E-Medic Silence TP 1 User Manual

Manual is about: All- in-one Medical Panel PC

Summary of Silence TP 1

  • Page 2: Table Of Contents

    - 1 - table of contents i. Safety & regulatory notice ............................................................ 2-9 a. Electronic magnetic compatibility .................................................. 2-6 b. Safety ..................................................................................

  • Page 3

    - 2 - i. Safety & regulatory notice a. Electronic magnetic compatibility guidance and manufacturer’s declaration – electromagnetic emissions the e-medic silence tp 2 & tp 1 are intended for use in the electromagnetic environment specified below. The user should assure the products are used in such a...

  • Page 4

    - 3 - the e-medic silence tp 2 & tp 1are intended for use in an electromagnetic environment in which radiated rf disturbances are controlled. The user of the e-medic silence tp 2 & tp 1 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile rf com...

  • Page 5

    - 4 - discharge (esd) iec 61000-4-2 ±8 kv air ±8 kv air concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. Electrical fast transient/burst iec 61000-4-4 ±2 kv for power supply lines ±1 kv for input/output lines ±2 kv for power supp...

  • Page 6

    - 5 - power frequency (50/60 hz) magnetic field iec 61000-4-8 3 a/m 3 a/m power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Note ut is the a.C. Mains voltage prior to application of the test level. Guidance and m...

  • Page 7

    - 6 - distance in meters (m). Field strengths from fixed rf transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. B interference may occur in the vicinity of equipment marked with the following symbol: note 1 at 80 mhz and ...

  • Page 8: B. Safety

    - 7 - b. Safety iso 7000-0434 : caution, consult accompanying documents. Iso 7000-1641 : follow operating instructions or consult instructions for use. Iec 60417 -5009 : stand-by. Iec 60417-5031 : direct current. Eu-wide legislation, as implemented in each member state, requires that waste electrica...

  • Page 9

    - 8 - collateral standard: safety requirements for medical electrical systems – en 60601-1-2 (iec 60601-1-2) medical electrical equipment part 1-2: general requirements for safety collateral standard: electromagnetic compatibility; requirements and tests accessory equipment connected to the analog a...

  • Page 10: C. Intended Use

    - 9 - c. Intended use the e-medic silence tp 2 & tp 1 are designed to serve as a medical pc for general purpose usage in the hospital environment and for clinic review purpose. It could be used as a computer device for surgical, radiology, pacs (picture archiving communication systems), lis (lab inf...

  • Page 11: II. Product Basic

    10 ii. Product basic a. Specifications main system cpu: for tp 2 . Support intel core i ivy bridge ultra low power cpu w/ bga1023 package, 17w tdp (max.) and integrated graphics cpu: for tp 1 . Support intel core i ivy bridge mobile cpu w/ pga988 package, 35w tdp (max.) and integrated graphics chips...

  • Page 12

    11 life time 35 million activations storage & expansion hdd 2.5” sata hdd drive bay x 1 + anti-vibration kit expansion slot . Pci-e (x16) x 1 (optional) . Card size: 100 x 155 x 18/0.5 mm or 68 x 220 x 18/0.5 mm . Mini pcie x 2 (52 pin card-edge type) i/o & control keys usb usb 2.0 x 4 com db-9 x 2 ...

  • Page 13

    12 altitudes . Operating: up to 3,000 m (9,842 feet) . Shipping: up to 12,192 m (40,000 feet) pressure . Operating: 700 – 1060 hpa . Storage/transportation: 186 – 1,060 hpa vibration . Operating: 15g/0.53 oz, 11 ms, half sine wave . Non-operating: 50g/1.76 oz, 11 ms, half sine wave shock . Operating...

  • Page 14

    13 b. The packing / outline / i/o / control buttons basic packing list -- . The medical pc: e-medic silence tp 2 (fanless) or tp 1 (smart fan) . Power adaptor (sinpro mpu101-105) . Ac power cord . Touch stylus pen . Dvd title (with drivers & user’s manual) . Screws (for rear cover vesa mounting) not...

  • Page 15

    14 . E-medic silence tp 2 . The i/o interface (basic configuration).

  • Page 16

    15 . The control buttons & led indicators control buttons & led indicators 1. Hdd indicator: dark - non-operating/stand-by amber blinking - accessing 2. Battery indicator: dark - non-operating amber - operating amber blinking - battery low 3. Power indicator: dark - off green - on 4. Power on/off bu...

  • Page 17: C. Others

    16 c. Others . To clean or disinfect the system to clean or disinfect the system, to avoid any damage to the touch panel and the plastic housing, please check below chemical resistance table to ensure you disinfectant does not contain harmful ingredients. We suggest to use ethanol and do check with ...

  • Page 18

    17 chemical plastic housing 40% sodium hydroxid damage acetone damage carbitol acetate damage cellosolve acetate damage ethanol / alcohol 500g load with 95% alcohol--> ok methylene chloride damage hydrochloric acid - 6% damage hexane damage isopropanol (isopropyl alcohol)/propanol damage methyl ethy...

  • Page 19

    18 . The internal battery (optional) for those systems installed with internal battery, below is the specification. Battery cell type lithium-ion rechargeable battery cell (2000mah/cell) configuration 2s2p typical capacity 4000mah (for two cells) minimum capacity 3800mah (for two cells) nominal volt...

  • Page 20: III. Bios

    19 iii. Bios immediately after powering on the system, during the post (power on self test) stage, press the key to enter the bios setup utility program. Follow the menu & the keys instruction to modify the bios configuration & settings. Main --.

  • Page 21

    20 advanced -- launch oprom support launch pxe oprom enables or disables boot option for legacy network devices. Launch storage oprom enables or disables boot option for legacy mass storage devices with option rom . Pci subsystem settings pci rom priority in case of multiple option roms (legacy and ...

  • Page 22

    21 extended tag if enabled allows device to use 8-bit tag field as a requester. No snoop enables or disables pci express device no snoop option. Maximum payload set maximum payload of pci express device or allow system bios to select the value maximum read request set maximum read request size of pc...

  • Page 23

    22 core-multi processing enable or disable core-multi processing mode. Execute disable bit xd can prevent certain classes of malicious buffer overflow attacks when combined with a supporting os (windows server 2003 sp1, windows xp sp2, suse linux 9.2, redhat enterprise 3 update 3.) limit cpuid maxim...

  • Page 24

    23 tv1 standard tv2 standard active lfp select the active lfp configuration. No lvds:vbios does not enable lvds. Int-lvds:vbios enables lvds driver by integrated encoder. Sdv0 lvds:vbios enables lvds driver by sdv0. Usb configuration legacy usb support allows usb devices to be used in ms-dos. Ehci h...

  • Page 25

    24 serial port 3 configuration set parameters of serial port 3 (com e). Serial port 4 configuration set parameters of serial port 4 (com f). Serial port console redirection serial port console redirection..

  • Page 26

    25 chipset -- host bridge/south bridge this screen provides information on host bridge/south bridge parameters..

  • Page 27

    26 boot -- setup prompt timeout number of seconds to wait for setup activation key. 65535(0xffff) means indefinite waiting. Bootup numlock state selects the keyboard numlock state. Quiet boot allows you to determine whether to display the ami logo at system startup. Disabled displays normal post mes...

  • Page 28

    27 always do not allow disabling ga20. Option rom messages sets display made for option rom. Interrupt 19 capture enables or disables option roms to trap int 19. Boot option priorities specifies the sequence of loading the operating system from the installed hard drives..

  • Page 29

    28 security – enables or disables the security chip. It is recommended that you use this function with the administrator/user password..

  • Page 30

    29 save & exit – save changes and exit exit system setup after saving the changes. Discard changes and exit exit system setup without saving any changes. Save changes and reset reset the system after saving the changes. Discard changes and reset reset system setup without saving the changes. Save ch...

  • Page 31

    30 baaske medical gmbh & co. Kg bacmeisterstr. 3 32312 lübbecke tel: +49 5741 236027 – 0 fax: +49 5741 236027 – 99 vertrieb@baaske.Net www.Baaske-medical.De.