Empi Active Manual - Technical Specifications
EMPI Active User Guide
18
EMPI Active User Guide
Technical Specifications
continued
Specifications – Guidance and Manufacturer’s Declaration
The Empi Active is intended for use in the electromagnetic environment specified
below. The customer or user of the Empi Active should assure that it is used in such
an environment.
Electromagnetic Emissions
Emission Tests
Compliance
Guidance Regarding Electromagnetic Environment
RF emissions CISPR 11
Group 1
The Empi Active uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment
RF emissions CISPR 11
Class B
The Empi Active is suitable for use in all establishments, including
domestic establishments and those directly connected to the public
low-voltage power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
N/A - Battery
Operated Device
Voltage fluctuations/
Flicker Emissions
IEC 61000-3-3
N/A - Battery
Operated Device
Electromagnetic Immunity
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic
Environment - Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast transient/
burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
N/A - Battery
Operated Device
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
N/A - Battery
Operated Device
Mains power quality should be that
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
T
(>95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in
U
T
) for 5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
T
(>95% dip in
U
T
) for 5 sec
N/A - Battery
Operated Device
Mains power quality should be that
of a typical commercial or hospital
environment.
Power frequency
(50/60Hz) magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical com-
mercial or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Technical Specifications
continued
Specifications – Guidance and Manufacturer’s Declaration
Electromagnetic Immunity continued
Immunity Test
IEC
60601
Test
Level
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the Empi Active device,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of
the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 V
d = [3.5]
√P
V
1
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 V/m
d = [3.5]
√P
80 MHz to 800 MHz
E
1
d = [7]
√P
800 MHz to 2.5 GHz
E
1
Where P is the maximum output power rating of the transmit-
ter in watts (W) according to the transmitter manufacturer and
d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electromagnetic site survey
a
, should be less than the
compliance level in each frequency range
b
.
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength
in the location in which the Empi Active device is used exceeds the applicable RF compliance
level above, the Empi Active device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating
the Empi Active device.