A.W. Sperry Instruments DSA-2002P Operating Instructions Manual - Ⅲ
― 6 ―
●Operating System
Dual Integration
●Display
Liquid crystal display with a maximum
count of 4000
●Low Battery Warning
"BATT" symbol is displayed on the digital
display.
●Overrange Indication
"OL" is displayed where input exceeds
the upper limit of a range
●Response Time
Approx. 2 seconds
●Sleep function
Automatically powered down in about 10
minutes after the last switch operation
●Data Hold
Available in all ranges provided the peak
measurement mode is deactivated.
●Location for use
Indoor use, Altitude up to 2000m
●Storage Temperature
ー20〜60℃, relative humidity up to 85%
and Humidity
without condensation
●Operating Temperature
0
〜40℃, relative humidity up to 85%
and Humidity
without condensation
●Conductor Size
Approx. 54.5
㎜ diameter max.
●Overload Protection
2400A AC for 10sec
1200V AC/DC for 10sec
600V AC for 10sec
●Withstand Voltage
5550V AC for 1 minute between electrical
circuit and housing cases or metal parts
of jaws
●Insulation Resistance
10M
Ω or greater at 1000V between
electrical circuit and housing cases or
metal parts of jaws
●Safety Standard
IEC 61010-1:Measurement CAT.
Ⅲ
600VAC,DC / CAT.
Ⅱ750VAC, 1000VDC,
pollution degree2.
●Dimensions
247
(L)
×105
(W)
×49
(D)
mm
●Weight
Approx. 470g(battery included)
●Power Source
Two R6P(DC1.5V) batteries or equivalent
●Current Consumption
Approx. 5mA max. (Approx. 20
μA in the
sleep mode)
●Accessories
Test leads M-7107
Two R6P batteries
Instruction manual
Recorder Output Plug M-8201
Carrying case M-9094
●Optional Accessories
Output Probe M-7014
Recorder M-5100A, etc.