Bacharach 3015-4521 Instructions

Summary of 3015-4521

  • Page 1

    These instructions describe how to replace the ir optical bench in the agm300 ammonia gas monitor. It is assumed that the user is familiar with the operation of the agm300 / adm800 ammonia gas monitoring system. If necessary, refer to instruction 3015-4275 for detailed operation and maintenance info...

  • Page 2

    Instruction 3015-4723 2 4. Important! For added safety, turn off the agm300 by pushing the internal ac power switch to its down position! This step is in addition to setting the monitor’s ac circuit breaker to off. Recheck the front panel monitor on indicator light to ensure that it is off. With ac ...

  • Page 3

    Instruction 3015-4723 3 adjusting the ir optical bench detector voltage 10. Check, and possibly adjust, the ir optical bench detector voltage by first going to the adm800 and pushing the button on the system screen that corresponds to the agm300 monitor being serviced. This action will display that ...

  • Page 4

    Instruction 3015-4723 4 13. Go back to the agm300 and find gain adj potentiometer r34 on the main board. Using a small flat-blade screwdriver, turn r34 clockwise to increase, or counterclockwise to decrease, the detector voltage to between 4.2 and 4.3 vdc. Make very small adjustments! 14. Secure the...