co 2 monitor 2800 instruction 19-9220 operation & maintenance rev. 6 – june 2004 product leadership • training • service • reliability.
Page 2: Warranty
Co 2 monitor 2800 instruction 19-9220 warranty bacharach, inc. Warrants to buyer that at the time of delivery this product will be free from defects in material and manufacture and will conform substantially to bacharach inc.’s applicable specifications. Bacharach’s li- ability and buyer’s rem...
Page 3: Contents
Instruction 19-9220 co 2 monitor 2800 i contents specifications.............................................................................. Ii 1.0 introduction.................................................................................1 1.1 general ...............................................
Page 4: Specifications
Co 2 monitor 2800 instruction 19-9220 ii specifications range: 0 - 50,000 ppm* or 0 - 5% resolution: 0.01% or 100 ppm accuracy: ±5000 ppm co 2 temperature: 0 - 40° c relative humidity: 10 - 95% non-condensing response time: size: 140 (l) x 65 (w) x 20 (d) mm (5.5 x 2.6 x 0.8 in.) weight: 200 grams (...
Page 5: 1.0 Introduction
Instruction 19-9220 co 2 monitor 2800 1.0 introduction 1.1 general the bacharach co 2 monitor model 2800 is easy to use, but itisessen- tialthattheseoperatinginstructionsare read prior to use. The co 2 monitor is designed to warn the user of increasing carbon diox- ide levels in the range of 0 to 50...
Page 6: 1.5 Important Note
Co 2 monitor 2800 instruction 19-9220 1.5 important note always ensure that the sensor grille located on the front of the unit is un- obstructed and open to the atmosphere. A quick instrument check can be performed by breathing directly into the sensor grille. This will cause the indicated co 2 leve...
Page 7: 2.0 Operation
Instruction 19-9220 co 2 monitor 2800 2.0 operation 2.1 reading mode when the instrument is switched on, there will be a 20 second (approxi- mately) warm up time before the co 2 level present in the atmosphere surrounding the user is displayed. The on/off button should be pressed quickly when switch...
Page 8: 3.0 Maintenance
Co 2 monitor 2800 instruction 19-9220 3.0 maintenance 3.1 battery low when the battery voltage falls below a predetermined level, the screens will alternate between the following: low batt and ppm co2 300 300 in addition, the alarm will emit three rapid tones every 30 seconds. The unit must be given...
Page 9: 3.4 Test Mode
Instruction 19-9220 co 2 monitor 2800 if this happens, retry the air calibration procedure, ensuring that the unit is only exposed to clean outside air. If the procedure is still unsuc- cessful then the unit must be returned to a bacharach service center (see section 4.2) for evaluation. This proced...
Page 10: 3.6 Data Logging
Co 2 monitor 2800 instruction 19-9220 3.5 fault condition warning the unit can warn the user of an internal fault, i.E. A source failure or blockage in the infrared path. When a fault occurs, both leds will illu- minate and the alarm will sound continuously. The following screen will display until t...
Page 11: 4.0 Parts & Service
Instruction 19-9220 co 2 monitor 2800 4.0 parts & service 4.1 parts and accessories accessories irda interface kit & bach-com software ..............................19-3301 12 vdc charger w/cigarette lighter adapter ..........................19-3302 earphone ............................................
Page 12
Co 2 monitor 2800 instruction 19-9220 4.2 bacharach service centers méxico bacharach de méxico playa regatas no. 473 tercer piso col. Militar marte delegación iztacalco, 08830 méxico d.F. México phones: +52-555-634-7740 +52-555-634-7741 fax: +52-555-634-7738 email: bacharach@prodigy.Net.Mx europe eu...
Page 13
Instruction 19-9220 co 2 monitor 2800
Page 16
Headquarters: 621 hunt valley circle, new kensington, pa 15068 ph: 724-334-5000 • fax: 724-334-5001 • toll free: 800-736-4666 website: www.Bacharach-inc.Com • e-mail: help@bacharach-inc.Com printed in u.S.A..