Fast Weigh Scales MS-600 User Manual - page 5
be weighed. The display will show the net weight of
the contents.
4. To return to normal weight display (gross weight),
remove any items from the weighing platform and
tap the
key again to re-zero the scale.
VII. Calibration
Calibration and adjustment are performed at the factory.
However, the weighing range can shift slightly during
shipping. Also, due to changes in local acceleration of gravity
around the world, adjustment is recommended when moving
the scale from one location to another. To check calibration:
1. Place the required calibration weight on the scale
and note the displayed value.
2. If the displayed value is not within acceptable
tolerance (see chart below), adjustment is required.
Model
Acceptable
Tolerance (±)
Calibration
Weight
Recommended
Accuracy Class
TR-100
0.03g
100g
OIML M2
TR-600
0.2g
500g
OIML M2
VIII. Adjustment
1. Turn the scale ON.
2. Press and hold the
M key until the display shows
“CAL”.
3. Press
Magain.Thedisplaywillflash“CAL”followed
by the required calibration weight (see above).
4. Place the required calibration weight on the platform.
5. The display will show “PASS” then return to normal
weighing. Calibration is complete!