    Battery pack sp with wifi perfusor® space / infusomat® space / infusomat® space p instructions for use valid for software 027l gb.

  • Page 2

    Contents before operation.................. ....................................................................... 3 chapter 1/2/3 intended use / general / safety instructions ........ 5 chapter 4 main menu ..................................................................... 9 chapter 5 options me...

  • Page 3: Caution:

    3 before operation the wireless option is supported since software version j and later of the pump. To upgrade the pump with newer software please contact your local b. Braun service. Handling the battery pack sp with wifi: the battery pack sp can be exchanged by any user. No special qualification i...

  • Page 4: Attention

    4 before operation note: in case of esd, pump may need plugged into wall outlet to re-start the battery. The battery is charged by the pump during connection to a/c power. When disconnected from a/c or in case of power failure, the pump automatically switches to battery power. Charging is displayed ...

  • Page 5: Intended Use

    5 chapter 1/2/3 intended use / general / safety instructions intended use the battery module of space guaranties operation independent from mains power during transport situation. The battery module contains in addition a wireless transceiver module to allow data transmission of a single pump during...

  • Page 6: Safety Instructions

    6 chapter 1/2/3 intended use / general / safety instructions when adding a wireless battery module to a pump the wireless module starts within 2 minutes. In case you remove the battery from the pump or switch off the pump when operated in battery mode the wireless module is switched off, too. In cas...

  • Page 7

    7 chapter 1/2/3 intended use / general / safety instructions please follow local ordinance and/or regulations for disposal fire or chemical burn hazard if mistreated. To avoid possible injury. ■ do not open or expose to heat above 80°c (176°f). ■ do not use damaged battery-packs. ■ do not attempt to...

  • Page 8

    8 chapter 1/2/3 intended use / general / safety instructions symbols on product symbol explanation caution, see documentation supplied with the product. Type cf unit with defibrillation protection protection class ii device labeling of electric and electronic devices according to directive 2002/96/e...

  • Page 9: Main Menu

    9 chapter 4 main menu main menu within the main menu the status of the wireless connection is displayed. In case the wireless transmitter is switched off, see also options menu, no symbol is shown in the status of the main menu. If the wireless is activated the status of the connection is shown. Wir...

  • Page 10: Signal Strength: The Signal

    10 chapter 4 main menu ip-address shows the actually assigned ip address to this infusion pump. Signal strength: the signal strength shows the quality of the connection. Baud rate is the maximum communication speed in mega bits per second (mbps). The maximum baud rate strongly depends on the used wi...

  • Page 11: Options Menu

    11 chapter 5 options menu options menu within the options menu the wireless mode can be switched on or off..

  • Page 12

    12 chapter 6 operation of a pump with wifi operation of a pump with wifi in case a wireless pump is connected to a spacestation the wireless transceiver is automatically switched off. When disconnected the wireless mode is automatically switched on again and operated with the previous settings. 6.1 ...

  • Page 13

    13 chapter 6 operation of a pump with wifi confirm the changes with “ok” and click on the button “accept”. To change the actual profile select an other one from the configuration list. The stored profiles can not be selected at the pump. In case the wireless communication is interrupted the pump sel...

  • Page 14: Ssid:

    14 chapter 6 operation of a pump with wifi 6.2 default configuration of the wireless battery: ssid: space_xxxxxx (xxxxxx represents the serial number of the pump with leading zero, e.G. 012345) ip-address (static): 192.168.101.41 network type: ad-hoc encryption: none (open network) 6.3 software appl...

  • Page 15: 6.4

    15 chapter 6 operation of a pump with wifi pumps. The data transfer within the bcc protocol is secured by checksums. To ensure the correctness of the data at the third party system it is strongly recommanded to use this checksums to identify possible transmission failures. Drug library upload via sp...

  • Page 16

    16 chapter 6 operation of a pump with wifi d. Continue on to step 4 3. If the module is connected to a wifi source you will need a pc with a browser that has port 80 access to this network. 4. Enter the target ip address in the web browser in order to reach the portal spaceonline (either 192.168.101...

  • Page 17

    17 chapter 6 operation of a pump with wifi c. Press “update software” a. -> “configuration” b. Username: config c. Password: config.

  • Page 18

    18 d. -> “configuration” -> select “maintenance and software updates” d. Wait about 10 minutes, until the image is loaded and the module has exited the update. 6. Checking the software version a. It shall again start by showing the main image with the latest ssid b. You can check the software versio...

  • Page 19: Ssid:

    19 6.4.2 update via fall-back image within the fall-back image the wireless is set to ad-hoc mode. To connect to the wireless module use the following settings: ssid: spacecomfallback network mode: ad-hoc after reconnecting to the spaceonline (http://192.168.101.41) you can select the software you w...

  • Page 20: 6.5

    20 do not restart the wireless module before you see this information in your browser. After the automatic restart of the wireless module you can connect to the module using the previous wireless settings before you have updated the module. In case you selected restore factory defaults after update ...

  • Page 21: 6.5.2   Login

    21 the local alarm displays at the infusion pumps. Therapeutic and diagnostic decisions must not be made solely on the basis of the displays of the web application. Spaceonline offers different micro- sites upon which different data or configurations are possible. 6.5.1 setting of a network connecti...

  • Page 22: 6.6

    22 status service config user name status service config password status page service page configuration page activated for in order to prevent unauthorized access you are recommended to change the individual passwords after first start-up. For further information regarding configuration please refe...

  • Page 23: 6.7

    23 spaceonline does show actual data network connection is established a connection status display is integrated in the header. The network connection between pc or browser and spaceonline is shown and the status of spaceonline is monitored using the two status displays, red/yellow/green. The displa...

  • Page 24: 6.8.1    User Settings

    24 6.8.1 user settings it is possible to administer the password of individual logins, confi- gure the bcc protocol and determine the network settings. The ward id can also be set. To restore the factory defaults please select „restore factory defaults after reboot“. An other option is the use of a ...

  • Page 25

    25 6.8.1.2 change user name the predefined user names can be changed in this menu. However, it is not possible to add new or to delete existing user names. 6.8.1.3 change access authorization the access authorization can be given for each individual user name. The user access rights can be set and m...

  • Page 26: 6.8.3  Database Settings

    26 for detailed information about compatible systems, please refer to http://www.Space.Bbraun.Com. The settings of the baud rate as well as parity, stopbits and databits, comply with the requirements of the pdm system. You also can setup a tcp/ip interface for the communication with spacecom via the...

  • Page 27: 6.8.5    Ftp Settings

    27 6.8.5 ftp settings spacecom allows to setup a ftp connection. Please enter the re- quired data to enable the ftp server at spacecom. The ftp access is for service purpose only. The ftp access requires explicit ssl connection. This has to be supported by the ftp client. 6.8.6 space onlinesuite sin...

  • Page 28: Technical Data

    28 chapter 8 technical data technical data the technical data of the pump keep valid in case here are no other data are described. Technical data battery module: general storage time: full charge at 3,5 month inside the pump 1 year outside the pump charging time: up to 5 h operating time: wireless a...

  • Page 29: Wireless Module

    29 chapter 8 technical data battery voltage: 6 to 8,2v wireless module operating system: linux 2.6.34 kernel processor: arm9 based memory: 32mb flash, 64mb ram radio module: ieee 802.11a/b/g/n (ieee 802.11g/n/b channel 1-13 ieee 802.11a/n channel 36,40,44,48-140) transmit power: 100mw supported mode...

  • Page 30: Gnu General Public License

    30 chapter 9 gnu general public license gnu general public license version 2, june 1991 copyright (c) 1989, 1991 free software foundation, inc., 51 franklin street, fifth floor, boston, ma 02110-1301 usa. Everyone is permitted to copy and distribute verbatim copies of this license document, but chan...

  • Page 31

    31 these restrictions translate to certain responsibilities for you if you distribute copies of the software, or if you modify it. For example, if you distribute copies of such a program, whether gratis or for a fee, you must give the recipients all the rights that you have. You must make sure that ...

  • Page 32: Gnu General Public License

    32 gnu general public license terms and conditions for copying, distribution and modification 0. This license applies to any program or other work which contains a notice placed by the copyright holder saying it may be distributed under the terms of this general public license. The “program”, below,...

  • Page 33

    33 and you may at your option offer warranty protection in ex- change for a fee. 2. You may modify your copy or copies of the program or any por- tion of it, thus forming a work based on the program, and copy and distribute such modifications or work under the terms of sec- tion 1 above, provided th...

  • Page 34

    34 themselves, then this license, and its terms, do not apply to those sections when you distribute them as separate works. But when you distribute the same sections as part of a whole which is a work based on the program, the distribution of the whole must be on the terms of this license, whose per...

  • Page 35

    35 distributed under the terms of sections 1 and 2 above on a medium customarily used for software interchange; or, c) accompany it with the information you received as to the offer to distribute corresponding source code. (this alternative is allowed only for noncommercial distribution and only if ...

  • Page 36

    36 terminated so long as such parties remain in full compliance. 5. You are not required to accept this license, since you have not signed it. However, nothing else grants you permission to modify or distribute the program or its derivative works. These actions are prohibited by law if you do not ac...

  • Page 37

    37 if any portion of this section is held invalid or unenforceable under any particular circumstance, the balance of the section is intended to apply and the section as a whole is intended to apply in other circumstances. It is not the purpose of this section to induce you to infringe any patents or...

  • Page 38

    38 program specifies a version number of this license which applies to it and “any later version”, you have the option of following the terms and conditions either of that version or of any later version published by the free software foundation. If the program does not specify a version number of t...

  • Page 39

    39 12. In no event unless required by applicable law or agreed to in writing will any copyright holder, or any other party who may modify and/or redistribute the program as permitted above, be liable to you for dam- ages, including any general, special, incidental or con- sequential damages arising ...

  • Page 40

    38917342 • drawing no. I0026700005 0915 3891 7342.2013f10.Km b. Braun melsungen ag sparte hospital care 34209 melsungen germany tel.: +49 (0) 56 61 71-0 fax: +49 (0) 56 61 71-20 44 www.Bbraun.Com manufactured by b. Braun melsungen ag 34209 melsungen germany tel +49 (0) 56 61 71-0