AA Portable Power Corp Precision AC- DC Power Supply User manual

Manual is about: Dual Channel 0-30V , 5A max or 0-60V single channe

Summary of Precision AC- DC Power Supply

  • Page 1

    User manual of precision ac- dc power supply: dual channel 0-30v , 5a max or 0-60v single channel aa portable power corp (http://www.Batteryspace.Com) address: 860 s, 19 th st, unit a, richmond, ca, 94804 tel: 510-525-2328 fax: 510-439-2808 email: sales@batteryspace.Com prepared & approved by max (2...

  • Page 2

    1. Introduction adjustable dc power supply are designed to be used in applications such as powering operational amplifier, push pull stages, logic circuit and definition systems where plus and minus voltages are required to track with an insignificant error, and in any application where three indepe...

  • Page 3

    Tracking switches. B. Because both voltage and current of the slave supply track the master supply, the maximum current and voltage are set using the master controls. Using the master supply output jacks, follow the instructions for "setting current limit" (4-2 section). Remember that the actual cur...

  • Page 4

    If the chassis or common of the equipment being powered is separate from both positive and negative polarity power inputs. The output of the slave (negative) supply is tracking the output of the master (positive) supply. The configuration is shown in power supply mast er sl ave l oad gnd fig .4-4 po...

  • Page 5

    Rating model specifications a b c d output voltage 2h 0~30v 2h 0~30v 2h 0~30v 2h 0~30v output current 2h 0~3a 2h 0~5a 2h 0~3a 2h 0~5a output fixed 5v,3a 5v,3a 5v,3a 5v,3a output on/off v/a initializes x x √ √ dimensions: 245(w)h140(h)h345(d)mm weight: approx.8kg 2-2 operation mode (1). Independent: ...

  • Page 6

    When the series tracking mode of operation is selected, the positive (red) terminal of the slave supply output is internally connected to the negative (black) terminal of the master supply. In the series tracking mode, the maximum output voltage of both master and slave supplies can be simultaneousl...

  • Page 7

    Supply is in the independent operating mode. B. Adjust "voltage" control and "current" control to the desired output voltage and current. C. Turn off the power supply and the equipment to be powered during hook-up. D. Connect the positive polarity of the device being powered to the red (+) terminal ...

  • Page 8

    Between chassis and ac cord 300mw or above (dc 500v) 3. Panel controls and indicators v 1 c.V . Laboratory dc power supply curr ent pow er 22 13 on 29 8 7 28 output o n/ of f + gnd - of f 12 21 19 20 c .V. C.C. Par. Slave vo ltage + - gnd 17 18 15 14 16 inde p. Seri es pa rall el track ing c.C. Mast...

  • Page 9

    Fig.4-1 con st an t volt ag e/co ns tant character is tic. Out pu t cu rren t co ns tant ran ge vo ltag e co ns tant current ran ge ou tp ut vo ltag e cro sso ver po in t vo m ax io max (3). Output voltage overshoot maybe voltage between output terminals exceeds the present value when the power is t...

  • Page 10

    Mode, a regulated output voltage is provided. The output voltage remains constant as the load increases up until the point where the preset current limit is reached. At that point, the output current becomes constant and the output voltage drops in proportion to further increases in load. The point ...

  • Page 11

    In this mode, maximum voltage of both supplies is set using the master voltage controls (voltage at output terminals of the slave supply tracks the voltage at the output terminals of the master supply). Also, in this mode of operation the positive terminal (red) of the slave supply is connected to t...

  • Page 12

    Fig .3-2 rear panel.