EDAN SE-1201 User Manual

Manual is about: Electrocardiograph

Summary of SE-1201

  • Page 2: About This Manual

    I about this manual p/n: 01.54.455318-10 release date: november 2011 © copyright edan instruments, inc. 2011. All rights reserved. Statement this manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with...

  • Page 3: Terms Used In This Manual

    Ii terms used in this manual this guide is designed to give key concepts on safety precautions. Warning a warning label advises against certain actions or situations that could result in personal injury or death. Caution a caution label advises against actions or situations that could damage equipme...

  • Page 4

    Iii table of contents chapter 1 safety guidance ........................................................................................................... 1 1.1 intended use ........................................................................................................................... 1...

  • Page 5

    Iv 4.2 configuring the electrocardiograph ..................................................................................... 33 4.3 screen description................................................................................................................ 34 4.3.1 about the main screen.......

  • Page 6

    V 8.1 loading orders ..................................................................................................................... 71 8.2 examining orders................................................................................................................. 72 8.3 deleting orders.....

  • Page 7

    Vi 13.1.3 cleaning the print head ............................................................................................. 113 13.2 disinfection ...................................................................................................................... 113 13.3 care and maintenance ....

  • Page 8: Chapter 1 Safety Guidance

    Se-1201 electrocardiograph user manual safety guidance - 1 - chapter 1 safety guidance this chapter provides important safety information related to the use of the 12-channel electrocardiograph. 1.1 intended use the intended use of se-1201 12-channel electrocardiograph (hereinafter called se-1201) i...

  • Page 9

    Se-1201 electrocardiograph user manual safety guidance - 2 - warning 2. Only qualified service engineers can install this equipment, and only service engineers authorized by the manufacturer can open the shell. 3. Explosion hazard - do not use the electrocardiograph in the presence of flammable anes...

  • Page 10

    Se-1201 electrocardiograph user manual safety guidance - 3 - warning 17. The use of equipment that applies high frequency voltages to the patient (including electrosurgical equipment and some respiration transducers) is not supported and may produce undesired results. Disconnect the patient data cab...

  • Page 11

    Se-1201 electrocardiograph user manual safety guidance - 4 - warning therefore anybody, who connects additional equipment to the signal input or output connector to configure a medical system, must make sure that it complies with the requirements of the valid version of the system standard iec/en 60...

  • Page 12

    Se-1201 electrocardiograph user manual safety guidance - 5 - warning 4. Do not heat or splash the battery or throw it into fire or water. 5. Do not destroy the battery; do not pierce battery with a sharp object such as a needle; do not hit with a hammer, step on or throw or drop to cause strong shoc...

  • Page 13: 1.3 List of Symbols

    Se-1201 electrocardiograph user manual safety guidance - 6 - caution 5. The device and accessories are to be disposed of according to local regulations after their useful lives. Alternatively, they can be returned to the dealer or the manufacturer for recycling or proper disposal. Batteries are haza...

  • Page 14

    Se-1201 electrocardiograph user manual safety guidance - 7 - mains supply battery indicator battery recharging indicator enter key delete key esc key reset key shift key fn key power on/off key 1mv/copy key mode key print/stop key tab key feed paper key lead switch key/left/right arrow key.

  • Page 15

    Se-1201 electrocardiograph user manual safety guidance - 8 - up/down arrow key review key gender key age group key recycle part number serial number date of manufacture manufacturer authorized representative in the european community the symbol indicates that the device complies with the european co...

  • Page 16: Chapter 2 Introduction

    Se-1201 electrocardiograph user manual introduction - 9 - chapter 2 introduction se-1201 gathers ecg signals of 12 leads simultaneously. It displays the operation menu, ecg parameters as well as electrocardiograms. The 12-channel ecg waves can be viewed on the lcd screen and printed out by using a h...

  • Page 17: 2.2 Keyboard and Keys

    Se-1201 electrocardiograph user manual introduction - 10 - symbol name explanation a mains supply indicator when the device is powered by the mains supply, this indicator is lit. B battery indicator when the device is powered by the battery, this indicator is lit. C battery recharging indicator when...

  • Page 18

    Se-1201 electrocardiograph user manual introduction - 11 - name explanation a function key press to select menu functions on the screen. B gender key press to select the gender for the patient when gender is selected in the patient information setup window. C age group key press to select the age gr...

  • Page 19: 2.3 Rear Panel

    Se-1201 electrocardiograph user manual introduction - 12 - l mode press to select a working mode among the auto, manual, rhythm, r-r analysis and off modes. Note: only if a working mode is selected in the work mode setup window, can the working mode be selected by pressing the mode key when the main...

  • Page 20: 2.4 Right Panel

    Se-1201 electrocardiograph user manual introduction - 13 - name explanation a potential equalization conductor potential equalization conductor provides a connection between the unit and the potential equalization bus bar of the electrical installation. B mains supply socket ac source: alternating c...

  • Page 21

    Se-1201 electrocardiograph user manual introduction - 14 - 1) patient cable socket : applied part of type cf with defibrillator proof : caution definitions of corresponding pins: pin signal pin signal pin signal 1 c2 (input)/ v2 (input) 6 sh 11 f (input)/ ll (input) 2 c3 (input)/ v3 (input) 7 nc 12 ...

  • Page 22

    Se-1201 electrocardiograph user manual introduction - 15 - definitions of corresponding pins: pin signal pin signal pin signal 1 cd dat3 5 clk 9 dat2 2 cmd 6 vss 10 cd 3 vss 7 dat0 11 wp 4 vcc 8 dat1 3) usb socket 1/usb socket 2/usb socket 3 caution only the usb equipment recommended by the manufact...

  • Page 23: 2.5 Bottom Panel

    Se-1201 electrocardiograph user manual introduction - 16 - 4) external input/output socket definitions of corresponding pins: pin signal pin signal 1 gnd 4 gnd 2 gnd 5 ecg signal (input) 3 gnd 6 ecg signal (output) 2.5 bottom panel figure 2-5 se-1201 bottom panel name explanation a speaker hole path...

  • Page 24: 2.6 Function Features

    Se-1201 electrocardiograph user manual introduction - 17 - 1) battery compartment rated voltage: 14.8v rated capacity: 2200mah warning 1. Improper operation may cause the battery to be hot, ignited or exploded, and it may lead to the decrease of the battery capacity. Therefore, it is necessary to re...

  • Page 25

    Se-1201 electrocardiograph user manual introduction - 18 - ♦ high resolution thermal recorder, recording frequency response ≤150hz ♦ supporting external usb printer ♦ supporting accurate digital filter to decrease the polarization voltage and other interferences ♦ supporting folded paper recorded wi...

  • Page 26: And Electrodes

    Se-1201 electrocardiograph user manual operation preparations - 19 - chapter 3 operation preparations warning before use, the equipment, patient cable and electrodes should be checked. Replace them if there is any evident defectiveness or aging which may impair the safety or the performance, and mak...

  • Page 27: 3.2 Preparing The Patient

    Se-1201 electrocardiograph user manual operation preparations - 20 - 3.1.2 connecting the patient cable to electrodes align all lead wires of the patient cable to avoid twisting, and connect the lead wires to the reusable electrodes or the alligator clips. Firmly attach them. The identifiers and col...

  • Page 28

    Se-1201 electrocardiograph user manual operation preparations - 21 - 3.2.2 preparing the skin thorough skin preparation is very important. The skin is a poor conductor of electricity and frequently creates artifacts that distort the ecg signals. By performing methodical skin preparation, you can gre...

  • Page 29

    Se-1201 electrocardiograph user manual operation preparations - 22 - standard 12-lead placement only for the reusable electrodes european label american label electrode placement c1 v1 fourth intercostal space at the right border of the sternum c2 v2 fourth intercostal space at the left border of th...

  • Page 30

    Se-1201 electrocardiograph user manual operation preparations - 23 - nehb placement european label american label electrode placement n st a1 attachment point of the second rib to the right sternal edge n ax a2 fifth intercostal space on the left posterior axillary line n ap v4 left mid-clavicular l...

  • Page 31

    Se-1201 electrocardiograph user manual operation preparations - 24 - limb electrode connection: 1) ensure that the electrodes are clean; 2) clean the electrode area which is a short distance above the ankle or the wrist with 75% alcohol; 3) daub the electrode area on the limb with gel evenly; 4) pla...

  • Page 32

    Se-1201 electrocardiograph user manual operation preparations - 25 - 3.3.3 attaching the disposable electrodes caution the disposable electrodes can only be used for one time. Disposable electrode: alligator clip: disposable electrodes must be used together with alligator clips. Disposable electrode...

  • Page 33

    Se-1201 electrocardiograph user manual operation preparations - 26 - 1) environment: ♦ make sure that there is no electromagnetic interference source around the equipment, especially large medical electrical equipment such as electrosurgical equipment, radiological equipment, magnetic resonance imag...

  • Page 34

    Se-1201 electrocardiograph user manual operation preparations - 27 - the electrocardiograph can be powered by either the mains supply or the battery. To turn on the electrocardiograph: ♦ when operating on ac power make sure that the mains supply meets the requirements (refer to a1.4 power supply spe...

  • Page 35

    Se-1201 electrocardiograph user manual operation preparations - 28 - to turn off the electrocardiograph: ♦ when operating on ac power hold down the key to display the hint system is shutting down… on the screen. Then the device will be off a few seconds later. Remove the plug from the outlet. ♦ when...

  • Page 36

    Se-1201 electrocardiograph user manual operation preparations - 29 - loading/replacing paper: 1) press the casing button downwards to open the recorder. 2) remove the remainder paper from the paper tray if necessary. 3) take off the wrapper of the new folded paper, and then put it in the paper tray....

  • Page 37

    Se-1201 electrocardiograph user manual operation preparations - 30 - 6) advance the recorder paper. If paper marker is set to yes, you can press feed to advance the recorder paper to the next black marker; if paper marker is set to no, you can press feed to advance the paper for 2.5cm. Pressing feed...

  • Page 38: 4.1 Navigation Tips

    Se-1201 electrocardiograph user manual basic operation guidance - 31 - chapter 4 basic operation guidance the following sections provide an overview of the main operations and functions. You can operate the electrocardiograph by using the touch screen (optional). Caution do not touch the lcd screen ...

  • Page 39

    Se-1201 electrocardiograph user manual basic operation guidance - 32 - 4.1.2 entering data take the patient information window for example: 1. Press f1 below patient on the main screen1to open the patient information window. 2. Press tab or shift + tab to move the cursor to the name textbox. To inpu...

  • Page 40

    Se-1201 electrocardiograph user manual basic operation guidance - 33 - 4.1.3 selecting an item take the work mode setup window for example: 1. In the work mode setup window, press tab or shift + tab to move the cursor among different check boxes. Press space to select a check box, and a check mark √...

  • Page 41: 4.3 Screen Description

    Se-1201 electrocardiograph user manual basic operation guidance - 34 - 4.3 screen description 4.3.1 about the main screen after the electrocardiograph is turned on, the main screen appears. Figure 4-1 se-1201 main screen1 figure 4-2 se-1201 main screen2 q u t s r j d e f h i g a b p o n m l k c.

  • Page 42

    Se-1201 electrocardiograph user manual basic operation guidance - 35 - name explanation a name patient name: within 60 ascii characters or the equivalent number of other characters that can be supported by equivalent memory used by 60 ascii characters b id when id mode is set to manual, the patient ...

  • Page 43

    Se-1201 electrocardiograph user manual basic operation guidance - 36 - m gain gain: 2.5 mm/mv, 5 mm/mv, 10 mm/mv, 20 mm/mv, 10/5 mm/mv or agc note: this setup modified on the main screen is only effective for the current patient. N speed in the manual mode, you can set speed to 5mm/s, 6.25mm/s, 10mm...

  • Page 44

    Se-1201 electrocardiograph user manual basic operation guidance - 37 - 4.3.2 about the system setup screen select setup on the main screen2 to display the system setup screen. Note: if you set the system password in the system maintenance window, you need enter the password before opening the system...

  • Page 45

    Se-1201 electrocardiograph user manual basic operation guidance - 38 - 4.3.3 about the order manager screen select order on the main screen2 to open the order manager screen. You can press load to load orders and select a loaded order to start an order examination. Or, you can press esc to return to...

  • Page 46

    Se-1201 electrocardiograph user manual basic operation guidance - 39 - 200 is the total number of orders that can be stored in the electrocardiograph. 2 is the current number of orders stored in the electrocardiograph. H order list orders will be loaded and displayed in the order list. The order inf...

  • Page 47

    Se-1201 electrocardiograph user manual basic operation guidance - 40 - figure 4-4 file manager screen1 figure 4-5 file manager screen2 a b c d e f h g i j k l m n h g.

  • Page 48

    Se-1201 electrocardiograph user manual basic operation guidance - 41 - name explanation a trans all press to transmit all the files to the pc. B export all press to export all the files from the electrocardiograph to the u disk or sd card. C del all press to delete all the files from the electrocard...

  • Page 49: 4.4 Work Mode Description

    Se-1201 electrocardiograph user manual basic operation guidance - 42 - 4.4 work mode description there are five work modes in se-1201. Auto: in the auto mode, the ecg data can analyzed, saved, printed and transmitted. The lead groups are switched automatically according to the lead sequence during t...

  • Page 50

    Se-1201 electrocardiograph user manual entering patient information - 43 - chapter 5 entering patient information 5.1 entering patient information manually operation procedures are as follows: 1. Configure the patient information setup window. (optional) 1) select the desired items. Select the desir...

  • Page 51: (Optional)

    Se-1201 electrocardiograph user manual entering patient information - 44 - first name within 30 ascii characters last name within 30 ascii characters age age unit: years, months, weeks or days gender patient gender (male/female/empty) pacemaker if you select pacemaker in the patient information setu...

  • Page 52

    Se-1201 electrocardiograph user manual entering patient information - 45 - 3. When the main screen is displayed, scan the patient’s bar code with the bar code reader, and then the patient information will appear in the corresponding box. Note: only bar code readers, complying with the standards bar ...

  • Page 53

    Se-1201 electrocardiograph user manual entering patient information - 46 - 6. Enter the patient id manually in the id textbox and press order, then the matched order will be loaded from smart ecg viewer software and the order information will be displayed in the corresponding textboxes. Or, you can ...

  • Page 54: 6.1 Printing An Ecg Report

    Se-1201 electrocardiograph user manual printing ecg reports - 47 - chapter 6 printing ecg reports note: 1. The working mode can not be changed during the printing course. Stop printing reports before changing the working mode. 2. Within three seconds after returning to the main screen, if you press ...

  • Page 55

    Se-1201 electrocardiograph user manual printing ecg reports - 48 - 2. When the main screen is displayed, press the mode key to select the auto mode. Press f3 to select a paper speed. Press f4 to switch the gain. Press f5 to set the emg filter or the lowpass filter. 3. Press the print/stop key to pri...

  • Page 56

    Se-1201 electrocardiograph user manual printing ecg reports - 49 - 2. When the main screen is displayed, press the mode key to select the manualmode. Press f3 to select a paper speed. Press f4 to switch the gain. Press f5 to set the emg filter or the lowpass filter. 3. Press the left or right arrow ...

  • Page 57

    Se-1201 electrocardiograph user manual printing ecg reports - 50 - 2. When the main screen is displayed, press the mode key to select the rhythmmode. Press f3 to select a paper speed. Press f4 to switch the gain. Press f5 to set the emg filter or the lowpass filter. 3. Press the print/stop key to be...

  • Page 58: 6.2 Copy Printing

    Se-1201 electrocardiograph user manual printing ecg reports - 51 - 2. When the main screen is displayed, press the mode key to select the r-r analysismode. Press f4 to switch the gain. Press f5 to set the emg filter or the lowpass filter. 3. Press the print/stop key to begin sampling, the sampling t...

  • Page 59: 6.3 Freezing Ecg Waves

    Se-1201 electrocardiograph user manual printing ecg reports - 52 - 6.3 freezing ecg waves you can freeze the ecg waves displayed on the main screen. Operation method: 1) press f3 to set the paper speed, press f4 to set the gain, and press f5 to set the filter on the main screen1. 2) select freeze to...

  • Page 60

    Se-1201 electrocardiograph user manual printing ecg reports - 53 - 6.4 printing a stored ecg report 6.4.1 printing on the file manager screen1/2 on the file manager screen 1: 1. Select file on the main screen to open the file manager screen1. 2. Select a file on the file manager screen1, and then pr...

  • Page 61: 6.5 Sample Ecg Reports

    Se-1201 electrocardiograph user manual printing ecg reports - 54 - 6.5 sample ecg reports 6.5.1 ecg reports in the auto mode quick mode (a).

  • Page 62

    Se-1201 electrocardiograph user manual printing ecg reports - 55 - (b) the above figure (a) and (b) show an ecg report in the auto mode. Template is selected, and record style is set to 6×2+1r. The ecg report includes: patient information, measure information, diagnosis information, report confirmed...

  • Page 63

    Se-1201 electrocardiograph user manual printing ecg reports - 56 - 2*5s+1r (12 leads are printed in 2 groups of 6 with the ecg wave of one lead on the bottom, and every group is printed for about 5s), v1.0 (software version), semip v1.6 (algorithm version), se-1201 (electrocardiograph model), instit...

  • Page 64

    Se-1201 electrocardiograph user manual printing ecg reports - 57 - save paper mode the above figure shows an ecg report in the auto mode. The style is 6×2+1r..

  • Page 65

    Se-1201 electrocardiograph user manual printing ecg reports - 58 - 6.5.2 ecg reports in the rhythm mode the above figure shows an ecg report in the rhythm mode, and rhythm style is set to three leads..

  • Page 66

    Se-1201 electrocardiograph user manual printing ecg reports - 59 - 6.5.3 ecg reports in the manual mode the above figure shows an ecg report in the manual mode, and manual style is set to 12 channels..

  • Page 67

    Se-1201 electrocardiograph user manual printing ecg reports - 60 - 6.5.4 ecg reports in the r-r analysis mode (a).

  • Page 68

    Se-1201 electrocardiograph user manual printing ecg reports - 61 - (b).

  • Page 69

    Se-1201 electrocardiograph user manual printing ecg reports - 62 - the above figure (a) and (b) show an ecg report in the r-r analysis mode. Figure (a) shows: current date & current time patient information (name, id, gender, age, height, weight) measuring time total beat number gain, speed, filter ...

  • Page 70

    Se-1201 electrocardiograph user manual printing ecg reports - 63 - 6.5.5 ecg reports printed by the usb printer.

  • Page 71

    Se-1201 electrocardiograph user manual printing ecg reports - 64 - 6.5.6 ecg report of patient with pacemaker the above figure shows the ecg report of a patient with a pacemaker..

  • Page 72

    Se-1201 electrocardiograph user manual printing ecg reports - 65 - 6.5.7 ecg report of nehb lead the above figure shows the ecg report of nehb lead..

  • Page 73: Ethernet

    Se-1201 electrocardiograph user manual transmitting ecg data - 66 - chapter 7 transmitting ecg data 7.1 transmitting ecg data to the pc ecg data in dat\pdf\scp\fda-xml format can be transmitted to the pc. To transmit ecg data in dat format, the smart ecg viewer software of the manufacturer must be i...

  • Page 74

    Se-1201 electrocardiograph user manual transmitting ecg data - 67 - c) the last section of the local ip item can be set at random. For the same network transmission, a) set the first three sections of the local ip item to the first three sections of the ip of the pc. B) the last section of the local...

  • Page 75: Ethernet

    Se-1201 electrocardiograph user manual transmitting ecg data - 68 - 7.1.2 transmitting ecg data in scp/fda-xml/pdf format through ethernet cable note: scp/fda-xml function can be activated on the advanced setup screen. For details, please contact the manufacturer or the local distributor. 1. Log int...

  • Page 76

    Se-1201 electrocardiograph user manual transmitting ecg data - 69 - 4. In the auto or rhythm mode, ecg data will be transmitted through the net cable automatically after an ecg report is printed out. In the off mode, 10s ecg data sampled after pressing the print/stop key will be transmitted through ...

  • Page 77

    Se-1201 electrocardiograph user manual transmitting ecg data - 70 - 1) double-click on the shortcut icon on the desktop to start up the pc ecg software. 2) click on the system setting button to open the system setting screen, and then click on the device tab. 3) select se serial machine from the sam...

  • Page 78: Chapter 8 Managing Orders

    Se-1201 electrocardiograph user manual managing orders - 71 - chapter 8 managing orders note: to use the order function, the smart ecg viewer software of the manufacturer must be installed in the pc. Select order on the main screen2 to open the order manager screen. Figure 8-1 order manager screen 8...

  • Page 79: 8.2 Examining Orders

    Se-1201 electrocardiograph user manual managing orders - 72 - note: if orders are modified on the smart ecg viewer software, the corresponding orders existing on the order manager screen will be refreshed after you load orders from the software. 5. If you select exam.Room filter on the order setup s...

  • Page 80: 8.4 Searching Orders

    Se-1201 electrocardiograph user manual managing orders - 73 - 8.4 searching orders select search on the order manager screen to display the following window. Select the search type, such as id, name, request no., exam. Room, time, enter the search information, and then press enter to confirm. All th...

  • Page 81

    Se-1201 electrocardiograph user manual managing orders - 74 - 8.5.2 order setup select setup on the order manager screen to make settings of orders. Item description condition choose from: default, id, order date, request no. Or state select default, orders will be displayed in sequence of the time ...

  • Page 82

    Se-1201 electrocardiograph user manual managing orders - 75 - exam.Room filter choose from: select or deselect select this item, a check mark √ appears in the box before exam.Room filter. You can enter an exact exam.Room in the textbox, such as electrocardiograph. If the room you entered exists in t...

  • Page 83: Chapter 9 Managing Files

    Se-1201 electrocardiograph user manual managing files - 76 - chapter 9 managing files if you want to save the ecg data in the electrocardiograph, you should set auto save to to ecg in the file setup window. Then the ecg data in the auto, off or rhythm mode will be saved on the file manager screen au...

  • Page 84: 9.1 Transmitting Files

    Se-1201 electrocardiograph user manual managing files - 77 - on the file manager screen, files can be printed, transmitted, exported, displayed, edited searched or deleted. Se-1201 can accommodate 200 files. If there is no file on the file manager screen, the following dialog box will pop up when yo...

  • Page 85: 9.3 Deleting Files

    Se-1201 electrocardiograph user manual managing files - 78 - caution it is forbidden to connect or disconnect a u disk, an sd card or a usb printer during the transmission course. Note: 1. Please insert the u disk or sd card recommended by the manufacturer. Please set the format to fat or fat32 when...

  • Page 86: 9.5 Importing Files

    Se-1201 electrocardiograph user manual managing files - 79 - 9.5 importing files operation procedures are as follows: 1. Connect the u disk or sd card recommended by the manufacturer to the electrocardiograph. 2. Save files to the directory of ecgdat\import of the u disk or sd card. 3. Select file o...

  • Page 87: 9.8 Previewing A File

    Se-1201 electrocardiograph user manual managing files - 80 - 9.8 previewing a file 1. Select preview on the file manager screen2 to open the file preview screen. 2. Select waveform to display the waveform screen. You can press the left or right arrow to switch between the lead groups when previewing...

  • Page 88

    Se-1201 electrocardiograph user manual managing files - 81 - 4. Select measure to display the measure screen which shows the measurement information. 5. Press print/stop or select record onthe preview screen to print the selected file.Or, press print/stop again to stop printing the file. 6. Select d...

  • Page 89: Chapter 10 System Setup

    Se-1201 electrocardiograph user manual system setup - 82 - chapter 10 system setup select setup on the main screen 2 to display the system setup screen. Figure 10-1 system setup screen 10.1 factory defaults table 10-1 factory defaults work mode setup items default mode options auto, manual, rhythm m...

  • Page 90

    Se-1201 electrocardiograph user manual system setup - 83 - filter setup items default ac filter on emg filter off dft filter 0.67hz lowpass filter 100hz record info setup-setup1 items default record style 6×2 record mode save paper record sequence sequential gain 10mm/mv record device thermal speed ...

  • Page 91

    Se-1201 electrocardiograph user manual system setup - 84 - id mode auto id hint on age mode age h/w unit cm/kg bp unit mmhg prompt confirmed by patient information refreshed on order acquired off user-defined cleared transmission setup items default auto transmission off ftp user name/ftp password/ ...

  • Page 92

    Se-1201 electrocardiograph user manual system setup - 85 - key volume medium notify volume medium date & time setup items default date mode dd-mm-yyyy time mode 24 hours power off/ lcd off cleared file setup items default auto save to ecg file format dat delete after trans. Or export off replace whe...

  • Page 93: 10.2 Work Mode Setup

    Se-1201 electrocardiograph user manual system setup - 86 - 10.2 work mode setup select work mode on the system setup screen, and then press enter to open the work mode setup window. Figure 10-2 work mode setup window item description mode options choose from: auto, manu, rhyt, r-r, or off note: only...

  • Page 94

    Se-1201 electrocardiograph user manual system setup - 87 - item description auto dispstyle choose from: 3×4, 3×4+1r, 3×4+3r, 6×2, 6×2+1r or 12×1 select 3×4 to display ecg waves of 12 leads in 4 groups of 3. Select 3×4+1r to display ecg waves of 12 leads in 4 groups of 3 with the ecg wave of one rhyt...

  • Page 95: 10.3 Filter Setup

    Se-1201 electrocardiograph user manual system setup - 88 - 10.3 filter setup select filter on the system setup screen, and then press enter to open the filter setup window. Figure 10-3 filter setup window item description ac filter choose from: on or off note: ac frequency can be set to 50hz or 60hz...

  • Page 96: 10.4 Record Info Setup

    Se-1201 electrocardiograph user manual system setup - 89 - 10.4 record info setup select record info on the system setup screen, and then press enter to open the record info setup window. 10.4.1 setup 1 press f1 to switch to the setup 1 window. Figure 10-4 record info setup1 item description record ...

  • Page 97

    Se-1201 electrocardiograph user manual system setup - 90 - item description record mode choose from: save paper or quickly select save paper,10s afterpressing the print/stop keyon themain screen,an ecg report is printed. The patient information, measure information, interpretation and ecg waves are ...

  • Page 98

    Se-1201 electrocardiograph user manual system setup - 91 - item description record device options corresponding usb printers hp deskjet 2010 hp deskjet 1050 hp deskjet 2000 hp 2010/1050/2000 hp deskjet 2050 hp laserjet p2015 hp2035 hp laserjet p2035 note: 1. During the usb printing course,pressing t...

  • Page 99

    Se-1201 electrocardiograph user manual system setup - 92 - item description paper marker select no, the device can not identify the start point of each page of the recorder paper while printing ecg reports. Sample time if record style is set to 12×1, you can set the time period. If the time period i...

  • Page 100

    Se-1201 electrocardiograph user manual system setup - 93 - 210×295mm 215×280mm.

  • Page 101

    Se-1201 electrocardiograph user manual system setup - 94 - 10.4.2 setup 2 press f2 to switch to the setup 2 window. Figure 10-5 record info setup2 item description auto record info select position marker, analysis, template, measure or minnesota code, the item will be printed in the ecg reports. Cho...

  • Page 102

    Se-1201 electrocardiograph user manual system setup - 95 - 10.5 patient information setup select patient info on the system setup screen, and then press enter to open the patient information setup window. Figure 10-6 patient information setup window item description patient options select first/last...

  • Page 103

    Se-1201 electrocardiograph user manual system setup - 96 - item description id hint choose from: on or off in the auto, rhythm or off mode, when id is set to manual and id hint is set to on, if you do not input the patient id before pressing the print/stop key, a hint will pop up to remind you to in...

  • Page 104: 10.6 Transmission Setup

    Se-1201 electrocardiograph user manual system setup - 97 - 10.6 transmission setup note: 1. To transmit ecg data in dat format to the pc, the smart ecg viewer software of the manufacturer must be installed in the pc. You should log into the smart ecg viewer software before transmission. 2. To transm...

  • Page 105: 10.7 Lead Setup

    Se-1201 electrocardiograph user manual system setup - 98 - item description ftp information enter data in the ftp path, ftp user name textboxes. Ip addresses set server ip, local ip, set gateway, set subnet mask for details, please refer to section 7.1.1 “transmitting ecg data in dat format through ...

  • Page 106: 10.8 Display&sound Setup

    Se-1201 electrocardiograph user manual system setup - 99 - item description rhythm lead1/2/3 in the r-r analysis mode: the r-r analysis report of the rhythm lead selected in the rhythm lead1 list box will be printed. Lead sequence choose from: standard or cabrera lead sequence lead group 1 lead grou...

  • Page 107

    Se-1201 electrocardiograph user manual system setup - 100 - item description brightness set the brightness within 0~20. Key volume choose from: low, medium, high or off select low, medium or high, the electrocardiograph gives a short sound when you press keys on the keyboard. Select off, there is no...

  • Page 108: 10.9 Date&time Setup

    Se-1201 electrocardiograph user manual system setup - 101 - 10.9 date&time setup select date&time on the system setup screen, and press enter to open the date&time setup window. Figure 10-10 date &time setup window note: please set date&time correctly when it’s the first time you use the electrocard...

  • Page 109: 10.10 File Setup

    Se-1201 electrocardiograph user manual system setup - 102 - 10.10 file setup select file on the system setup screen, and press enter to open the file setup window. Figure 10-11 file setup window item description auto save choose from: off, to ecg or to external memory select off, ecg data will not b...

  • Page 110

    Se-1201 electrocardiograph user manual system setup - 103 - item description delete after trans. Or export choose from: on or off select on, the files will be automatically deleted from the file manager screen after they are transmitted to the pc or exported to the u disk or sd card. Replace when me...

  • Page 111

    Se-1201 electrocardiograph user manual system setup - 104 - 10.11 system maintenance setup select maintenance on the system setup screen, and press enter to open the system maintenance window. Figure 10-12 system maintenance window item description export settings press to export the system settings...

  • Page 112: 10.12 Other Setup

    Se-1201 electrocardiograph user manual system setup - 105 - 10.12 other setup select other on the system setup screen, press enter to open the other setup window. Figure 10-13 other setup window item description institution input the institution name manually within 40 ascii characters. Note: the to...

  • Page 113

    Se-1201 electrocardiograph user manual system setup - 106 - item description external output the external output socket is equipped in the electrocardiograph, through which the electrocardiograph can send signals to the external equipment. Choose from: off, standard or triggered if external input is...

  • Page 114: Chapter 11 Hint Information

    Se-1201 electrocardiograph user manual hint information - 107 - chapter 11 hint information hint information and the corresponding causes provided by the electrocardiograph are listed in table 11-1. Table 11-1 hint information and causes hint information causes lead off electrodes fall off the patie...

  • Page 115: Chapter 12 Troubleshooting

    Se-1201 electrocardiograph user manual troubleshooting - 108 - chapter 12 troubleshooting 1. Operating problems q1: i was trying to select a file from the file list on the file manage screen, but the file was in the middle of the long list. Is there any way to make the selection faster? A1: actually...

  • Page 116

    Se-1201 electrocardiograph user manual troubleshooting - 109 - q6: the hint “memory full! Replace the earliest file?” pops up every time when i operate the electrocardiograph. What am i supposed to do? A6: the hint is used to remind you that the amount of stored file reaches 200. You can just make a...

  • Page 117

    Se-1201 electrocardiograph user manual troubleshooting - 110 - reload the paper in the paper tray, close the recorder casing firmly. If the problem still exists, please contact the manufacturer or the local distributor for further disposal. Q4: i want to print the hospital name in the report, but i ...

  • Page 118

    Se-1201 electrocardiograph user manual troubleshooting - 111 - 3. Transmitting problems q1: the ecg doesn’t respond to any keys after a long time of transmission. It transmits nothing for there is no new data appearing on the screen of the pc software. What should i do? A1: some error may occur duri...

  • Page 119: 13.1 Cleaning

    Se-1201 electrocardiograph user manual cleaning, care and maintenance - 112 - chapter 13 cleaning, care and maintenance caution 1. Turn off the power before cleaning and disinfection. The mains supply must be switched off if it is used. 2. Prevent the detergent from seeping into the equipment. 13.1 ...

  • Page 120: 13.2 Disinfection

    Se-1201 electrocardiograph user manual cleaning, care and maintenance - 113 - caution do not clean the unit and accessories with abrasive fabric and avoid scratching the electrodes. 13.1.3 cleaning the print head 1. Open the recorder casing and remove the paper. 2. Wipe the print head gently with a ...

  • Page 121

    Se-1201 electrocardiograph user manual cleaning, care and maintenance - 114 - 13.3.1 recharge and replacement of battery 1) capacity identification the battery capacity can be identified according to the battery symbol in the top right corner of the lcd screen. : full capacity; : 3/4 capacity; : 1/4...

  • Page 122

    Se-1201 electrocardiograph user manual cleaning, care and maintenance - 115 - warning 3. When the battery’s useful life is over, contact the manufacturer or the local distributor for disposal or dispose of the battery according to local regulations. 4. Remove the battery from the electrocardiograph ...

  • Page 123

    Se-1201 electrocardiograph user manual cleaning, care and maintenance - 116 - 13.3.4 maintenance of the main unit and the patient cable caution besides the maintenance requirements recommended in this manual, comply with local regulations on maintenance and measurement. The following safety checks s...

  • Page 124

    Se-1201 electrocardiograph user manual cleaning, care and maintenance - 117 - 1) main unit ♦ avoid excessive temperature, sunshine, humidity and dirt. ♦ put the dustproof coat on the main unit after use and prevent shaking it violently when moving it to another place. ♦ prevent any liquid from seepi...

  • Page 125: Chapter 14 Accessories

    Se-1201 electrocardiograph user manual accessories - 118 - chapter 14 accessories warning only the patient cable and other accessories supplied by the manufacturer can be used. Or else, the performance and electric shock protection can not be guaranteed. 14.1 standard accessories table 14-1 standard...

  • Page 126

    Se-1201 electrocardiograph user manual accessories - 119 - alligator clip/banana socket adapters 01.57.040173 adult disposable adhesive electrodes 11.57.471056 pediatric disposable adhesive electrodes 11.57.471057 disposable resting tab electrodes (1 piece) 11.57.471031-10 electrode gel 11.25.78047 ...

  • Page 127: 15.1 Warranty

    Se-1201 electrocardiograph user manual warranty & service - 120 - chapter 15 warranty & service 15.1 warranty edan warrants that edan’s products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within warranty period. The warranty ...

  • Page 128: A1.1 Safety Specifications

    Se-1201 electrocardiograph user manual technical specifications - 121 - appendix 1 technical specifications a1.1 safety specifications comply with: iec/en 60601-1+a1+a2, iec/en60601-1-2+a1, iec/en60601-2-25, ansi/aami ec11, iec/en 60601-2-51 anti-electric-shock type: class І with internal power supp...

  • Page 129: A1.3 Physical Specifications

    Se-1201 electrocardiograph user manual technical specifications - 122 - a1.3 physical specifications dimensions 361mm×262mm×135mm (91.6in×66.5in×34.3in) weight approx. 4.2 kg (1.9 lbs) (excluding recorder paper and battery) display 800×480 tft lcd screen a1.4 power supply specifications operating vo...

  • Page 130

    Se-1201 electrocardiograph user manual technical specifications - 123 - a1.5 performance specifications recording recorder: thermal dot-matrix recorder printing density 8 dots per mm / 200 dots per inch (amplitude axes) 40 dots per mm / 1000 dots per inch (time axes, @ 25 mm/s) recorder paper: folde...

  • Page 131

    Se-1201 electrocardiograph user manual technical specifications - 124 - ac filter: on/off dft filter: 0.05hz/0.15hz/0.25hz/0.32hz/0.5hz/0.67hz emg filter: 25hz/35hz/45hz/off filter lowpass filter:150hz/100hz/75hz cmrr ≥115db sampling frequency 1000 hz pacemaker detection amplitude ±2 to ±700 mv widt...

  • Page 132: Appendix 2 Emc Information

    Se-1201 electrocardiograph user manual emc information - 125 - appendix 2 emc information guidance and manufacture’s declaration - electromagnetic emissions- for all equipment and systems guidance and manufacture’s declaration – electromagnetic emission the 12-channel electrocardiograph is intended ...

  • Page 133

    Se-1201 electrocardiograph user manual emc information - 126 - guidance and manufacture’s declaration - electromagnetic immunity - for all equipment and systems guidance and manufacture’s declaration – electromagnetic immunity the 12-channel electrocardiographis intended for use in the electromagnet...

  • Page 134

    Se-1201 electrocardiograph user manual emc information - 127 - for 25 cycles t (>95% dip in u t ) for 5 sec for 25 cycles t (>95% dip in u t ) for 5 sec powered from an uninterruptible power supply or a battery. Note u t is the a.C. Mains voltage prior to application of the test level. Guidance and ...

  • Page 135

    Se-1201 electrocardiograph user manual emc information - 128 - metres (m). Field strengths from fixed rf transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range. B interference may occur in the vicinity of equipment marked wit...

  • Page 136

    Se-1201 electrocardiograph user manual emc information - 129 - recommended separation distances between portable and mobile rf communications equipment and the equipment or system - for equipment or system that are not life-supporting recommended separation distances between portable and mobile rf c...

  • Page 137: Appendix 3 Abbreviation

    Se-1201 electrocardiograph user manual abbreviation - 130 - appendix 3 abbreviation abbreviation statement lcd liquid crystal display bp blood pressure ecg electrocardiogram/electrocardiograph hr heart rate avf left foot augmented lead avl left arm augmented lead avr right arm augmented lead la left...

  • Page 138

    - 131 -.