Summary of D-TRONplus

  • Page 1

    Insulin pump reference manual.

  • Page 2

    Front back display cartridge compartment adapter control keys powerpack compartment year of manufacture errors and alerts serial number infrared interface.

  • Page 3

    Accu-chek d-tronplus insulin pump reference manual 1 dear accu-chek d-tronplus insulin pump user, congratulations on your new insulin pump. You and your insulin pump are unique. Therefore, it is important that you or your health care team fill in the following information. Accu-chek d-tronplus insul...

  • Page 4

    Accu-chek d-tronplus insulin pump reference manual 3 explanation of the symbols and markings accu-chek d-tronplus insulin pump reference manual 2 rx only ipx7 read instructions for use. Marking of conformity to the european medical device directive 93/42/eec (mdd) with the number of the notified bod...

  • Page 5

    Accu-chek d-tronplus insulin pump reference manual 5 introduction: safe and efficient use of your accu-chek d-tronplus insulin pump 9 1 warnings and precautions 11 1.1 warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 1.2 precautions . . . . . . . . ....

  • Page 6

    Accu-chek d-tronplus insulin pump reference manual 7 7 alerts and errors 133 7.1 overview of the alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .135 7.1.1 alert a1: cartridge low warning . . . . . . . . . . . . . . . . . . . .136 7.1.2 alert a2: low powerpack warning . . . . ...

  • Page 7

    Introduction: safe and efficient use . . . Accu-chek d-tronplus insulin pump reference manual 8 10 care of your accu-chek d-tronplus insulin pump 171 10.1 repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .171 10.2 maintenance and cleaning . . . . . . . ....

  • Page 8

    Warnings and precautions 1 accu-chek d-tronplus insulin pump reference manual 11 accu-chek d-tronplus insulin pump reference manual 10 this section includes: 1.1 warnings 1.2 precautions caution: rx only. This device is restricted to sale by or on the order of a physician. Prior to starting your acc...

  • Page 9

    Warning: never change the cartridge or prime the infusion set with an infusion set inserted into your body. You risk uncontrolled insulin delivery. With disconnectable infusion sets, make sure you have discon- nected the tubing from your site prior to changing or priming. Always follow the instructi...

  • Page 10

    Accu-chek d-tronplus insulin pump reference manual 15 accu-chek d-tronplus insulin pump reference manual 14 pen/syringe and insulin with you. Without insulin, diabetic ketoacidosis may develop and this may require in-patient hospital treatment. Caution: if an “error e4: occlusion” occurs, immediatel...

  • Page 11

    Accu-chek d-tronplus insulin pump reference manual 17 accu-chek d-tronplus insulin pump reference manual 16 due to the large variety of devices with electromagnetic radiation such as cellular phones, their influence on your insulin pump can not be ruled out completely. It is advisable to keep your i...

  • Page 12

    Correct wrong accu-chek d-tronplus insulin pump reference manual 19 accu-chek d-tronplus insulin pump reference manual 18 your insulin pump, sterile products and accessories should be used for the first time in the presence of your doctor and/or your health care team. Regular medical exa- minations ...

  • Page 13

    • to tighten or loosen the powerpack, use only the battery tool supplied with your accu-chek d-tronplus insulin pump (never use knives, coins, screwdrivers and the like). The powerpack is correctly inserted and tight- ened when the position of the slot is parallel to the cartridge compart- ment (bay...

  • Page 14

    Description and functions ... 2 accu-chek d-tronplus insulin pump reference manual 23 this section includes: 2.1 safety and warning features 2.2 handling and display 2.3 operating status (stop/run) 2.4 operating logic and overview of accu-chek d-tronplus insulin pump functions 2.5 standard configura...

  • Page 15

    Accu-chek d-tronplus insulin pump reference manual 25 accu-chek d-tronplus insulin pump reference manual 24 stop- warning alerts and errors keys unlocked scrolling temporary basal rate beep one long beep every minute reminds you that your insulin pump is in stop and that no insulin is being delivere...

  • Page 16

    Accu-chek d-tronplus insulin pump reference manual 27 accu-chek d-tronplus insulin pump reference manual 26 d menu this key is used to select or move to a function or a value. F check this key is used to confirm the selected function or value and to activate it. The check key is also used to confirm...

  • Page 17

    Accu-chek d-tronplus insulin pump reference manual 29 accu-chek d-tronplus insulin pump reference manual 28 description of the display symbols your insulin pump is functioning (flashing symbol) standard/scroll bolus (quick delivery) extended bolus (delivery over a programmed period of time) basal ra...

  • Page 18

    Accu-chek d-tronplus insulin pump reference manual 31 accu-chek d-tronplus insulin pump reference manual 30 date format (european or american) stop – no insulin delivery beep volume keylock activated or deactivated/basal rate programming and profile selection locked or unlocked setup menu percentage...

  • Page 19

    Accu-chek d-tronplus insulin pump reference manual 33 accu-chek d-tronplus insulin pump reference manual 32 2.4 operating logic and overview of accu-chek d-tronplus insulin pump functions the operation of your accu-chek d-tronplus insulin pump is logical and simple. Most operations are carried out u...

  • Page 20

    Keylock your insulin pump is equipped with a unique keylock function that can be activated or deactivated. The keylock function allows you to lock the four accu-chek d-tronplus insulin pump keys and serves as additional safety measure against unintentional activation of accu-chek d-tronplus insulin ...

  • Page 21

    Accu-chek d-tronplus insulin pump reference manual 37 accu-chek d-tronplus insulin pump reference manual 36 2.5 standard configuration warning: incorrect programming of your insulin pump may cause inappropriate insulin delivery. Your insulin pump must be pro- grammed with your personal settings prio...

  • Page 22

    Accu-chek d-tronplus insulin pump reference manual 39 accu-chek d-tronplus insulin pump reference manual 38 if the parameters set in your insulin pump do not meet your needs, changes can be made using the accu-chek dialog pump programming tool. Consult your doctor and/or your health care team about ...

  • Page 23

    Accu-chek d-tronplus insulin pump reference manual 41 accu-chek d-tronplus insulin pump reference manual 40 your insulin pump configuration and settings (basal rate, active profile, automatic off, etc.) and the event memory (bolus history, history of daily totals, history of alerts and errors) remai...

  • Page 24

    Accu-chek d-tronplus insulin pump reference manual 43 accu-chek d-tronplus insulin pump reference manual 42 note that your insulin pump must be used with pre-filled 3.0 ml humalog ® * pen cartridges and/or user-filled 3.0 ml glass car- tridges with rubber membrane only. Your insulin pump is not guar...

  • Page 25

    Accu-chek d-tronplus insulin pump reference manual 45 accu-chek d-tronplus insulin pump reference manual 44 • the cartridge compartment is sealed. • the luer-lock end of the infusion set is protected. Caution: use your insulin pump for insulin delivery in the lock position only. Caution: if your ins...

  • Page 26

    3 set up your accu-chek d-tronplus insulin pump accu-chek d-tronplus insulin pump reference manual 47 accu-chek d-tronplus insulin pump reference manual 46 this section includes: 3.1 insert and change the powerpack 3.1.1 insert the powerpack 3.1.2 change the powerpack 3.2 review or set the time and ...

  • Page 27

    Accu-chek d-tronplus insulin pump reference manual 49 accu-chek d-tronplus insulin pump reference manual 48 3.1.1 insert the powerpack if there is currently no powerpack in your accu-chek d-tronplus insulin pump: 1. Remove the powerpack from the packaging by pushing it through the cardboard base. A ...

  • Page 28

    Accu-chek d-tronplus insulin pump reference manual 51 accu-chek d-tronplus insulin pump reference manual 50 caution: if you discover incomplete characters, numbers or symbols on the display, your insulin pump needs to be returned since you may not be notified of any changes of your insulin pump in a...

  • Page 29

    Accu-chek d-tronplus insulin pump reference manual 53 accu-chek d-tronplus insulin pump reference manual 52 after inserting the powerpack, your insulin pump performs the following start-up procedure: • the display indicates the software version (i.E. 3.00) and 77. Your insulin pump beeps a sequence ...

  • Page 30

    Accu-chek d-tronplus insulin pump reference manual 55 accu-chek d-tronplus insulin pump reference manual 54 3.2 review or set time and date caution: incorrect setting of time and date can lead to incorrect insulin delivery. Make sure that time and date are programmed correctly. This en- sures correc...

  • Page 31

    Accu-chek d-tronplus insulin pump reference manual 57 accu-chek d-tronplus insulin pump reference manual 56 2. Press d 9 times to select the setup menu. 3. Confirm with f . The time and date are displayed. 4. You have two options for returning to run : • press d until “end” appears on the display an...

  • Page 32

    Accu-chek d-tronplus insulin pump reference manual 59 accu-chek d-tronplus insulin pump reference manual 58 3.3 program the basal rate profiles warning: incorrect programming of your insulin pump may cause inappropriate insulin delivery. Your insulin pump must be pro- grammed with your personal sett...

  • Page 33

    Accu-chek d-tronplus insulin pump reference manual 61 accu-chek d-tronplus insulin pump reference manual 60 2. Choose basal rate profile a or b by pressing d 4 or 5 times. The daily basal rate total for 24 hours is displayed. (in this ex- ample the daily basal rate total is 18 units.) 3. Confirm wit...

  • Page 34

    Accu-chek d-tronplus insulin pump reference manual 63 accu-chek d-tronplus insulin pump reference manual 62 3.3.2 copy an hourly basal rate a basal rate can be copied for 1 or more hours. To copy an hourly basal rate 1. Make sure your accu-chek d-tronplus insulin pump is in run and the keylock is de...

  • Page 35

    Accu-chek d-tronplus insulin pump reference manual 65 accu-chek d-tronplus insulin pump reference manual 64 5. Change the selected hourly basal rate with a or s . 6. Repeat steps 4 and 5 as required. 7. Confirm your changes with f . The changes you just made are marked to save but not saved yet. The...

  • Page 36

    4 prepare the cartridge and attach the adapter ... Accu-chek d-tronplus insulin pump reference manual 67 accu-chek d-tronplus insulin pump reference manual 66 this section includes: 4.1 fill the cartridge 4.1.1 prepare the 3.15 ml plastic cartridge (steps 1 to 6) 4.1.2 fill the 3.15 ml plastic cartr...

  • Page 37

    Accu-chek d-tronplus insulin pump reference manual 69 accu-chek d-tronplus insulin pump reference manual 68 2. Filling needle with protective cap 3. Combifill base 4. Combifill extension 5. 3.0 ml glass cartridge 6. Instructions for use your insulin pump can be used with 3.0 ml glass cartridges. In ...

  • Page 38

    Accu-chek d-tronplus insulin pump reference manual 71 accu-chek d-tronplus insulin pump reference manual 70 4.1. 2 fill the 3.15 ml plastic cartridge (steps 7 to 15) move the plunger rod slowly and steadily when filling the cartridge. Do not move the plunger rod side-to-side as the seal between the ...

  • Page 39

    Accu-chek d-tronplus insulin pump reference manual 73 accu-chek d-tronplus insulin pump reference manual 72 4.1.3 fill the insulin pump 3.0 ml glass cartridge (steps 16 to 24) 16. Clean the rubber membrane of the new, empty glass car- tridge with an alcohol wipe before filling it. 17. Snap the glass...

  • Page 40

    Accu-chek d-tronplus insulin pump reference manual 75 accu-chek d-tronplus insulin pump reference manual 74 4.2 return the piston rod the piston rod must be returned in order to insert a new cartridge. After you return the piston rod, your insulin pump performs a self-test. To return the piston rod ...

  • Page 41

    Accu-chek d-tronplus insulin pump reference manual 77 accu-chek d-tronplus insulin pump reference manual 76 caution: do not interrupt the self-test by pressing keys or inserting a car- tridge or adapter. Interruption of the self-test may lead to malfunctions of your insulin pump. Caution: if your in...

  • Page 42

    Accu-chek d-tronplus insulin pump reference manual 79 accu-chek d-tronplus insulin pump reference manual 78 4.4 connect the adapter and infusion set in order for your insulin pump system to be complete, the adapter and infusion set must be connected to each other and then to your insulin pump and ca...

  • Page 43

    Accu-chek d-tronplus insulin pump reference manual 81 accu-chek d-tronplus insulin pump reference manual 80 warning: your insulin pump has been designed for use with accu-chek infusion sets only. The use of other infusion sets can endanger your health and may void the warranty. Infusion sets are ste...

  • Page 44

    Accu-chek d-tronplus insulin pump reference manual 83 accu-chek d-tronplus insulin pump reference manual 82 to prime the infusion set 1. Make sure your insulin pump is in stop and the keylock is deactivated or the keys are unlocked. 2. Press d 3 times to select the prime infusion set menu. 3. Confir...

  • Page 45

    Accu-chek d-tronplus insulin pump reference manual 85 accu-chek d-tronplus insulin pump reference manual 84 stop the priming the priming process can be stopped at any time by pressing any key. Your insulin pump returns to stop . Caution: air bubbles in the cartridge, adapter or the infusion set caus...

  • Page 46

    5 operate your accu-chek d-tronplus insulin pump accu-chek d-tronplus insulin pump reference manual 87 accu-chek d-tronplus insulin pump reference manual 86 this section includes: 5.1 start and stop insulin delivery 5.2 program a bolus 5.2.1 standard bolus 5.2.2 scroll bolus 5.2.3 extended bolus 5.3...

  • Page 47

    Accu-chek d-tronplus insulin pump reference manual 89 accu-chek d-tronplus insulin pump reference manual 88 to put your insulin pump from run into stop 1. Make sure your insulin pump is in run and the keylock is deactivated or the keys are unlocked. 2. Press d 2 times. The -s-- symbol flashes. 3. Pr...

  • Page 48

    Accu-chek d-tronplus insulin pump reference manual 91 accu-chek d-tronplus insulin pump reference manual 90 5.2 program a bolus the bolus amount is determined by your health care team’s guidelines, your blood glucose level, your food intake and your activity level. Warning: the amount of insulin del...

  • Page 49

    Accu-chek d-tronplus insulin pump reference manual 93 accu-chek d-tronplus insulin pump reference manual 92 5. 5 seconds after the confirmation, your insulin pump beeps** three times and starts to deliver the total bolus amount pro- grammed. To start with the a key: 1. Make sure your insulin pump is...

  • Page 50

    Accu-chek d-tronplus insulin pump reference manual 95 accu-chek d-tronplus insulin pump reference manual 94 during bolus delivery: if the bolus delivery has begun, it can be cancelled at any time by pressing s or a for 3 seconds after which your insulin pump beeps* a melody and an “alert a8: bolus c...

  • Page 51

    Accu-chek d-tronplus insulin pump reference manual 97 accu-chek d-tronplus insulin pump reference manual 96 press f and your insulin pump returns to run . No bolus is delivered! During bolus delivery: if the scroll bolus delivery has begun, it can be cancelled at any time by pressing s or a for 3 se...

  • Page 52

    Accu-chek d-tronplus insulin pump reference manual 99 accu-chek d-tronplus insulin pump reference manual 98 6. Press/scroll a to lengthen or s to shorten the bolus duration. Info: if a bolus duration of 00:00 is programmed, a normal scroll bolus is delivered. Info: if you first program the time and ...

  • Page 53

    Accu-chek d-tronplus insulin pump reference manual 101 accu-chek d-tronplus insulin pump reference manual 100 5.3 change the cartridge, adapter and infusion set warning: never change the cartridge or prime the infusion set with an infusion set inserted into your body. You risk uncontrolled insulin d...

  • Page 54

    Accu-chek d-tronplus insulin pump reference manual 103 accu-chek d-tronplus insulin pump reference manual 102 6. Set up your insulin pump with the new cartridge, adapter and infusion set. Please refer to section 4 “prepare the cartridge and attach the adapter and infusion set” for further informatio...

  • Page 55

    Accu-chek d-tronplus insulin pump reference manual 105 accu-chek d-tronplus insulin pump reference manual 104 6. Set up your insulin pump with the new adapter and infusion set. Please refer to section 4 “prepare the cartridge and attach the adapter and infusion set” for further information. 5.5 chan...

  • Page 56

    Accu-chek d-tronplus insulin pump reference manual 107 accu-chek d-tronplus insulin pump reference manual 106 5. Remove the infusion set from the adapter and dispose of it properly. You may use the battery tool to unscrew the luer connection between the infusion set and adapter if it cannot be un- s...

  • Page 57

    6 program other functions (special functions) accu-chek d-tronplus insulin pump reference manual 109 accu-chek d-tronplus insulin pump reference manual 108 this section includes: 6.1 review the data memory 6.1.1 review the bolus history 6.1.2 review the history of daily totals 6.1.3 review the histo...

  • Page 58

    Accu-chek d-tronplus insulin pump reference manual 111 accu-chek d-tronplus insulin pump reference manual 110 3. Press s to review the previous bolus. 4. Use a or s to review further or previous boluses. If there are less than 10 boluses in the bolus history, there are dashes in the display. 5. Pres...

  • Page 59

    Accu-chek d-tronplus insulin pump reference manual 113 accu-chek d-tronplus insulin pump reference manual 112 6.1.3 review the history of alerts and errors the history of alerts and errors enables you to review the last 10 alerts or errors star- ting with the newest entry in chronological order. To ...

  • Page 60

    Accu-chek d-tronplus insulin pump reference manual 115 accu-chek d-tronplus insulin pump reference manual 114 5. Press f to confirm the desired profile. Your insulin pump returns to run . Info: if f is not pressed within 20 seconds*, your insulin pump automa- tically returns to run (timeout). The ch...

  • Page 61

    Accu-chek d-tronplus insulin pump reference manual 117 accu-chek d-tronplus insulin pump reference manual 116 rary basal rate over” occurs. Silence and confirm the alert by pressing f twice. Your insulin pump remains in run . Please refer to section 7.1.7 “alert a7: temporary basal rate over” for fu...

  • Page 62

    Accu-chek d-tronplus insulin pump reference manual 119 accu-chek d-tronplus insulin pump reference manual 118 6.4 automatic off this safety function triggers an “error e3: automatic off” (your insulin pump stops insulin delivery!) if no key is pressed within a certain time. Automatic off is set to o...

  • Page 63

    Accu-chek d-tronplus insulin pump reference manual 121 accu-chek d-tronplus insulin pump reference manual 120 to adjust the beep volume 1. Make sure your insulin pump is in run and the keylock is deactivated or the keys are unlocked. 2. Press d 9 times to select the setup menu. 3. Confirm with f . 4...

  • Page 64

    Accu-chek d-tronplus insulin pump reference manual 123 accu-chek d-tronplus insulin pump reference manual 122 5. Confirm with f . The current bolus increment flashes. 6. Press/scroll a or s to increase or decrease the bolus increment. 7. Confirm with f . Your insulin pump returns to run . Info: if f...

  • Page 65

    Accu-chek d-tronplus insulin pump reference manual 125 accu-chek d-tronplus insulin pump reference manual 124 6.8 keylock your insulin pump is equipped with a unique keylock function that can be activated or deactivated. The keylock function allows you to lock the four accu-chek d-tronplus insulin p...

  • Page 66

    Accu-chek d-tronplus insulin pump reference manual 127 accu-chek d-tronplus insulin pump reference manual 126 9. If no further key is pressed, your insulin pump automatically locks all four keys after 10 seconds (keylock timeout). It then beeps** three times and the keylock symbol appears on the dis...

  • Page 67

    Accu-chek d-tronplus insulin pump reference manual 129 accu-chek d-tronplus insulin pump reference manual 128 the time remaining (in days) until the occurrence of the “error e5: end of use time” for t1 or t2 respectively can be reviewed in the setup menu at any time. Please refer to sections 7.1.5 “...

  • Page 68

    Accu-chek d-tronplus insulin pump reference manual 131 accu-chek d-tronplus insulin pump reference manual 130 6.10 data transfer the data memory of your accu-chek d-tronplus insulin pump stores the last 1400 events (i.E. Alerts and errors, programming operations and insulin delivery records). These ...

  • Page 69

    7 alerts and errors accu-chek d-tronplus insulin pump reference manual 133 accu-chek d-tronplus insulin pump reference manual 132 this section includes: 7.1 overview of the alerts 7.1.1 alert a1: cartridge low warning 7.1.2 alert a2: low powerpack warning 7.1.3 alert a3: set time/date 7.1.4 alert a4...

  • Page 70

    Accu-chek d-tronplus insulin pump reference manual 135 accu-chek d-tronplus insulin pump reference manual 134 no. Meaning of the alert what you have to do page a1 cartridge low warning silence and confirm the alert. 136 change the cartridge before it is completely empty. A2 low powerpack warning sil...

  • Page 71

    Accu-chek d-tronplus insulin pump reference manual 137 accu-chek d-tronplus insulin pump reference manual 136 7.1.2 alert a2: low powerpack warning your accu-chek d-tronplus insulin pump checks the powerpack voltage before each insulin delivery (at least every 3 minutes). If it falls below a certain...

  • Page 72

    Accu-chek d-tronplus insulin pump reference manual 139 accu-chek d-tronplus insulin pump reference manual 138 set up your insulin pump with a new cartridge, adapter and infusion set tubing and prime. Please refer to section 4 “prepare the cartridge and attach the adapter and infusion set” for furthe...

  • Page 73

    Accu-chek d-tronplus insulin pump reference manual 141 accu-chek d-tronplus insulin pump reference manual 140 set up your insulin pump with a new cartridge, adapter and infusion set and prime. Please refer to section 4 “prepare the cartridge and attach the adapter and infusion set” for further infor...

  • Page 74

    Accu-chek d-tronplus insulin pump reference manual 143 accu-chek d-tronplus insulin pump reference manual 142 7.1.5 alert a5: end of use time alert an “alert a5: end of use time alert” occurs 56 days before the end of use time. Do the following: 1. Silence and confirm the alert by pressing f twice. ...

  • Page 75

    Accu-chek d-tronplus insulin pump reference manual 145 accu-chek d-tronplus insulin pump reference manual 144 no. Meaning of the error what you have to do page e1 empty cartridge silence and confirm the error. 146 act according to section 7.2.1 “error e1: empty cartridge”. When finished, put your in...

  • Page 76

    Accu-chek d-tronplus insulin pump reference manual 147 accu-chek d-tronplus insulin pump reference manual 146 7.2.2 error e2: powerpack depleted your insulin pump checks the powerpack voltage before each insulin delivery (at least every 3 minutes). If it falls below a certain value, an “error e2: po...

  • Page 77

    Accu-chek d-tronplus insulin pump reference manual 149 accu-chek d-tronplus insulin pump reference manual 148 in rare cases, the cartridge itself can be the cause of the detected error. Should an “error e4: occlusion” occur again after carrying out the corrective actions described, the whole system ...

  • Page 78

    Accu-chek d-tronplus insulin pump reference manual 151 accu-chek d-tronplus insulin pump reference manual 150 do the following: 1. Silence and confirm the error by pressing f twice. 2. Remove or disconnect your infusion set from the site. 3. Remove the cartridge, adapter and infusion set (tubing) fr...

  • Page 79

    8 troubleshooting guide accu-chek d-tronplus insulin pump reference manual 153 accu-chek d-tronplus insulin pump reference manual 152 this section includes: 8.1 therapy situations 8.1.1 troubleshooting high blood glucose levels (hyperglycemia) 8.1.2 troubleshooting low blood glucose levels (hypoglyc...

  • Page 80

    Accu-chek d-tronplus insulin pump reference manual 155 accu-chek d-tronplus insulin pump reference manual 154 insulin pump or infusion set causes blood in tubing disconnected or dislodged infusion set empty or reused cartridge failure to complete prime incorrect time setting incorrect programming or...

  • Page 81

    Insulin causes alcohol consumption basal rates too high bolus dose too high for amount of food eaten temporary basal rate too high accu-chek d-tronplus insulin pump reference manual 157 accu-chek d-tronplus insulin pump reference manual 156 8.1.2 troubleshooting low blood glucose levels (hypoglycemi...

  • Page 82

    9 special situations accu-chek d-tronplus insulin pump reference manual 159 accu-chek d-tronplus insulin pump reference manual 158 this section includes: 9.1 recommendations for daily use 9.2 interruption of pump therapy 9.3 your insulin pump and water 9.4 electromagnetic fields and hazardous areas ...

  • Page 83

    Accu-chek d-tronplus insulin pump reference manual 161 accu-chek d-tronplus insulin pump reference manual 160 interruption of use of your insulin pump for a longer period of time warning: contact your doctor and/or your health care team for an alter- nate therapy plan when you interrupt the insulin ...

  • Page 84

    Accu-chek d-tronplus insulin pump reference manual 163 accu-chek d-tronplus insulin pump reference manual 162 9.3 your insulin pump and water avoid deliberate immersion in water caution: your insulin pump is watertight, however you should avoid deliberate immersion in water e.G. Bathing, taking show...

  • Page 85

    Accu-chek d-tronplus insulin pump reference manual 165 accu-chek d-tronplus insulin pump reference manual 164 guidance and manufacturer’s declaration – electromagnetic emissions emissions test rf emissions cispr 11 rf emissions cispr 11 harmonic emissions iec 61000-3-2 voltage fluctuations/ flicker ...

  • Page 86

    Accu-chek d-tronplus insulin pump reference manual 167 accu-chek d-tronplus insulin pump reference manual 166 the accu-chek d-tronplus is intended for use in the electromagnetic environment spe- cified below. The customer or the user of the accu-chek d-tronplus should assure that it is used in such ...

  • Page 87

    Accu-chek d-tronplus insulin pump reference manual 169 accu-chek d-tronplus insulin pump reference manual 168 9.5 sports you can perform sporting activities while wearing your insulin pump as long as you protect your insulin pump while engaging in any sporting activities with no body con- tacts. Do ...

  • Page 88

    Accu-chek d-tronplus insulin pump reference manual 170 10 care of your accu-chek d-tronplus insulin pump accu-chek d-tronplus insulin pump reference manual 171 this section includes: 10.1 repair 10.2 maintenance and cleaning 10.2.1 system check 10.2.2 your insulin pump 10.2.3 if your insulin pump is...

  • Page 89

    Accu-chek d-tronplus insulin pump reference manual 173 accu-chek d-tronplus insulin pump reference manual 172 10.2.2 your insulin pump caution: check your insulin pump for chips or cracks at least once per day, especially if it is dropped. Do not use your insulin pump if it is chipped or cracked. If...

  • Page 90

    Accu-chek d-tronplus insulin pump reference manual 175 accu-chek d-tronplus insulin pump reference manual 174 10.2.4 adapter the adapter is a sterile product intended for single use only. Sterility is guaranteed for unopened packaging up to the indicated expiration date. Do not use if package is dam...

  • Page 91

    Accu-chek d-tronplus insulin pump reference manual 177 accu-chek d-tronplus insulin pump reference manual 176 10.3 store your insulin pump caution: when your insulin pump will not be in use for a long period of time, it must be properly stored, to prevent a malfunction. To store your insulin pump: •...

  • Page 92

    Accu-chek d-tronplus insulin pump reference manual 179 accu-chek d-tronplus insulin pump reference manual 178 • you visually inspect the cartridge. Make sure that the actual amount of insulin in the cartridge is equal to the calculated amount on the remaining cartridge content display. • the infusio...

  • Page 93

    11 technical data accu-chek d-tronplus insulin pump reference manual 181 maximum dimen- sions (l × d × h) weight pump casing water protection temperature ranges air humidity ranges atmospheric pressure ranges connections power supply service life of powerpack data storage time 105 × 48 × 21.5 mm (4....

  • Page 94

    Average occlusion volume before error delivery basal rate bolus temporary basal rate safety system maximum infusion pressure before error flow (delivery rate) maximum quantity delivered at first error accu-chek d-tronplus insulin pump reference manual 183 precision of flow rate the trumpet curve sho...

  • Page 95

    List of symbols accu-chek d-tronplus insulin pump reference manual 185 accu-chek d-tronplus insulin pump reference manual 184 warning, explanation see page 3 caution, explanation see page 3 info, explanation see page 3 read instructions for use sterilized using ethylene oxide sterilized using radiat...

  • Page 96

    Glossary 13 accu-chek d-tronplus insulin pump reference manual 187 accu-chek d-tronplus insulin pump reference manual 186 adapter it physically connects the cartridge and the infusion set. The adapter’s built-in needle pierces the cartridge’s rubber membrane, thus allowing insulin to flow. Combined ...

  • Page 97

    Accu-chek d-tronplus insulin pump reference manual 189 accu-chek d-tronplus insulin pump reference manual 188 hyperglycemia excessively high blood glucose levels. This is observed when the person with diabetes receives too little insulin, eats too much, performs less activity or exercise than usu- a...

  • Page 98

    Accu-chek d-tronplus insulin pump reference manual 191 accu-chek d-tronplus insulin pump reference manual 190 personal settings settings are values like your hourly basal rate, your basal rate profile, the actual time and date, boluses, temporary basal rates, etc. Which can be programmed on your ins...

  • Page 99

    Accu-chek d-tronplus insulin pump reference manual 193 accu-chek d-tronplus insulin pump reference manual 192 14 addresses ditributed by: united kingdom: roche diagnostics ltd. Bell lane lewes uk-east sussex bn7 1lg phone +44 1273 480444 fax +44 1273 480266 lewes.Insulinpumps@roche.Com australia: ro...

  • Page 100

    15 index accu-chek d-tronplus insulin pump reference manual 195 accu-chek d-tronplus insulin pump reference manual 194 a accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, 40, 46, 47 accu-chek dialog pump programing tool . 37, 38, 65, 110, 119, 12...

  • Page 101

    Accu-chek d-tronplus insulin pump reference manual 197 accu-chek d-tronplus insulin pump reference manual 196 chips or cracks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16, 173 cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ....

  • Page 102

    G glass cartridge filling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73 glossary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187 h handling . . . . . . . . . . . . . . . . ....

  • Page 103

    Accu-chek d-tronplus insulin pump reference manual 201 stop -warning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177 sun . ...

  • Page 104

    Accu-chek d-tronplus insulin pump reference manual 202.

  • Page 105

    K eylock activated and keys locked • standar d bolus • remaining cartridge content • backlight • scr oll bolus • e xtended bolus • review or set the time and date • automatic off • adjust the beep volume • adjust the bolus incr ement • lock or unlock the basal rate pr ogr amming and pr ofile selecti...

  • Page 106

    No. Meaning of the alert what you have to do page a1 cartridge low warning silence and confirm the alert. 136 change the cartridge before it is completely empty. A2 low powerpack warning silence and confirm the alert. 137 replace the powerpack within the next 24 hours. A3 set time/date silence and c...

  • Page 107

    Alert and error code list on the back of your accu-chek d-tronplus insulin pump alerts errors serial number year of manufacture symbol for “protection against the effects of temporary immersion in water” according to iec 60529 symbol for “protection against electric shock” symbol for “see instructio...

  • Page 108: Gut Zum Druck

    04543157001 /v01/08.04 accu-chek, dialog, disetronic, d-tron, d-tronplus are trademarks of a member of the roche group. Disetronic medical systems ag ch -3401 burgdorf, switzerland www.Disetronic.Com disetronic medical systems ag labeling services gut zum druck an: absender: datum: wir bitten sie, d...