A&D HL-1000WP Instruction Manual

Other manuals for HL-1000WP: Instruction Manual
Manual is about: Digital Scale HL-WP Series

Summary of HL-1000WP

  • Page 1

    Digital scale hl-wp series hl-1000wp hl-300wp hl-3000wp wm+pd4000754.

  • Page 2: Safety Precautions

    1 safety precautions all safety messages are identified by the following, “warning” or “caution”, of ansi z535.4 (american national standard institute: product safety signs and labels). The meanings are as follows: warning a potentially hazardous situation which, if not avoided, could result in deat...

  • Page 3: Parts Description

    2 parts description weighing platform display on/off key g/lb model: hl-1000wp g/lb-oz model: hl-300wp/hl-3000wp turns the scale power on/off. Units key (g/lb and g/lb-oz models only) eighing units, changes w g or lb (oz). Re-zero key s display to zero or ubtracts weight of container. Set s spirit l...

  • Page 4: Preparation

    3 preparation installing the batteries / connecting the ac adapter installing the batteries caution when replacing batteries, be careful not to apply too much force to the weighing pan. An excessive force could damage the scale. Remove the battery cover and insert six batteries (r6p / lr6 / aa size)...

  • Page 5: Weighing

    4 weighing before weighing be aware of the automatic power-off function. If this unit is left on and the stable indicator is displayed, an automatic power-off function turns power off after approximately 5 minutes. To disable this function, press the [on/off] key while pressing the [re-zero] key. P-...

  • Page 6: Calibration

    5 calibration when to calibrate calibration may be required when the scale initially installed, if it is moved a substantial distance, or in accordance with local regulations. This is necessary because the weight of a mass in one location is not necessarily the same in another location. Also, with t...

  • Page 7: 1. Enter Calibration Mode.

    6 calibration by gravity compensation if you have calibration weights at the capacity of the scale, you can calibrate the scale with the weights instead of compensating for cceleration due to gravity. A the scale is calibrated at 9.798 m/sec 2 at the factory. If the gravity acceleration at your loca...

  • Page 8: Specifications

    7 specifications model hl-1000wp hl-300wp hl-3000wp 1000 g 300 g 3000 g capacity 2.2 lb 10.58 oz 6 lb 9.8 oz 0.5 g 0.1 g 1 g resolution 0.001 lb 0.01 oz 0.1 oz non-linearity ±1 g ±0.2 g ±2 g repeatability 0.5 g 0.1 g 1 g span drift ±0.015% / °c typ (5°c~35°c / 41°f~95°f) operating temp. -10°c~40°c /...

  • Page 9: World Map

    8 world map.

  • Page 10: A&d Engineering, Inc

    3-23-14 higashi-ikebukuro, toshima-ku, tokyo 170-0013 japan telephone: [81] (3) 5391-6132 fax: [81] (3) 5391-6148 a&d engineering, inc . 1555 mccandless drive, milpitas, ca. 95035 u.S.A. Telephone: [1] (408) 263-5333 fax: [1] (408) 263-0119 a&d instruments ltd. Unit 24/26 blacklands way, abingdon bu...