A&D UM-101 Instruction Manual

Manual is about: Digital Sphygmomanometer

Summary of UM-101

  • Page 1

    Instruction manual manuel d’instructions manual de instrucciones manuale di istruzioni bedienungsanleitung 使用手册 wm+pd4001301.

  • Page 2

    English 1 um-101 digital sphygmomanometer instruction manual contents 1. Dear customers ................................................................................2 2. Preliminary remarks .........................................................................3 3. Notes for proper use............

  • Page 3: Features

    English 2 1. Dear customers congratulations on purchasing a state-of-the-art a&d sphygmomanometer, one of the most advanced monitors available on the market today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read throug...

  • Page 4: Compliance

    English 3 2. Preliminary remarks compliance compliance with european directive 93/42 eec for medical products the device conforms to the following requirements: european directive 93/42 eec for medical products; medical products act; european standards for electrical medical equipment en 6060-1 (gen...

  • Page 5: Training

    English 4 training the healthcare provider should stop the examination if there is an abnormality, such as a patient feeling excessive arm pain, and remove the cuff to protect the patient. Repair do not attempt to open the device. Contact your nearest a&d authorized representative and they will repa...

  • Page 6: During Use

    English 5 during use this device should be used by trained professionals. Stop the examination immediately, if the patient feels pain during a measurement or if the device does not work properly. Stop using the device if you notice any abnormalities (for example; liquid inside the device) and reques...

  • Page 7: Main Body and Accessories

    English 6 4. Parts identification main body and accessories function on/off button turns on or off the device. Marker button during measurement............ After turned off ...................... Puts a marker at a pressure value indicates a lifetime counter numerical display during measurement........

  • Page 8

    English 7 5. Symbols symbols function / meaning recommended action standby and turn the device on. Marker pressure value holding and lifetime counter. Battery installation guide sn serial number date of manufacture type bf: device, cuff and tubing are designed to provide special protection against e...

  • Page 9: Caution

    English 8 6. Using the monitor installing / changing the batteries 1. Remove the battery cover. 2. Insert a new set of batteries into the battery compartment as shown. Make sure the polarities (+) and (-) are correct. Use only lr6, aa or equivalent batteries. 3. Close the battery cover. Caution inse...

  • Page 10: Attaching The Arm Cuff

    English 9 attaching the arm cuff 1. Wrap the cuff around the upper arm, about 2 ~ 3 cm above the elbow as shown on right. Place the cuff directly against the skin, as clothing may cause a faint pulse and could result in a measurement error. 2. Constriction of the upper arm, caused by rolling up a sh...

  • Page 11: Lifetime Counter

    English 10 4. When inflation is complete. Turn the exhaust valve screw to release air slowly. Measure the systolic pressure and the diastolic pressure by stethoscopy. 5. The pulse rate is shown on the numerical display when the measurement is complete, and meets the following conditions. When you pr...

  • Page 12

    English 11 9. Troubleshooting problem possible reason recommended action batteries are empty. Replace all batteries with new ones. Nothing appears in the display, even when the power is turned on. Battery polarities are not in the correct position. Reinstall the batteries with their negative and pos...

  • Page 13

    English 12 11. Technical data model um-101 measurement method stethoscopy with stethoscope measurement range numerical display pressure: 0 ~ 300 mmhg pulse: 30 ~ 200 beats / minute pressure bar display pressure: 20 ~ 280 mmhg measurement accuracy numerical display pressure: ±3 mmhg pulse: ±5 % press...

  • Page 14: A&d Co.,ltd.

    A&d co.,ltd. 1-243 asahi, kitamoto-shi, saitama 364-8585 japan telephone: [81] (48) 593-1111 fax: [81] (48) 593-1119 a&d instruments ltd. Unit 24/26 blacklands way, abingdon business park, abingdon, oxon ox14 1dy united kingdom telephone: [44] (1235) 550420 fax: [44] (1235) 550485 a&d mercury pty. L...