A&D UA-621 Instruction Manual

Other manuals for UA-621: Operation Manual, Operation Manual
Manual is about: Upper Arm Blood Pressure Monitor

Summary of UA-621

  • Page 1

    Digital blood pressure monitor model ua-621 instruction manual 1wmpd4001898d.

  • Page 2: Contents

    1 contents dear customers ............................................................................................ 1 preliminary remarks...................................................................................... 2 precautions...............................................................

  • Page 3: Preliminary Remarks

    2 preliminary remarks this device conforms to the european directive 93/42 eec for medical products. This is made evident by the mark of conformity. (0366: the reference number to the involved notified body). The device is designed for use on adults only, not newborns or infants. Environment for use...

  • Page 4: Parts Identification

    3 do not apply the cuff on an arm receiving an intravenous drip or blood transfusion. It may cause injury or accidents. Do not use the dvice where flammable gases such as anesthetic gases are present. It may cause an explosion. Do not use the device in highly concentrated oxygen environments, such a...

  • Page 5: Symbols

    4 symbols symbols function / meaning recommended action standby and turn the device on. Battery installation guide direct current type bf: device, cuff and tubing are designed to provide special protection against electrical shocks. The indicator that appears while measurement is in progress. It bli...

  • Page 6: Operation Mode

    5 symbols function / meaning recommended action weee label refer to instruction manual/booklet sn serial number operation mode 1. Normal measurement with storing data ( refer to page 9) press the start button. Blood pressure is measured and data is stored within the device. The device can store the ...

  • Page 7: Using The Monitor

    6 using the monitor installing / changing the batteries 1. Remove the battery cover. 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+) and (-) are correct. 3. Replace the battery cover. Use only r6p, lr6 or aa batteries. ...

  • Page 8: Using The Monitor

    7 using the monitor selecting the correct cuff using the correct cuff size is important for an accurate reading. If the cuff is not the proper size, the reading may yield an incorrect blood pressure value. The arm size is printed on each cuff. The index and proper fit range, on the cuff, tell you if...

  • Page 9: Using The Monitor

    8 using the monitor applying the arm cuff 1. Wrap the cuff around the upper arm, about 1 to 2 cm above the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse and result in a measurement error. 2. Constriction of the upper arm, caused by rolling up a shirts...

  • Page 10: Measurements

    9 measurements model ua-621 is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. If your systolic pressure is expected to exceed 230 mmhg, read "measurement with the desired systolic pressure" on the next page. Normal measurement 1. Place the cuff on th...

  • Page 11: Measurements

    10 measurements measurement with the desired systolic pressure if your systolic pressure is expected to exceed 230 mmhg, use this procedure. 1. Place the cuff on the arm (preferably the left arm). 2. Press and hold the start button to display the last data number. 3. Press and hold the start button ...

  • Page 12: About Memory

    11 about memory the device automatically stores up to thirty blood pressure and pulse measurements in memory. Data stored in memory are assigned a data number in the order of the newest to the oldest. The oldest data displays as "n01". The m symbol in the upper left corner of the display indicates t...

  • Page 13: Example:

    12 example: moderate hypertension mild hypertension high normal what is an irregular heartbeat model ua-621 provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartb...

  • Page 14: About Blood Pressure

    13 about blood pressure what is blood pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmhg). One's natu...

  • Page 15: Troubleshooting

    14 take measurements at the same time every day using the procedure described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note date and time when recording your blood pressure. Consult your doctor to interpre...

  • Page 16: Maintenance

    15 maintenance do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the a&d service group. The a&d service group will provide tech...

  • Page 17

    16 useful life device: 5 years (when used six times a day) cuff: 2 years (when used six times a day) accessory ac adapter the adapter is to connect the device to a power source at home. Tb-233 please contact your local a&d dealer for purchasing. The ac adapter is required to be inspected or replaced...

  • Page 18

    A&d company, limited 1-243 asahi, kitamoto-shi, saitama 364-8585 japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d instruments ltd. Unit 24/26 blacklands way, abingdon business park, abingdon, oxfordshire ox14 1dy united kingdom telephone: [44] (1235) 550420 fax: [44] (1235) 550485 a&d...