Campbell LLAC4 Instruction Manual

Manual is about: LLAC4 Four-Channel, Low-Level AC Conversion Module

Summary of LLAC4

  • Page 1

    Llac4 four-channel, low-level ac conversion module revision: 3/12 c o p y r i g h t © 2 0 0 4 - 2 0 1 2 c a m p b e l l s c i e n t i f i c , i n c ..

  • Page 3: Warranty

    Warranty “products manufactured by campbell scientific, inc. Are warranted by campbell scientific, inc. (“campbell”) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless otherwise specified in the corresponding campbell...

  • Page 4: Assistance

    Assistance products may not be returned without prior authorization. The following contact information is for us and international customers residing in countries served by campbell scientific, inc. Directly. Affiliate companies handle repairs for customers within their territories. Please visit www...

  • Page 5: Llac4 Table of Contents

    Llac4 table of contents pdf viewers: these page numbers refer to the printed version of this document. Use the pdf reader bookmarks tab for links to specific sections. 1. General description.....................................................1 2. Specifications ........................................

  • Page 6

    This is a blank page..

  • Page 7: Ac Conversion Module

    Llac4 four-channel, low-level ac conversion module 1. General description the llac4 is a peripheral device used to increase the number of low-level ac signals a datalogger can monitor. The module enables four datalogger control ports to emulate pulse-counting channels by converting the low-level ac ...

  • Page 8: 2.  Specifications

    Llac4 four-channel, low-level ac conversion module 2. Specifications power: 8 to 16 vdc 0.1 ma @ 12 vdc signal conditioning: minimum ac input voltage versus output square wave frequency: input sine wave (mv rms) 20 200 2000 5000 output square wave range (hz) 1.0 to 20 0.5 to 200 0.3 to 10,000 0.3 to...

  • Page 9: 4.  Wiring

    Llac4 four-channel, low-level ac conversion module table 2. Control port compatibility datalogger input (hz) control port measurement type cr1000 cr3000 400,000 c1,c2… c8 high frequency cr800 400,000 c1,c2,c3,c4 high frequency cr5000 5,000,000 c7,c8 high frequency cr23x 2,500 c5,c6,c7,c8 high freque...

  • Page 10: 5.  Program Examples

    Llac4 four-channel, low-level ac conversion module 5. Program examples using the cr1000, let's measure four different wind speed sensors that produce low-level ac signals. These will include the rm young 05103, 05106, 05305, propeller type anemometers, and rm young 03101 cup anemometer. Program exam...

  • Page 11

    Llac4 four-channel, low-level ac conversion module 'main program beginprog scan (1,sec,0,0) pulsecount (rmy05103ws,1,15,0,1,0.098,0.0) pulsecount (rmy05106ws,1,16,0,1,0.098,0.0) pulsecount (rmy05305ws,1,17,0,1,0.1024,0.0) pulsecount (rmy03101ws,1,18,0,1,0.750,0.20) if rmy03101ws rmy03101ws = 0.0 end...

  • Page 12

    Llac4 four-channel, low-level ac conversion module 6 'main program beginprog scan (1,sec) pulsecount (rmy05103ws,p_ll,0,1,0.098,0.0) pulsecount (rmy05106ws,c1,0,1,0.098,0.0) pulsecount (rmy05305ws,c2,0,1,0.1024,0.0) pulsecount (rmy03101ws,p_sw,0,1,0.750,0.20) if rmy03101ws rmy03101ws = 0.0 endif cal...

  • Page 14

    Campbell scientific companies campbell scientific, inc. (csi) 815 west 1800 north logan, utah 84321 united states www.Campbellsci.Com • info@campbellsci.Com campbell scientific africa pty. Ltd. (csaf) po box 2450 somerset west 7129 south africa www.Csafrica.Co.Za • cleroux@csafrica.Co.Za campbell sc...