Calibrating the Humidity Display
Your hygrometer is factory calibrated and should not need any
further adjustments. The following instructions should be
performed only to correct an error in calibration.
1.
Test the humidity display settings by placing the Caliber I
V™
next to another unit that is known to be accurate.
2.
Determine the amount of calibration needed by finding the
difference between the display value and the expected value.
For example, if you are expecting 70% but your unit displays 71%,
the difference in calibration is 71-70 = 1. If you need the display
lower as in this example, change the calibration by -1%. If you need
the display higher, add (+) the value.
3. Press the Set button for 3 seconds to place the unit into
calibration mode.
4. In calibration mode, you will see the current adjustment setting
flashing on the display. If this is the first time you are
calibrating your unit, it will show 0%. If you have already
calibrated the unit previously, you will see the current
adjustment.
5. Press the Min/Max/Clear button to adjust the display based
on your calibration. Pressing the button once will change the
adjustment by 1%. The Caliber I
V™
adjustment range is +/-6%
in the sequence: 0, +1, +2, +3, +4, +5, +6, -6, -5, -4, -3, -2, -1, 0.
6. When you have completed your adjustment, press the Set
button one more time and it will take the unit out of
calibration mode. If you leave the unit in calibration mode for more
than one minute without pressing any buttons, the unit will
automatically will change back to operations mode.
Restoring Factory Calibration
To restore your Caliber IV back to factory calibration, simply
follow the calibration instructions above and adjust the display
back to 0%. This will restore the unit back to factory calibration.
Limited Warranty
The Caliber IV™ is covered by a limited warranty against defects
from materials or workmanship for a full 1 Year with proof of
purchase. Any misuse or excessive wear and tear is not covered
under the limited warranty.
Contact Information
If you have any questions or warranty inquiries please call us
directly at 516-520-5258 or email
customersupport@cigaroasis.com.
V2.18