Basic American Zenith 7000 Instructions For Use And Service Manual - page 22
TROUBLESHOOTING STAFF CONTROL
TROUBLESHOOTING PENDANT (& T-CABLE)
Basic American Medical Products, a division of GF Health Products, Inc.
www.grahamfield.com
999-0843-190A Service Manual
22
1. STAFF CONTROL WORKS BUT PENDANT/
HAND CONTROLLER FUNCTION DOESN’T
a. Check to see if the functions are locked out on the Staff
Control panel - if orange lights show on panel then that
function is locked out. Press the Hold/Select Button and
the blue pendant symbol at the same time until the orange
light goes out. Test the function. If the PENDANT STILL
DOESN’T WORK = go to step b.
b. Check the connections at the seat pan (Hand Controller
to T-Cable) and Staff Control Cables at the foot end. If the
Hand Controller still DOESN’T WORK = move to Step c.
c. UNPLUG THE POWER CORD FROM THE WALL OUTLET.
d. Unplug the Hand Controller plug from the end of the
T-Cable on either side of the beds seat pan. Unplug the
cables at the foot end of the bed that connect to the staff
control cables.
e. Plug the Hand Controller directly into the cylindrical plug
on the short Black Control Cable, making sure the pendant’s
phone jack plug is properly seated in the control cable
plug. Plug in the power cord and test the Pendant.
• ALL BUTTONS WORK = Replace T-Cable
See page 28 for order information.
• NOTHING WORKS = Replace the Hand Controller.
See page 28 for order information.
1. STAFF CONTROL LIGHTS ARE BLINKING
NOTHING ON STAFF CONTROL WORKS
BUT PENDANT WORKS
a. This could mean that your staff control is not getting
enough power from the Control Box.
b. Unplug the power cord from the wall outlet. Unplug
the staff control cables from Control Cable and T-Cable.
c. If you have a spare staff control assembly, plug the
cables into the T-Cable and Control Cable and test.
WORKS & LIGHTS DON’T BLINK =
Replace the Staff Control Assembly in your
footboard (See Page 23 for dis-assembly.
Please see Page 28 for order number.)
DOESN’T WORK & LIGHTS STILL BLINK =
Replace the Control Box. Please see
Page 27 for order number.
2. STAFF CONTROL - PENDANT WORKS BUT
NOTHING ON STAFF CONTROL WORKS
a. Check the Staff Control connections at the foot end -
(Staff Control cables plugged securely into T-Cable
and Control Cable). Also check if other end of black
Control Cable is plugged securely into the Control Box.
Test the staff control . . .
STILL DOESN’T WORK = move to Step b.
b. Unplug the power cord from your wall outlet. Remove
the Cable Cover (Shroud) on the inside of the foot-
board and locate the 2 terminal ends that plug directly
into the back of the Staff Control Panel in the footboard.
Make sure they are properly seated into the panel.
Plug in the power cord and test the staff control . . .
STILL DOESN’T WORK = Replace the Staff Control
assembly (See Page 23 for dis-assembly. See
Page 28 for order information).
IF YOU NEED TO REPLACE YOUR
PENDANT/HAND CONTROLLER
PLEASE USE ORDER KIT
# 999-0808-112.
ZENITH SERIES