A&D TM-2655 Instruction Manual

Manual is about: Automatic Blood Pressure Monitor

Summary of TM-2655

  • Page 1

    Automatic blood pressure monitor 1wmpd4000206j 1303.

  • Page 2

    ? 2013 a&d company, limited all rights reserved. No part of this publication may be reproduced, transmitted, transcribed, or translated into any language in any form by any means without the written permission of a&d company, limited. The contents of this manual and the specifications of the instrum...

  • Page 3: Warning Definitions

    I warning definitions the warnings described in this manual have the following meanings: warning important information to alert you to a situation that might cause injury and/or damage to your property if instructions are not followed. Caution important information to alert you to a situation that m...

  • Page 4: Compliance

    Ii compliance compliance with the european directive 93/42/eec for medical products this device conforms to the following requirements: european directive 93/42/eec for medical products act; medical products act; european standards for electrical equipment en 60601-1 (general safety provisions), en ...

  • Page 5: Safety Precautions

    Iii safety precautions warning grounding to avoid electrical shock, connect the power cable to an electrical outlet having a ground terminal for grounding the device. Fuse to avoid a fire hazard, use only a fuse of the proper type, voltage and current rating as specified on the rear panel. Never byp...

  • Page 6: General Precautions

    Iv general precautions obey the following precautions for safe and correct usage. Precautions unique to the tm-2655/tm-2655p appear on the relevant pages in this manual. Read the manual thoroughly before use. When installing and storing the device install or store the device away from moisture. Do n...

  • Page 7: Contents

    1 contents warning definitions....................................................................................................... I compliance....................................................................................................................... Ii safety precautions................

  • Page 8

    2 9. Maintenance 9-1 replacing the arm cuff cover .................................................................................................................. 21 9-2 replacing the fuses ...............................................................................................................

  • Page 9: 1.Introduction

    3 1.Introduction 1-1 features the tm-2655/tm-2655p is a blood pressure monitor that measures systolic and diastolic blood pressure and pulse. The tm-2655 is a standard model; the tm-2655p is equipped with a printer unit. The features of the devices are as follows: compact design due to its compactne...

  • Page 10

    4 1-2 unpacking and inspection caution handle this device carefully at all times. Strong shock to the device may cause trouble during operation. Note save the packing material for later use. Unpack the tm-2655/tm-2655p carefully and verify that the following items are contained. Fig.1 items containe...

  • Page 11: 1-3 Specifications

    5 1-3 specifications table 1 specifications performance specifications general power supply 120vac 60 hz (for usa area) 230vac 50 hz (for europe area) (factory-preset) power consumption approx. 40va protection against electrical shock class i , type b display led emc en 60601-1-2 blood pressure meas...

  • Page 12: 2. Part Names

    6 2. Part names 2-1 main unit fig.2 tm-2655/tm-2655p whole view note the printer unit is available only for the tm-2655p. The human sensor area appears the same for both models, but it functions only for the tm-2655p. The illustration above is not to scale..

  • Page 13: 2-2 Rear Panel

    7 2-2 rear panel fig.3 tm-2655/tm-2655p rear panel (the illustration above is not to scale.) 2-3 symbols the symbols used with the tm-2655/tm-2655p have the following functions or meanings. Table 2 symbol descriptions symbols function/meaning starts and stops a measurement. Turns the device on. Turn...

  • Page 14

    8 what is an irregular heartbeat? The tm-2655p provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important t...

  • Page 15: 3.Installation

    9 3.Installation follow the procedure below to install the tm-2655/tm-2655p. 1. To ensure that the tm-2655/tm-2655p works properly, install the device in an environment where: the temperature range is from 10 °c to 40°c (50°f to 104°f). The humidity is less than 85% (non condensing). It is away from...

  • Page 16: 4. Operation

    10 4. Operation 1. Turn on the power switch located on the rear panel. When the power switch is turned on, all of the display symbols will appear for several seconds. Then; (tm-2655) “0” (zero) appears in the systolic pressure display, indicating that the device is ready for measurement. (tm-2655p) ...

  • Page 17

    11 8. When the measurement is complete, the air is automatically released from the cuff and the cuff will become loose. 9. The measurement results appear in the display. 10. (tm-2655p) the measurement results are output to the printer. Fig.9 measurement results note to stop the measurement at any ti...

  • Page 18: 5.Setting The Clock

    12 5.Setting the clock 5-1 clock adjustment mode the clock is set in the clock adjustment mode. The clock adjustment mode display and switches used are as shown below: fig.10 switches for setting the clock and the clock adjustment mode display table 3 switches for setting the clock select switch (1)...

  • Page 19

    13 8. Press the switch to display “14”. 9. Press the select switch to make the minutes section of the time display flash. 10. Press the switch to display “07”. 11. Press the select switch to return to the measurement mode. Note if no operation is performed for one minute, while setting the minutes s...

  • Page 20

    14 6. Installing the printer paper (this chapter is applicable only to the tm-2655p.) with the power turned on, install the printer paper as follows: 1. Cut the top end of the printer paper with scissors. If the end is not smooth, it may cause a paper jam or damage the printer head. 2. Press lightly...

  • Page 21

    15 6. Lower the lever to the down position (1) . Press the paper feed switch again, and confirm that the paper is cut automatically. Note lower the lever for cutting the paper. Lower the lever for printing the measurement data. 7. Remove the printer paper from the paper outlet, and close the cover. ...

  • Page 22: 7.  Setting The Functions

    16 7. Setting the functions 7-1 function setting mode the tm-2655/tm-2655p is provided with various functions, which allow the user to set the device to suit the user’s requirements. The functions are set in the function setting mode. The function setting mode display and switches used are as shown ...

  • Page 23

    17 7-2 description of the functions table 5 functions no. Function systolic display diastolic display description f01 f01 off not available. Off no printing 1 high-speed printing 2 4-line printing 3 4-line printing 4 graph printing f02 f02 5 table printing off no bit pattern printing f03 printer set...

  • Page 24: 7-3 Printing Samples

    18 7-3 printing samples high-speed printing graph printing 4-line printing table printing bit pattern printing fig.13 printing samples note bit pattern printing is available only for the tm-2655p. With this format, some items such as the company name can be printed. For details, contact your nearest...

  • Page 25

    19 8. Communication specifications the tm-2655/tm-2655p is equipped with two rs-232c channels. Various settings for each channel are available in the function setting mode. Refer to “7-2 description of the functions” for details. 8-1 channel 1 : miniature 8-pin din caution the personal computer and ...

  • Page 26

    20 8-2 channel 2: d-sub 9-pin caution the personal computer and medical equipment connected to the device are not allowed to be in the patient area. The personal computer used must conform to iec60950 specifications table 7 channel 2 specifications standard conforms to eia rs-232c transmission async...

  • Page 27: 9. Maintenance

    21 9. Maintenance 9-1 replacing the arm cuff cover replace the arm cuff cover as follows: 1. Locate the cover at the bottom of the arm insertion section. Slide the cover downward to open it. A screw is exposed. Using the screwdriver, loosen the screw and remove the front frame. 2. Remove the four sc...

  • Page 28: 9-2  Replacing The Fuses

    22 9-2 replacing the fuses replace the fuses as follows: 1. Remove the caps of both fuse holders located on the rear panel. 2. Replace both fuses with new ones. 3. Re-install the fuse holder caps. Fig.19 fuse holders note use only fuses of the proper type, voltage and current rating as specified on ...

  • Page 29: 9-4 Error Codes

    23 resetting the counter hold down the counter switch for 4 seconds or more. The counter data will be reset to zero. Printing the counter graph press and hold the counter switch, press the paper feed switch to display the current measurement count. Total : the total measurement count of the product....

  • Page 30

    24 housing case clean the case gently using a soft cloth moistened with water and mild detergent or alcohol. Display panel clean the display gently so as not to scratch the panel surface. Use a soft cloth moistened with water. Arm cuff cover when the arm cuff cover becomes dirty or it is worn out, r...

  • Page 31: 10. Troubleshooting

    25 10. Troubleshooting if the tm-2655/tm-2655p does not function properly or an error code appears, try the following corrective actions. Table 10 troubleshooting problem check this corrective action is the power cable connected properly? Connect the power cable properly. Nothing appears in the disp...

  • Page 32: 11. External Dimensions

    26 11. External dimensions fig.20 external dimensions note the illustration above indicates the tm-2655p with the printer unit. The dimensions are the same for the tm-2655 and tm-2655p..

  • Page 33: Appendix: Emc Information

    27 appendix: emc information medical electrical equipment needs special precautions regarding emc and needs to be installed and put into service according to the emc information provided in the following. Portable and mobile rf communication equipment (e.G. Cell phones) can affect medical electrical...

  • Page 34

    28 guidance and manufacturer’s declaration – electromagnetic immunity the a&d unit is intended for use in the electromagnetic environment specified below. The customer or the user of the a&d unit should assure that it is used in such an environment. Immunity test iec 60601 test level compliance leve...

  • Page 35

    29 guidance and manufacturer’s declaration – electromagnetic immunity the a&d unit is intended for use in the electromagnetic environment specified below. The customer or the user of the a&d unit should assure that it is used in such an environment. Immunity test iec 60601 test level compliance leve...

  • Page 36

    30 memo.

  • Page 37

    31 memo.

  • Page 38

    32 memo.

  • Page 39

    13.

  • Page 40

    1-243 asahi , kitamoto-shi, saitama 364-8585, japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d instruments limited unit 24/26 blacklands way, abingdon business park, abingdon, oxfordshire ox14 1dy united kingdom telephone: [44] (1235) 550420 fax: [44] (1235) 550485 a&d engineering, in...