A&D UA-774 Instruction Manual - Using The Monitor
English 5
Using The Monitor
Installing / Changing The Batteries
1. Slide the battery cover up to open it.
2. Remove the used batteries and insert
new batteries into the battery
compartment as shown, taking care that
the polarities (+) and (-) are correct.
3. Slide the battery cover down to close it.
Use only R6P, LR6 or AA batteries.
CAUTION
Insert the batteries as shown in the battery compartment. If not, the device will not work.
When
(LOW BATTERY mark) blinks in the display, replace all batteries with
new ones. Do not mix old and new batteries. It may shorten the battery life, or
cause the device to malfunction.
(LOW BATTERY mark) does not appear when the batteries are drained.
Battery life varies with the ambient temperature and may be shorter at low
temperatures.
Use the specified batteries only. The batteries provided with the device are for
testing monitor performance and may have a limited life.
Remove the batteries if the device is not to be used for a long time.
The batteries may leak and cause a malfunction.
The stored data are cleared when the batteries are removed.
Connecting The Air Hose
Insert the air connector plug into
the air socket firmly.
Selecting The Correct Cuff
Using the correct cuff size is important for an accurate reading. If the cuff is not
the proper size, the reading may yield an incorrect blood pressure value.
The arm size is printed on each cuff.
The index and proper fit range, on the cuff, tell you if you are attaching the
correct cuff. (Refer to "Attaching The Arm Cuff" on the next page)
If the index points outside of the range, contact your local dealer to purchase a
replacement cuff.
Arm Size
Recommended Cuff Size
Order No.
32 cm to 42 cm
Large cuff
UA-CUFBLALB
22 cm to 32 cm
Medium cuff
UA-CUFCAULB, UA-CUFDAULB
-
Small
cuff
-
Arm size: The circumference at the biceps.
Note: This UA-774 monitor is NOT designed for using a small cuff.
Step 1
Step 3
Step 2
Air Connector Plug
Air Socket
Air Hose