Bacharach PGM-IR Instruction

Manual is about: Portable Gas Monitor for Carbon Dioxide CO2 Gas

Summary of PGM-IR

  • Page 1

    Instruction 3015-5508 (rev 2) 1 portable gas monitor (pgm-ir) infrared sensor replacement instruction 3015-5508 rev. 2 – march 2013 1. Scope these instructions describe how to replace the ir sensor in the portable gas monitor (example shown in figure 1). It is assumed that the user is familiar with ...

  • Page 2

    Portable gas monitor (pgm-ir) infrared sensor replacement 2 instruction 3015-5508 (rev 2) figure 2. Internal components showing the ir sensor (bench) of a pgm-ir 2. Items required • ir sensor replacement kit which includes: o ir sensor assembly (p/n: see table 1) o cable ties, qty: 2 (p/n: 3015-2835...

  • Page 3

    Portable gas monitor (pgm-ir) infrared sensor replacement instruction 3015-5508 (rev 2) 3 figure 3. Disassembling the chassis 4. Remove the ir sensor step description 5. Review the internal components of the monitor, shown below, before proceeding. Note that figure 4 shows r37 – an analog potentiome...

  • Page 4

    Portable gas monitor (pgm-ir) infrared sensor replacement 4 instruction 3015-5508 (rev 2) figure 5. Monitor components in various pgm-ir models step description 8. Disconnect the tubing from the purge-air bag “t”-fitting, and then lay aside the chassis-half containing the purge-air bag. (this step d...

  • Page 5

    Portable gas monitor (pgm-ir) infrared sensor replacement instruction 3015-5508 (rev 2) 5 5. Install the new ir sensor step description 12. Record calibration label information (found on the new ir sensor) into table 2 for use later in this procedure. Refer to figure 6. Note that calibration informa...

  • Page 6

    Portable gas monitor (pgm-ir) infrared sensor replacement 6 instruction 3015-5508 (rev 2) 6. Adjust power level for ir emitter step description 23. To access the monitor’s set-up mode and display the ir emitter power adjustment screen, locate and view the portable monitor’s display. 24. With the mon...

  • Page 7

    Portable gas monitor (pgm-ir) infrared sensor replacement instruction 3015-5508 (rev 2) 7 8. Update calibration information step description 33. Scroll to tempcos and press enter. Enter your recorded tempco value from table 2 using the keypad buttons. Press enter to save the new value. • for co 2 se...

  • Page 8

    Portable gas monitor (pgm-ir) infrared sensor replacement 8 instruction 3015-5508 (rev 2) world headquarters: 621 hunt valley circle, new kensington, pa 15068-7074 usa toll free: 1-800-736-4666 • phone: +1-724-334-5000 • fax: +1-724-334-5001 website: www.Mybacharach.Com • e-mail: help@mybacharach.Co...