A&D UA-767 Plus BT-C Instruction Manual

Manual is about: Digital Blood Pressure Monitor

Summary of UA-767 Plus BT-C

  • Page 1

    Digital blood pressure monitor model ua-767 bt-c instruction manual 1wmpd4001966a.

  • Page 3: Contents

    1 contents dear customers.................................................................... 2 preliminary remarks ............................................................. 2 precautions........................................................................... 2 parts identification..............

  • Page 4: Dear Customers

    2 dear customers congratulations on purchasing a state-of-the-art a&d blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read through this manual caref...

  • Page 5: Parts Identification

    3 parts identification battery cover 1.5v batteries (r6p or aa) battery compartment display air connector plug air socket arm cuff dc jack start button air hose index mark proper fit range ( 8.7"-12.6 " ) part of display pressure bar indicator memory systolic pressure diastolic pressure pulse rate b...

  • Page 6: Symbols

    4 symbols symbols function / meaning recommended action standby and power on. Battery installation guide direct current sn serial number date of manufacture type bf: device, cuff and tubing are designed to provide special protection against electrical shocks. The indicator that appears while measure...

  • Page 7: Using The Monitor

    5 using the monitor installing / changing the batteries 1. Slide the battery cover up to open it. 2. Remove the used batteries and insert new batteries into the battery compartment as shown, taking care that the polarities (+) and (-) are correct. 3. Slide the battery cover down to close. Use only r...

  • Page 8: Using The Monitor

    6 using the monitor pairing the pbt-c series device with a continua-compliant manager model ua-767 bt-c is a pbt-c series device. Pbt-c series devices are designed to be connected only with continua managers. To pair the device with a manager, please follow these steps: 1. Remove the four batteries ...

  • Page 9: Using The Monitor

    7 using the monitor attaching the arm cuff 1. Wrap the cuff around the upper arm, about 2 - 3 cm above the elbow, as shown. Place the cuff directly against the skin, as clothing may cause a faint pulse, and result in a measurement error. 2. Constriction of the upper arm, caused by rolling up a shirt...

  • Page 10: Measurements

    8 at heart level the last data at initial display starts inflation pressurizing measurement in progress systolic pressure diastolic pressure pulse rate exhausts remaining air automatically start button measurements model ua-767 bt-c is designed to detect the pulse and to inflate the cuff to a systol...

  • Page 11: Measurements

    9 measurements measurement with the desired systolic pressure if your systolic pressure is expected to exceed 230 mmhg or you use the optional small cuff, use this procedure. 1. Place the cuff on the arm (preferably the left arm). 2. Press and hold the start button until a number about 30 to 40 mmhg...

  • Page 12: Pressure Bar Indicator

    10 what is an irregular heartbeat model ua-767 bt-c blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure me...

  • Page 13: About Blood Pressure

    11 reference material: journal of hypertension 1999, vol 17 no.2 mmhg 110 105 100 95 90 85 80 optimal 120 130 140 150 160 170 180 systolic blood pressure mmhg normal high-normal grade 1 hypertension (mild) grade 2 hypertension (moderate) grade 3 hypertension (severe) dia stolic bl ood pre ssure abou...

  • Page 14: Troubleshooting

    12 typical fluctuation within a day (measured every five minutes) pm am time dia stolic bl ood pre ssure systolic diastolic sleep described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note date and time when ...

  • Page 15: Maintenance

    13 maintenance do not open the device. It uses delicate electrical components and an intricate air unit that could be damaged. If you cannot fix the problem using the troubleshooting instructions, request service from your dealer or from the a&d service group. The a&d service group will provide tech...

  • Page 16

    14 medical electrical equipment needs special precautions regarding emc and needs to be installed and put into service according to the emc information provided in the following. Portable and mobile rf communication equipment (e.G. Cell phones) can affect medical electrical equipment. The use of acc...

  • Page 17

    15 guidance and manufacturer’s declaration – electromagnetic immunity the a&d unit is intended for use in the electromagnetic environment specified below. The customer or the user of the a&d unit should assure that it is used in such an environment. Immunity test iec 60601 test level compliance leve...

  • Page 18

    16 guidance and manufacturer’s declaration – electromagnetic immunity the a&d unit is intended for use in the electromagnetic environment specified below. The customer or the user of the a&d unit should assure that it is used in such an environment. Immunity test iec 60601 test level compliance leve...

  • Page 20

    A&d company, limited 1-243 asahi, kitamoto-shi, saitama 364-8585 japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d instruments ltd. Unit 24/26 blacklands way, abingdon business park, abingdon, oxfordshire ox14 1dy united kingdom telephone: [44] (1235) 550420 fax: [44] (1235) 550485 a&d...