Bartington MAG-03PSU Operation & Maintenance Manual

Manual is about: POWER SUPPLY UNIT

Summary of MAG-03PSU

  • Page 1

    Gmw _________________________________________________________________________________ gmw 955 industrial road, san carlos, ca 94070 tel: (650) 802-8292 fax: (650) 802-8298 email: sales@gmw.Com web site: http://www.Gmw.Com bartington mag-03psu power supply unit operation/maintenance manual distribute...

  • Page 2

    Om0065 issue 5 page 2 of 21 list of contents 1 introduction 2 power 3 filters 4 buffers 5 cables 6 analog voltage measurements 7 mag-03psu with mag-03 series sensors 8 battery charging 9 low battery alarm 10 analog output 11 taking measurements 11.1 connecting up 11.2 filter section 11.3 additional ...

  • Page 3

    Om0065 issue 5 page 3 of 21 list of figures figure 1 mag-03 & mag-03psu simplified schematic figure 2 mag-03psu assembly figure 3 pc36 schematic - mag-03psu figure 4 pc36 assembly and filter selection figure 5 mag-03psu wiring details parts lists: pl0065 - mag-03psu assembly pl0063 - pc36 assembly p...

  • Page 4

    Om0065 issue 5 page 4 of 21 1 introduction this manual describes the operation and maintenance of the mag-03psu power supply unit. The unit provides a battery backed power supply of ± 12 v for the mag-03 series of three axis fluxgate magnetic sensors. It also contains filters for the analog outputs ...

  • Page 5

    Om0065 issue 5 page 5 of 21 5 cables cables can be supplied to connect any of the mag-03 series of sensors to the mag-03psu. Details are given in the operation manual for the sensor. The cable carries the supply voltage to the magnetic field sensor; 0 v connection is via the screen. Analog signals a...

  • Page 6

    Om0065 issue 5 page 6 of 21 8 battery charging a 12 v dc (nominal unregulated) output mains adaptor for battery charging purposes is supplied as an accessory. Before use ensure that this is of the appropriate voltage for the local mains supply, either 110 v or 220 v/240 v ac, 60 hz or 50 hz. As an a...

  • Page 7

    Om0065 issue 5 page 7 of 21 be selected if it is required to block the static background field and provide only an ac response. The decision to use either a dc or ac response will depend upon several factors. (a) the need or otherwise to measure the static field. (b) the dynamic range of any associa...

  • Page 8

    Om0065 issue 5 page 8 of 21 operation close to a source of high electromagnetic fields should be avoided to prevent pick-up which will be evident as noise in the output signal. 12 detailed circuit description (see figures 1 to 5) 12.1 power an unregulated voltage from the battery charger appears via...

  • Page 9

    Om0065 issue 5 page 9 of 21 13.2 open circuit conditions or no supply voltage ensure that power is supplied to the magnetic field sensor. Remove the cover of the mag-03psu power supply unit and establish that a voltage of ± 12 to ± 13 v is obtained on pins 5 and 7 respectively relative to pin 6 on p...

  • Page 10

    Om0065 issue 5 page 10 of 21 (ii) if it is believed that charging is not taking place connect a voltmeter across the battery. Whilst charging is taking place the battery voltage should increase to 6.7 v. (iii) if the battery voltage is not 6.7 v remove the power supply connector and check for around...

  • Page 11

    Om0065 issue 5 page 11 of 21 14 specifications mechanical materials : high strength abs dimensions : 133mm x 84mm x 46mm electrical sensor analog input and power output : 10 pole hirose plug type rm15trd-1op outputs : 3 bnc connectors sensor power supply : ± 12 v at 1 w battery : panasonic type lcr6...

  • Page 12

    Om0065 issue 5 page 12 of 21 dr1011(1) mag-03 & mag-03psu simplified schematic figure 1.

  • Page 13

    Om0065 issue 5 page 13 of 21 dr0065(7) mag-03psu assembly figure 2.

  • Page 14

    Om0065 issue 5 page 14 of 21 dr1013(2) pc36 schematic figure 3.

  • Page 15

    Om0065 issue 5 page 15 of 21 dr1014(1) pc36 assembly & filter selection figure 4.

  • Page 16

    Om0065 issue 5 page 16 of 21 dr1015(2) mag-03psu wiring details figure 5.

  • Page 17

    Om0065 issue 5 page 17 of 21 pl0065/6 mag-03psu assembly 1 pc 36 assembly pm 0063 1 2 front panel assembly pm 0855 1 3 rear panel assembly pm 0858 1 4 label cotswold dr 0067 1 screen printers 5 filter label bartington dr 0856 1 bond on inside of item 8 instruments 6 enclosure lower half dr 0859 1 us...

  • Page 18

    Om0065 issue 5 page 18 of 21 19 heatshrink tubing 2 2:1 shrink ratio 6.4 x 20mm long black 20 no. 6 self tapping screw 1 6.5mm long steel, bright zinc plated clear passivated 21 loctite 7400 as reqd 22 stick on feet 4 polyeurethane self adhesive Ø12.7 x 3.6 black 23 fr4 nut 1 used to locate ground i...

  • Page 19

    Om0065 issue 5 page 19 of 21 pl0063/5 pc36 assembly 1 pcb pc 36 pm 0061 1 2 d1,2 diode in4001 fare in4001 2 3 ic1 lm317t fare lm317t 1 4 ic2 lm2940ct-5-0 1 5 ic3 nma0512s fare 330-759 1 6 ic4 op400gp fare op400gp 1 7 ic5 vn4302ezb voltage 1 detector ic54 series 8 a1 sub-miniature piezo electrospeed ...

  • Page 20

    Om0065 issue 5 page 20 of 21 30 cable 7/0.2 grey x 150mm 1 31 cable 7/0.2 yellow x 150mm 1 32 cable 7/0.2 violet x 150mm 1 33 cable 7/0.2 green x 150mm 1 34 cable 7/0.2 blue x 150mm 1 35 cable 7/0.2 black x 150mm 2 36 cable 7/0.2 green/yellow 1 to complete esd shield x 100mm 37 m3 solder tag 1 38 4....

  • Page 21

    Om0065 issue 5 page 21 of 21 2 10-way panel jack plug fare 147-058 1 3 2.1mm dc inlet mapp ft96e 1 4 slimline rocker switch rsco 197-7692 1 5 m2 x 12mm pan pozi rsco 560-546 2 screw 6 m2 plain washer rsco 560-316 2 7 m2 nut rsco 560-271 2 8 led (red) rsco 826-492 1 is = 1.6ma algaas 9 mounting bezel...