Campbell CS300 Instruction Manual

Manual is about: CS300 Pyranometer

Summary of CS300

  • Page 1

    Cs300 pyranometer revision: 1/11 c o p y r i g h t © 1 9 9 4 - 2 0 1 1 c a m p b e l l s c i e n t i f i c , i n c ..

  • Page 2

    Warranty and assistance this equipment is warranted by campbell scientific (canada) corp. (“csc”) to be free from defects in materials and workmanship under normal use and service for six (6) months from date of shipment unless specified otherwise. ***** batteries are not warranted. ***** csc's obli...

  • Page 3: Please Read First

    Please read first about this manual please note that this manual was originally produced by campbell scientific inc. (csi) primarily for the us market. Some spellings, weights and measures may reflect this origin. Some useful conversion factors: area: 1 in 2 (square inch) = 645 mm 2 length: 1 in. (i...

  • Page 4: Cs300 Table of Contents

    Cs300 table of contents pdf viewers note: these page numbers refer to the printed version of this document. Use the adobe acrobat® bookmarks tab for links to specific sections. 1. General description.....................................................1 2. Specifications ...............................

  • Page 6: Cs300 Pyranometer

    Cs300 pyranometer 1. General description the cs300 measures incoming solar radiation with a silicon photovoltaic detector mounted in a cosine-corrected head. Output from the detector is a current, which is converted to voltage by a potentiometer potted in the sensor head. The resistance of the poten...

  • Page 7: 3.  Installation

    Cs300 pyranometer 3. Installation the cs300 should be mounted such that it is never shaded by the tripod/tower or other sensors. The sensor should be mounted with the cable pointing towards the nearest magnetic pole, e.G., in the northern hemisphere point the cable toward the north pole. Mounting he...

  • Page 8

    Cs300 pyranometer cs300 pyranometer pn 18356 base ossarm cm225 stand cm200 series cr figure 3-1. Cm225 pyranometer mounting stand and cm202 crossarm cs300 pyranometer pn 18356 base 015arm figure 3-2. 015arm pyranometer mounting arm 3.

  • Page 9: 4.  Wiring

    Cs300 pyranometer 4. Wiring a schematic diagram of the cs300 is shown in figure 4-1. Red black clear 500 to 600 : typical single-ended analog input ground/ag ground/g figure 4-1. Cs300 schematic connections to campbell scientific dataloggers are given in table 4-1. When short cut for windows softwar...

  • Page 10

    Cs300 pyranometer solar radiation can be recorded as an average flux density (w m -2 ) or daily total flux density (mj m -2 ). The appropriate multipliers are listed in table 5-1. Negative values should be set to zero before being processed. Table 5-1. Multipliers required for average flux and total...

  • Page 11

    Cs300 pyranometer 'define data tables datatable(table1,true,-1) datainterval(0,60,min,10) average(1,slrw,fp2,false) endtable datatable(table2,true,-1) datainterval(0,1440,min,10) totalize(1,slrmj,ieee4,false) endtable 'main program beginprog scan(10,sec,1,0) 'measure cs300 pyranometer voltse (slrw,1...

  • Page 12

    Cs300 pyranometer ; set negative values to zero 2: if (xf) (p89) 1: 1 x loc [ slrw ] 2: 4 3: 0 f 4: 30 then do 3: z=f x 10^n (p30) 1: 0 f 2: 0 n, exponent of 10 3: 1 z loc [ slrw ] 4: end (p95) ; convert mv to mj/m2 for 10 second execution interval 5: z=x*f (p37) 1: 1 x loc [ slrw ] 2: .00005 f 3: 2...

  • Page 13

    Cs300 pyranometer 13: real time (p77) 1: 1220 year,day,hour/minute (midnight = 2400) 14: resolution (p78) 1: 1 high resolution 15: totalize (p72) 1: 1 reps 2: 2 loc [ slrmj ] 16: resolution (p78) 1: 0 low resolution 5.2 total solar radiation if solar radiation is totalized in units of kj m -2 , ther...

  • Page 14: 7.  Troubleshooting

    Cs300 pyranometer recalibrate the cs300 every three years. Obtain an rma number before returning the cs300 to campbell scientific, inc. For recalibration. 7. Troubleshooting symptom: -9999 or radiation values around 0 1. Check that the sensor is wired to the single-ended channel specified by the mea...

  • Page 15

    Cs300 pyranometer 10

  • Page 17

    Campbell scientific companies campbell scientific, inc. (csi) 815 west 1800 north logan, utah 84321 united states www.Campbellsci.Com • info@campbellsci.Com campbell scientific africa pty. Ltd. (csaf) po box 2450 somerset west 7129 south africa www.Csafrica.Co.Za • cleroux@csafrica.Co.Za campbell sc...