Pt series pressure transmitters user manual abus technologies inc..
Abus technologies inc. 2 pt warning v this manual should be passed on to the end user. V the contents of this manual are subject to change without prior notice. V all rights reserved. V abus gives no warranty of any kind with regard to this manual, including, but not limited to, fitness for a partic...
Abus technologies inc. 3 pt . Catalogue contents page no. 1. Introduction applications 4 4 2. Presentation 1. Features 2. Technical parameters 4 4 5 3. Dimensions 5 4. Ordering details 6 5. Connections output characteristics 7 8 6. Installation recommendation 9 7 7. Operation 1. Zero and span adjust...
Abus technologies inc. 4 pt 1. Introduction pt series of p ressure transmitters are precision engineered to fit most industrial pressure measurement applications. The compact, rugged design make these instruments suitable for applications including hydraulics and pneumatics, vacuum, test equipment, ...
Abus technologies inc. 5 pt 2.2 technical parameters accuracy: ±0.25%fs (bfsl) ±0.5%fs (limit point calibration) ranges: absolute: 0…1 to 0…50 bar, 0…15 to 0…750 psi vacuum: (-1)… bar, (-15)…0 psi gauge: 0…100mb to 0…1000 bar, 0…1.5 to 0…15000 psi output: 0...5 v, 0...10 v, 0.5...4.5 v, 4...20 ma (c...
Abus technologies inc. 6 pt 4. Ordering details type description type pt10 absolute pressure transmitter 0 ~ 25 bar pt20 gauge/compound pressure transmitter -1 ~ 0 ~ 1800 bar pt30 vacuum pressure transmitter -760 ~ 0 mmhg range __ specify range code output d 4 ~ 20 ma v 0 ~ 10 v a others, specify pr...
Abus technologies inc. 7 pt 5. Connections wire signal output excitation 2-wire 4 ~ 20ma 12 ~ 36 vdc 3-wire 0 ~ 10 vdc 12 ~ 18 vdc.
Abus technologies inc. 8 pt output characteristics (for pt20-12xxxx pressure transmitter) c d a b c d.
Abus technologies inc. 9 pt 6. Installation recommendation 1. Please confirm the configuration for pressure transmitter i.E. 2- wire, 3-wire and accordingly follow the wiring procedure. Ensure that the excitation voltage is supplied to the correct points and is with in the specified limits for that ...
Abus technologies inc. 10 pt 7.2 example for usage 8. Maintenance troubleshooting s. No. Specifications eliminate 1 no output • check the wiring configuration. • check the input power supply. 2 no variation in output • check for leakage at pressure port / point. • check the span set of indicating in...
Abus technologies inc. 11 pt 9. Safety precautions 1. The unit should be powered for 15 minutes before use. 2. Use in ambient temperature of 0-60˚c. 3. Avoid vibrations, shock, excessive dust, corrosive chemical materials or gaseous environment. 4. Input wire should not be too long. If measured sign...