AAT SOLO+ User Manual - page 75
75
GB
Malfunction, control unit
Code
Malfunction
Troubleshooting
01
Keypad
Contact the manufacturer or authorized
personnel!
02
Joystick
Contact the manufacturer or authorized
personnel!
03
Serial number
Contact the manufacturer or authorized
personnel!
04
Parameter data
Turn off the SOLO, wait for 10 sec., turn
on the SOLO.
Malfunction still present?
Yes: Contact the manufacturer!
0
5
Software sequence Turn off the SOLO, wait for 10 sec., turn
on the SOLO.
Malfunction still present?
Yes: Contact the manufacturer!
0
6
Wrong command Turn off the SOLO, wait for 10 sec., turn
on the SOLO.
Malfunction still present?
Yes: Contact the manufacturer!
0
7
No contact to the monitoring
of the getting up function
Contact the manufacturer or authorized
personnel!