Summary of H-10 PRO

  • Page 1

    3015-9005 rev 1 1 instruction manual p/n: 3015-9005 revision 1 march 2016 product leadership • training • service • reliability www. .Com information@itm.Com 1.800.561.8187.

  • Page 2: Information@itm.Com

    2 3015-9005 rev 1 notice specifications and information contained in this document may change without notice due to product improvements and enhancements. Bacharach, inc. Shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, per...

  • Page 3: Information@itm.Com

    3015-9005 rev 1 3 section 1. Overview 1.1. Introduction the h-10 pro is the most versatile, high performance leak detector available in the industry today. It detects refrigerants to pinpoint small, medium, and large leaks quickly and efficiently. Review this manual carefully and completely to assur...

  • Page 4: Information@itm.Com

    4 3015-9005 rev 1 ! Warning: except for replacement of consumables such as sensors, filters, and battery, this analyzer should only be opened and/or serviced by authorized bacharach personnel. Failure to comply may void the warranty. ! Warning: do not store instrument or its sensors with solvents or...

  • Page 5: Information@itm.Com

    3015-9005 rev 1 5 spec description sensitivity [3] 0.006 oz/yr stationary 0.1 oz/yr moving per sae j2791 auto adjust to background yes, automatic mode only warm-up 2 min (typical) indi ca tio n visual red led in probe tip audible internal beeper or stereo headphone jack weight 5.1 lb. (2.3 kg) dimen...

  • Page 6: Information@itm.Com

    3015-9005 rev 1 6 section 2. Operation 2.1. Overview the h-10 pro sensor uses positive ion emission technology, commonly known as a heated diode. It is very sensitive to only halon substances (refrigerants) making this product highly resistive to false alarming, while retaining superior sensitivity ...

  • Page 7: Information@itm.Com

    3015-9005 rev 1 7 2.2. Panel controls www. .Com information@itm.Com 1.800.561.8187.

  • Page 8: Information@itm.Com

    8 3015-9005 rev 1 2.3. Before first use… before you use the unit the first time: 1. Remove the sticky-label disc from the cap of the calibration reference bottle. This disc can be reused to help seal the vial when the leak detector is not in use, or it can be discarded. Do not attempt to unscrew the...

  • Page 9: Information@itm.Com

    3015-9005 rev 1 9 has a leak rate of approximately 0.5 oz/year. Medium switch position is used for chlorine-based (cfc and hcfc) gases like r12 and r22. The medium setting will indicate approximately a 0.1 oz/yr (or greater) leak rate. This position supports finding leaks of approximately 0.5 oz/yr ...

  • Page 10: Information@itm.Com

    10 3015-9005 rev 1 procedure will reliably and repeatedly pinpoint the leak with each pass over the leak site. If a large leak is present, the auto zero circuit may reduce sensitivity to an unacceptable level to find small and medium sized leaks. If this condition exists, use the manual mode to pinp...

  • Page 11: Information@itm.Com

    3015-9005 rev 1 11 may operate the unit using the supplied wall adapter. ! Note: the sensor does not operate when the low battery led is on. 5. Check for sufficient airflow by pointing the probe tip toward the floor, covering it with your finger, then releasing your finger. If proper flow exists, th...

  • Page 12: Information@itm.Com

    12 3015-9005 rev 1 6. Allow two minutes for the sensor to warm up, after which the flashing probe light and sound indicator will idle at approximately 1 click per second. 7. Test operation by quickly touching the probe tip to the top of the calibration reference bottle (make sure sticky label is rem...

  • Page 13: Information@itm.Com

    3015-9005 rev 1 13 unit. Check for moisture before inserting the probe into areas to be checked for leaks. 4. If leaks have not been detected using the large leak setting: try using the medium or small setting as appropriate. If leaks have been detected using the large leak setting after a leak is l...

  • Page 14: Information@itm.Com

    14 3015-9005 rev 1 section 3. Calibration and performance 3.1. Heater adjustments the sensor heater circuit can be adjusted to control the sensitivity of the unit. A heat setting that is too high causes instability due to excessive sensitivity and shortens sensor life. A heat setting that is too low...

  • Page 15: Information@itm.Com

    3015-9005 rev 1 15 this procedure until the green ok led briefly glows. If the red high led glows, the heat is set too high and the heater adjustment should be turned slightly counterclockwise using a small screwdriver. Allow the unit to stabilize for about 1 minute and then repeat the test. Repeat ...

  • Page 16: Information@itm.Com

    16 3015-9005 rev 1 section 4. Maintenance 4.1. Replacing the filter to protect the pump from damage due to foreign particles and moisture, replace the filter as it becomes dirty. With moderate use (15 to 30 minutes a day), it is recommended that the filter be replaced once per month. In dirty enviro...

  • Page 17: Information@itm.Com

    3015-9005 rev 1 17 3. Insert the new airflow indicator ball into the tip. 4. Reattach the probe tip to the probe assembly. 4.3. Replacing the sensor the sensor needs to be replaced when the h-10 pro no longer responds to the calibration reference bottle, even with the heater adjustment turned fully ...

  • Page 18: Information@itm.Com

    18 3015-9005 rev 1 section 5. Troubleshooting 5.1. Diagnosing issues p=problem • c/s=cause/solution(s) p no response to calibration reference bottle. C/s heat adjustment is set too low or bottle is empty. Readjust heater (see “heater adjustment” section) or replace bottle (3015-0864). No air flow (i...

  • Page 19: Information@itm.Com

    3015-9005 rev 1 19 p erratic response occurs in all leak positions. C/s filter is clogged. Replace the filter (3015-0784). Dirt is in the sensor. Remove sensor and blow it out with clean air (not over 10 psi). If unsuccessful, replace the sensor (3015-0486). Replace the filter (3015-0784). Sensor ha...

  • Page 20: Information@itm.Com

    20 3015-9005 rev 1 5.2. Limited warranty the purchaser is warranted that this leak detector will be free of defects in material and workmanship for 3 years from date of purchase. This warranty does not cover sensors, reference leaks, filters, airflow balls, lamps, batteries, or probe tips. Damages c...

  • Page 21: Information@itm.Com

    3015-9005 rev 1 21 5.4. Spare parts part number description 3015-8004 h-10 pro refrigerant leak detector, battery, charger, and north american plug 3015-8005 h-10 pro refrigerant leak detector, battery, charger, and plug set (north american and international) 3015-0326 14” flexible probe extension 3...

  • Page 22: Information@itm.Com

    22 3015-9005 rev 1 5.5. Authorized service centers replacement parts and service can be obtained by contacting one of the following bacharach service centers. ∇ ∇ ∇ www. .Com information@itm.Com 1.800.561.8187.

  • Page 23: Information@itm.Com

    3015-9005 rev 1 23 the technical documentation file required by this directive is maintained at the corporate headquarters of bacharach, inc. Note: the directives or harmonized standards to which this product has been evaluated may have changed. Bacharach, inc. However continues to market this produ...