    Instruction guide – model ua-704 mini manual inflate blood pressure monitor latex free.

  • Page 2: Than Six Months.

    Please read this important information before using your monitor. ☞ please remember that only a medical practitioner is qualified to interpret your blood pressure measurements. Use of this device should not replace regular medical examinations. ☞ have your physician review your procedure for using t...

  • Page 3

    Tm congratulations on purchasing a state-of-the-art lifesource blood pressure monitor—one of the most technologically advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. Physicians agree that daily self-monitoring ...

  • Page 4: E-1

    E-1 what display symbols mean . . . . . . . . . . . . . . . . . . . E-2 monitor components . . . . . . . . . . . . . . . . . . . . . . . . E-3 how ua-704 works . . . . . . . . . . . . . . . . . . . . . . . . . . E-4 before you start . . . . . . . . . . . . . . . . . . . . . . . . . . . . E-4 selecti...

  • Page 5: E-2

    E-2 display symbol recommended action remain still and refrain from talking; measurement in progress. Press exhaust button to release air. Inflate cuff 30 to 40 mmhg above your normal systolic pressure. Take measurement again and consult with your physician. Replace the battery with new one, when th...

  • Page 6: E-3

    Mmhg mmhg(/min) ua-704 digit al blood pressure monit or measuring presse exhaust infla te irrgu lar hea rt bea t batter y low sys. D ia.(pu l.) e-3 monitor components ? Cuff battery compartment battery cover aa battery start button air tube systolic pressure display exhaust valve inflator bulb in ca...

  • Page 7: E-4

    E-4 you must install 1 type aa (1.5 volt) battery and attach the cuff to the monitor before using it. To install battery (or replace it if the “low battery” symbol appears on display), proceed as follows: 1. Remove battery cover by gently pushing down on handle and pulling cover off monitor. 2. Plac...

  • Page 8: E-5

    E-5 using the correct cuff size is important for an accurate reading. A cuff that is too large will produce a reading that is lower than the correct blood pressure; a cuff that is too small yields a measurement that is higher. With your arm hanging at the side of your body, measure the circumference...

  • Page 9: E-6

    E-6 taking your blood pressure tips for blood pressure monitoring: ☞ relax for about 5 to 10 minutes before measurement. ☞ remove constricting clothing and place cuff on bare arm. ☞ unless your physician recommends otherwise, use left arm to measure pressure. Now you are ready. Follow these simple s...

  • Page 10: E-7

    E-7 ☞ important: measure pressure at the same time each day. 4. Press the start button 5. Watch for all display symbols to appear briefly, followed by a flashing “0”, indicating that monitor is ready for measurement. 6. If the mark is displayed, the cuff has some air trapped in it. Press the exhaust...

  • Page 11: E-8

    E-8 taking your blood pressure 9. When the correct pressurization has been reached, the automatic exhaust mechanism will gradually reduce the pressure in the cuff. Sit quietly during measurement. The symbol will blink with your pulse beat. Note: exhaust rate cannot be adjusted during measurement. If...

  • Page 12: E-9

    E-9 taking your blood pressure 12. Remove cuff and make a note of your blood pressure and pulse rate on the chart (see page e-13), indicating date and time of measurement. Additional logbook charts can be downloaded and printed for free at www.Lifesourceonline.Com or www.Autocontrol.Com. We advise t...

  • Page 13: E-10

    E-10 what is blood pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts; diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmhg). What affects blood pressure? ...

  • Page 14: E-11

    E-11 assessing high blood pressure the following standards for assessing high blood pressure (without regard to age) have been established as a guideline. Bp classification systolic (mmhg) diastolic (mmhg) normal and prehypertension 120-139 or 80-89 stage 1 hypertension 140-159 or 90-99 stage 2 hype...

  • Page 15: E-12

    E-12 about blood pressure ❤ why measure blood pressure at home? It is now well known that, for many individuals, blood pressure readings taken in a doctor’s office or hospital setting might be elevated as a result of apprehension and anxiety. This response is commonly called “white coat hypertension...

  • Page 16: E-13

    Tm in canada (toll-free): 1-800-463-5414 e-13 ❤.

  • Page 17: Troubleshooting

    Troubleshooting ✓ problem display is blank, even when the power is turned on. The device will not measure. Readings are too high or too low. Other probable cause battery is drained. Battery terminals are not in the correct position. The cuff is not fastened properly. You moved your arm or body durin...

  • Page 18: Specifications

    Specifications ✓ in canada (toll-free): 1-800-463-5414 e-15 ✓ model . . . . . . . . . . . . . . . . . . .Ua-704 type . . . . . . . . . . . . . . . . . . . . .Oscillometric display . . . . . . . . . . . . . . . . . .Digital, 10-mm character height pressure/pulse displayed alternately measurement rang...

  • Page 19: E-16

    E-16 contact information our products are designed and manufactured using the latest scientific and technological methods, and offer accurate, easy to use, home monitoring and treatment options. Our full product line includes: aneroid blood pressure kits blood pressure cuffs digital blood pressure m...

  • Page 20

    Contact information a division of a&d engineering, inc. 1555 mccandless drive milpitas, ca 95035 www.Lifesourceonline.Com lifesource is a trademark of a&d medical. ©2003 a&d medical. All rights reserved. Specifications subject to change without notice. I-man:704 printed in china 02/04 important! If ...