce (oz)..

  • Page 50

    48 9-9 clock and calendar function the balance is equipped with a clock and calendar function. When the “glp output (info)” parameter is set to “1” or “2” and the “time/date output (5-td)” parameter is set to “1”, “2” or “3”, the time and date are added to the output data. Set or confirm the time an...

  • Page 51

    49 confirming the date 6 the current date is displayed with all the digits blinking. To change the display order of year (y), month (n) and day (d), press the mode key. The date is output in the order as specified. When the date is correct and the operation is to be finished, press the cal key and p...

  • Page 52: 9-10 Comparator Function

    50 9-10 comparator function the results of the comparison are indicated by hi ok lo on the display. Operating conditions: no comparison comparison when the weighing data is stable or overloaded, excluding “near zero” comparison when the weighing data is stable or overloaded, including “near zero” co...

  • Page 53

    51 weighing input mode press the re-zero key. The balance displays 0.0 g. Place a sample, with a mass that corresponds to the upper limit value, on the pan. Press the print key to store the upper limit value. Remove the sample. The balance displays cp lo . 9 with cp lo displayed, press the print key...

  • Page 54

    52 9 with cp lnt displayed, press the print key. The current setting of the tolerance value is displayed with all the digits blinking. When the current setting is not to be changed, press the print or cal key to proceed to step10. When the current setting is to be changed, press the re-zero key. The...

  • Page 55

    53 setting the function table 2 press and hold the sample key until ba5fnc of the function table is displayed, then release the key. 3 press the sample key several times to display cp fnc . 4 press the print key. 5 press the sample key several times to display cp-b 0 . 6 press the re-zero key to dis...

  • Page 56

    54 10. Id number and glp report the id number is used to identify the balance when good laboratory practice (glp) is used. The id number is maintained in non-volatile memory even if the ac adapter is removed. The glp output format is selected at “glp output (info)” of the function table and can be o...

  • Page 57: 10-2 Glp Report

    55 10-2 glp report set the following parameters to output the report. To print the report, set the “glp output (info)” parameter to “1” and use mode 3 of the ad-8121b. For details on using the printer, refer to “15-1 connection to the ad-8121b printer”. To output the report to a personal computer us...

  • Page 58

    56 calibration report using an external weight when the setting is “info 1”: when the setting is “info 2”: calibration test report using an external weight (calibration test does not perform calibration.) when the setting is “info 1”: when the setting is “info 2”:.

  • Page 59

    57 title block and end block when weight values are recorded as glp data, a “title block” is inserted at the beginning and an “end block” is inserted at the end of a group of weight values in the glp report. Note to output the report to an ad-8121b , use mode 3 of the ad-8121b. Caution if the data m...

  • Page 60: 11. Data Memory

    58 11. Data memory data memory is a function to store weighing data, calibration data and unit mass in memory. Of the data in memory, the balance can only display the weighing data. The weighing data and calibration data in memory are available for outputting at one time to a printer or personal com...

  • Page 61

    59 11-2 memory for weighing data the data memory function can store 200 sets of weighing data (100 set if time and date are added). Even if the ac adapter is removed, the data is maintained in non-volatile memory. It is not necessary that the printer or computer be continually connected to the balan...

  • Page 62

    60 the following commands can not be used during data storage. Q query command for weighing data. S query command for stable weighing data. Si query command for weighing data. Sir query command for continuous weighing data. Setting the function table parameter settings for each output mode are as fo...

  • Page 63

    61 recalling the memory data confirm that the “data memory (data)” parameter is set to “2”. 1 press and hold the print key until recall is displayed, then release the key. 2 press the print key to enter the memory recall mode. The type of data appears in the upper left of the display as shown to the...

  • Page 64

    62 11-3 memory for calibration and calibration test data calibration data (when and how it is performed) and calibration test data can be stored in memory. All the data in memory is available to be output at one time to a printer or personal computer. Up to 50 data sets of the latest calibration or ...

  • Page 65

    63 11-4 memory for unit mass in the counting mode the data memory function can store 50 sets of unit mass for the counting mode. Even if the ac adapter is removed, the data is maintained in non-volatile memory. Among the 50 sets, “p01” is the memory function for the standard counting mode. The unit ...

  • Page 66

    64 weighing input mode in the weighing input mode, the specified number of samples is placed on the pan to store the unit mass. Re-storing the unit mass or performing automatic counting accuracy improvement (acai) on the re-stored unit mass is possible. Follow the procedure described in “5-2 countin...

  • Page 67

    65 11-5 memory for comparator settings the data memory function can store 20 sets of upper and lower limit values for the comparator mode. The reference value or tolerance value for the comparator mode can not be stored in memory. The upper and lower limit values in memory can be recalled easily usi...

  • Page 68

    66 digital input mode in the digital input mode, the upper and lower limit values are entered digitally using the keys. The display in the digital input mode is shown below. Use the following keys to store upper and lower limit values in digital input mode. Sample key to select the digit to be chang...

  • Page 69

    67 recalling the upper and lower limit values (quick selection mode) the procedure below describes an easy way to recall the upper and lower limit values to be used for weighing. When the recalled upper and lower limit values are to be changed, refer to “storing the upper and lower limit values”. 1 ...

  • Page 70

    68 storing the tare value to store a new tare value: recall the stored tare value to be changed (“t01” to “t20”). Then, change the recalled tare value using the digital input mode or the weighing input mode, and store the new value. Notes when the re-zero key is pressed with nothing placed on the we...

  • Page 71

    69 weighing input mode in the weighing input mode, a sample tare container is placed on the pan to store the tare value. Notes pressing the cal key will interrupt the operation and the balance will return to the tare value confirmation mode (step 3 in “storing the tare value”). To go to the digital ...

  • Page 72

    70 recalling the tare value (quick selection mode) the procedure below describes an easy way to recall the tare value to be used for weighing. When the recalled tare value is to be changed, refer to “storing the tare value”. 1 set the “data memory (data)” parameter to “5”. 2 press the mode key to en...

  • Page 73

    71 11-7 data memory: quick selection mode the data memory has a quick selection mode, to recall data in memory quickly. Using the quick selection mode, the comparator settings or the tare value, whichever is selected in the function table, can be recalled, by a simple operation, using the mode key. ...

  • Page 74

    72 11-8 data memory: confirmation and storage mode the confirmation and storage mode can change and store the recalled data. Of the three types of memory data listed below, one may be selected in the function table, and is available for changes using this mode. Unit mass in the counting mode compara...

  • Page 75: 12. Programmable-Unit

    73 12. Programmable-unit this is a programmable unit conversion function. It multiplies the weighing data in kilograms by an arbitrary coefficient set in the function table and displays the result. The coefficient must be within the range between the minimum and maximum shown below. If the coefficie...

  • Page 76: 13. Density Measurement

    74 13. Density measurement the gp series balance is equipped with a density mode. It calculates the density of a solid using the weight of a sample in air and the weight in liquid. The density mode was not selected for use when the balance was shipped from the factory. To use the mode, change the fu...

  • Page 77

    75 (2) selecting a parameter to set. 1 press the sample key to select a parameter to set. Each pressing of the key switches the parameter. 2 press the re-zero key to enter the inputting mode of the parameter selected. (3) setting the parameter. A. Entering the weight of a sample in air. 1 confirm th...

  • Page 78

    76 c. Entering the density of a liquid. Two ways to set the density of a liquid are available in the function table, liquid density input (ldin): by entering the water temperature or by entering the density directly. • enter the water temperature (ldin 0) the water temperature currently set (unit: °...

  • Page 79

    77 14. I/o unit specifications (standard) 14-1 rs-232c//external contact input d-sub 25 pin numbers d-sub 25 pin assignments pin no. Signal name interface type direction description 1 fg - frame ground 2 rxd rs-232c input receive data 3 txd rs-232c output transmit data 4 rts rs-232c input ready to s...

  • Page 80

    78 external contact input by connecting pin 18 to pin 7, or pin 19 to pin 7, for more than 100 ms, the same operation as performed by pressing the re-zero key or the print key, will be performed. Use of example the external input connector (ax-hdb-25p/ctf) and the foot switch (ax-sw128) are sold sep...

  • Page 81

    79 15. Connection to peripheral equipment 15-1 connection to the ad-8121b printer set the following parameters to use the ad-8121b printer. Function setting description dout prt 0-3 selects a print mode. Dout ap-p 0-2 selects the polarity for the auto print mode. Dout ap-b 0-2 selects the auto print...

  • Page 82

    80 15-3 using windows communication tools (winct) when windows is used as an operating system in a personal computer, the provided winct software can be used to transmit the weighing data to the personal computer. The winct software has two communication methods: “rscom” and “rskey”. For details on ...

  • Page 83: 16. Commands

    81 16. Commands 16-1 command list note a command has a terminator added, that is specified using “5if crlf” of the function table, and is sent to the balance. Commands to query weighing data c cancels the s or sir command. Q requests the weighing data immediately. S requests the weighing data when s...

  • Page 84

    82 16-2 acknowledge code and error codes when the “serial interface function (5if)” parameter is set to “ercd 1”, the balance outputs code or error code to each command as follows: (06h) acknowledge in ascii code. When the balance receives a command to request data and can not process it, the balanc...

  • Page 85: 17. Maintenance

    83 17. Maintenance do not disassemble the balance. Contact the local a&d dealer if the balance needs service or repair. Use the original packing material for transportation. Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that is moistened with warm water a...

  • Page 86: 18-2 Error Codes

    84 sample and container has the sample absorbed or lost moisture due to the ambient conditions such as temperature and humidity? Has the temperature of the container been allowed to equalize to the ambient temperature? Refer to “3-2 during use”. Is the sample charged with static electricity? Refer t...

  • Page 87

    85 display error code description weighing pan error the weight value is too light. Confirm that the weighing pan is properly installed and calibrate the balance. Sample mass error the balance can not store the sample for the counting mode or for the percent mode because it is too light. Use a large...

  • Page 88: 18-3  Asking For Repair

    86 display error code description memory type error type of memory set in the function table and type of data stored are different. For details, refer to “11. Data memory”. Ec, e00 communications error a protocol error occurred in communications. Confirm the format, baud rate and parity. Ec, e01 und...

  • Page 89: 19. Specifications

    87 19. Specifications gp-12k gp-20k gp-30k gp-30ks gp-32k gp-32ks gp-40k weighing capacity 12 kg 21 kg 31 kg 31 kg 41 kg maximum display (see bottom for the description of ∗) 12.0084 kg 21.0084 kg 31.0084 kg 31.008 kg 6.1009 kg ∗ 41.0080 kg minimum weighing value (1 digit) 0.1 g 1 g/0.1 g 0.5 g repe...

  • Page 90

    88 gp-60k gp-60ks gp-61k gp-61ks gp-100k gp-100ks gp-102k weighing capacity 61 kg 61 kg 101 kg 101 kg maximum display (see bottom for ∗) 61.084 kg 61.0084 kg 101.084 kg 101.08kg 61.009 kg ∗ minimum weighing value (1 digit) 1 g 0.1 g 1 g 10 g/1 g repeatability (standard deviation) 0.7 g 0.2 g 1 g 5 g...

  • Page 91: 20. External Dimensions

    89 20. External dimensions standard type: gp-12k/20k/30k/32k/40k/60k/61k/100k/102k note the size in parentheses is for the gp-100k/102k unit: mm.

  • Page 92

    90 separate display type: gp-30ks/32ks/60ks/61ks/100ks note the size in parentheses is for gp-100ks unit: mm.

  • Page 93: 21. Options

    91 21. Options ad-8121b printer compact dot-matrix printer statistical function, clock and calendar function, interval print function, graphic print function, dump print mode 5 x 7 dots, 16 characters per line print paper (ax-pp143, 45 (w) x 50 (l) mm , ø65 mm) ac adapter or alkaline battery op-04: ...

  • Page 94: 22. Terms/index

    92 22. Terms/index terms stable value the stable weight data, indicated by the illuminated stabilization indicator. Environment ambient conditions such as vibration, drafts, temperature, static electricity and magnetic fields which affect the weighing operation. Store to save the weighing data, unit...

  • Page 95

    93 index keys and symbols on/off key ..................................................11 sample key ...........................................11, 12 mode key...............................................11, 13 cal key.........................................................11 print key..............

  • Page 96

    94 gross indicator ........................................................ 41 grounding terminal................................................ 7, 8 - h - hold .......................................................... 37, 40 hold function .............................................................

  • Page 97: Memo

    95 memo.

  • Page 98: Memo

    96 memo.