CA.MI Mobile Instruction Manual

Manual is about: Flo series PORTABLE COMPRESSOR NEBULIZER

Summary of Mobile

  • Page 1

    Portable compressor nebulizer instruction manual www.Nebulisercentre.Co.Uk.

  • Page 2

    Mobile is a micro piston compressor system for aerosoltherapy, with universal power supply, thanks to the switching adaptor in equipment (model fuhua ue20-120167spa1 or ue24wcp-120167spa) and operated also by rechargeable batteries 1.2 ni-mh type aa, easily available from stores, with capacity betwe...

  • Page 3

    Instructions for the correct disposal of the product for the purpose and effect of european comunity directive 2002/96/ec: at the end of its useful life the product must not be disposed of with domestic waste. It may be delivered to the relevant separate waste collection centres prepared by municipa...

  • Page 4

    Class ii insulation device ce marking in conformity with ec directive 93/42/cee and subsequent changes warning: read instruction manual store in cool and dry place storing temperature: - 25 ÷ 70° c type b device ~ alternate current direct current hz mains frequency on / off a copy of ec declaration ...

  • Page 5

    How to return the unit for repair in compliance with the new european rules, ca-mi indicates some important points to preserve the hygiene of mobile and of the patients. These rules must be respected in order to guarantee hygiene, safety, quality and well being to anybody operating mobile. Every ret...

  • Page 6

    To insert batteries, position mobile turning it over a flat surface, with enough room. To access the batteries compartment, open the cover set on the base (fig. 1 – 2), pulling the two levers on the base and raising the cover until complete removal. Fig. 1 fig. 2 the batteries shall be placed in the...

  • Page 7

    Warning: before proceeding, always make sure not to start recharge cycle witj not-rechargeable alkaline batteries. Never recharge non-rechargeable alkaline batteries. Insert batteries ni-mh type aa – 1.2v in mobile, then connect it to the ac/dc adapter (or car plug cable) and to the electrical mains...

  • Page 8

    Tab ii – luminous indications during recharge of batteries signal of led phase problem / cause solution green rapid flashing during recharging recharging in progress wit until recharge has been completed red and green alternate flashing during recharging - one or more batteries fully discharged, slo...

  • Page 9

    Problem solutions 1. Mobile does not work after pressing the on/off switch - check the correct connection of ac/dc adapter to mobile and electrical mains - if you’re using batteries, verify charge conditions and their correct position in the compartment - for further information please see tables i ...