Instruction manual mode d'emploi manual instructivo manuale d’uso bedienungsanleitung english français español italiano deutsch 1wmpd4002731 precision health scale uc-355pbt-ci.
1wmpd4002731 startup guide ............................................................................................. 1 safety precautions..................................................................................... 2 wireless communication function ..........................................
English 1 startup guide kg ● receivers can also be unpaired. (p.11) set the unit (p.8) confirm your weight ※ data is transmitted. Data is stored in the scale if there is no receiver. (p.11) step off the scale confirm that step and off are alternately displayed step on the scale pair with a receiver ...
English 2 safety precautions the following items contain information to prevent the risk of harm or injury to yourself or other people. Please read these items carefully and be sure to use this product correctly. Danger this symbol indicates "an immediate danger which could result in death or seriou...
English 3 wireless communication function this scale uses the bluetooth wireless to communicate with a receiver device. At the end of measurement, the measurement data is automatically sent from this scale to the receiver device. Bluetooth is communication technology that enables devices to communic...
English 4 parts names scale display section display symbols display section symbol name description ○ ready/stable symbol this symbol is displayed when the measurement display is stable. Battery indicator ↓ ↓ low battery indicator the normal battery power indicator with charge remaining. If this s...
English 5 installing (or replacing) the batteries 1. Open the battery compartment cover on the underside of the scale. 2. Install four new aa size alkaline batteries with the positive (+) and negative (-) terminals properly aligned with the corresponding symbols in the battery compartment. 3. Close ...
English 6 registering with a receiver (pairing) this scale needs to be registered with a receiver in order to use bluetooth communication with the receiver. The process of registering this scale with a receiver is known as pairing. When pairing is completed, measurement data from this scale is autom...
English 7 skip registering with a receiver (pairing) 1. Install four aa size alkaline batteries into this scale. for details, see "installing (or replacing) the batteries (p.5)". 2. Press the switch when "pr" is displayed. press the switch to turn the power off. ※ you can also measure your weigh...
English 8 setting the unit 1. Confirm that the scale is turned off, and press and hold the switch. When the switch is continuously pressed, "unit" and "upr" are alternately displayed. ≪♪beep≫ 2. Release the switch when "unit" is displayed. the display changes from "unit" to display the unit that i...
English 9 measuring your weight step on weighing 1. Confirm that the scale is turned off, and step on the scale gently. measurement starts when you step on the scale. A progress bar is displayed during measurement. Stand up straight and remain as still as possible during measurement. ≪♪beep≫ 2. St...
English 10 measuring your weight on a carpet place the scale on a solid level surface, such as tile, wood, or laminate flooring. If the scale must be used on a carpet, be sure to use the supplied tray. 1. Place the supplied tray on the carpet. 2. Place the scale onto the tray in line with the illust...
English 11 other functions unpairing 1. Confirm that the scale is turned off, and press and hold the switch. when the switch is continuously pressed, "unit" and "upr" are alternately displayed. ≪♪beep≫ 2. Release the switch when "upr" is displayed. ≪♪beep≫ 3. Confirm that "end" is displayed. the...
English 12 scale communication this scale is equipped with a continua certified bluetooth wireless communication function. Measurement data can be transmitted by connecting to a bluetooth device that is compatible with continua connection specifications. Devices that are continua certified carry the...
English 13 specifications model uc-355pbt-ci maximum capacity 250 kg / 550 lb minimum display 0.1 kg units from 0 to 100.0 kg 0.2 kg units from 100.0 to 250.0 kg 0.2 lb units from 0 to 220.0 lb 0.5 lb units from 220.0 to 550.0 lb communication method bluetooth ver.2.1 class 1 ssp hdp compatible sett...
English 14 before requesting service symptoms corrective action e is displayed exceeds the maximum capacity (250 kg / 550 lb). E-0 is displayed exceeds the maximum capacity (250 kg / 550 lb). Err is displayed exceeds the maximum capacity (250 kg / 550 lb). E-10 is displayed pairing or unpairing has ...
1-243 asahi, kitamoto-shi, saitama 364-8585 japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d instruments limited unit 24/26 blacklands way, abingdon business park, abingdon, oxfordshire ox14 1dy united kingdom telephone: [44] (1235) 550420 fax: [44] (1235) 550485 a&d engineering, inc....