Acaia ORION BEAN DOSER Operation Manual - Chapter 5: Calibration
11
Action
Operation
Enter Unit Settings
When Unit is displayed, short press Dose to enter Unit menu.
Toggle Unit Options
Rotate the dial to toggle between unit options: g (gram) and oZ (ounce).
Short press Dose to confirm.
4.5 About
Action
Operation
Enter Set Menu
Long press Power and rotate the dial to SEt. Press Dose to enter Settings
menu.
Enter Unit Settings
When About is displayed, short press Dose to display the firmware
version.
Chapter 5: Calibration
5.1 Calibrating the Orion
In weighing mode, press the Dose button 5 times to enter the user calibration mode.
Place a 500 g calibration weight on the platform and allow the Orion to calibrate.
1. While the Orion is in Weighing Mode and the weighing unit is in gram, press the Dose
button quickly multiple times until LED display shows [ CAL ] and then [ 500 ].
2. Place a 500 g calibration weight on platform and wait for the scale to execute
calibration procedure.
3. The Display shows [ -End- ] when the calibration procedure is completed.
4. Remove the calibration weight from the platform when calibration is completed.
* Note: Calibration weights of 100 g, 1000 g, 1500 g, 2000 g, and 3000 g can also be
used to calibrate the Orion.