A&D TM-2430 Instruction Manual - Mode III Settings
Page 19
Mode III Settings
Set up procedure
Before you enter into mode III, read the procedure below. Also, refer to the
example on the next page for the setting procedure.
Each blocks starting time must match the previous blocks finish time.
The end of block 6 automatically equalizes to start time of the block 1.
If you enter the block 1 start time in any other block, these parameters are
saved and this sequence is finished.
When selecting 120 minutes for the current frequency, you must adjust the
start time of the next block so that the current block fits a multiple of 120
minutes. If you do not fit to the next start time, an error code is displayed.
The recorder displays
as 60 minutes and
as 120 minutes.
When you enter the sequence of mode III settings, the recorder initializes each start
time to the start time of block 1 and each frequency to "-" (not used).
To read the current settings, press the
START STOP
key in this sequence.
Display & key
Items
Sequence
number
Parameters (monitor)
Meaning
Initial value
01
0 ~ 23 o'clock
Start time of first block
02
- , 5, 10, 15, 20, 30, 60, 120 minutes Frequency of first block
03
0 ~ 23 o'clock
Start time of second block
04
- , 5, 10, 15, 20, 30, 60, 120 minutes Frequency of second block
05
0 ~ 23 o'clock
Start time of third block
06
- , 5, 10, 15, 20, 30, 60, 120 minutes Frequency of third block
07
0 ~ 23 o'clock
Start time of fourth block
08
- , 5, 10, 15, 20, 30, 60, 120 minutes Frequency of fourth block
09
0 ~ 23 o'clock
Start time of fifth block
10
- , 5, 10, 15, 20, 30, 60, 120 minutes Frequency of fifth block
11
0 ~ 23 o'clock
Start time of sixth block
12
- , 5, 10, 15, 20, 30, 60, 120 minutes Frequency of sixth block
13
0 ~ 23 o'clock
End of sixth block
The "-" means "not used".
AUTO
ON/OFF
START
STOP
Sequence number
Parameter
Sign of setting parameters
for automatic measurement
Selecting a
parameter
Selecting a sequence number