Fema S40-T User manual

Manual is about: Series S40 Panel meter 72x36mm size for Pt100 / RTD for Thermocouples

Summary of S40-T

  • Page 1

    Series s40 panel meter 72x36mm size panel meter for pt100/rtd (2 and 3 wires), and thermocouples j, k, t, e, s, r, n, c, l and x with temperature display in degrees celsius (ºc) or fahrenheit (ºf). Selectable pt100/rtd resolution at 1º or 0.1º, and manual offset available. Thermocouple measure with ...

  • Page 2: Meter S40-T

    User’s manual s40-t fema electrÓnica - page 2 meter s40-t panel meter 72x36mm size for pt100/rtd and thermocouples panel meter for temperature signals, accepts pt100/rtd with 2 and 3 wires, and thermocouples j, k, t, e, s, r, n, c, l and x. Temperature display in degrees celsius (ºc) or fahrenheit (...

  • Page 3

    User’s manual s40-t user’s manual s40-t fema electrÓnica - page 3 button up logo alarms units power signal option1 option2 button sq button le front view rear view earth connection - although a terminal is offered for earth connection, the connection is optional. The instrument does not need this co...

  • Page 4

    User’s manual s40-t fema electrÓnica - page 4 type range ºc max. Error at 25ºc* range ºf offset drift span drift* *includes offset drift pt100/rtd 800 / -200 ºc 1562 / -328 ºf 0.05º/ºc 0.10º/ºc thermocouple j 1200 / -200 ºc 2192 / -328 ºf 0.05º/ºc 0.20º/ºc thermocouple k 1372 / -200 ºc 2372 / -328 º...

  • Page 5

    User’s manual s40-t user’s manual s40-t fema electrÓnica - page 5 information menu to enter the “information menu” press the sq button. The “information menu” allows to visualize information and does not allow to modify the configuration of the instrument. It is not affected by the “password” functi...

  • Page 6

    User’s manual s40-t fema electrÓnica - page 6 configuration menu operating the configuration menu to enter the “configuration menu” press the sq button. The “configuration menu” allows to change the configuration of the instrument. Access to the “configuration menu” can be password protected with th...

  • Page 7

    User’s manual s40-t user’s manual s40-t fema electrÓnica - page 7 configuration menu sensor pt100 2wire degrees ºc resolution 0.1º offset 0 on break to high alpha 385 alarms 1,2,3 and 4 active off (not managed) type maximum setpoint 1000 hysteresis 0 counts delay 0.0 seconds setpoint2 off display st...

  • Page 8

    User’s manual s40-t fema electrÓnica - page 8 input menu the input menu selects the input signal range. Options avail- able are pt100/rtd 2 or 3 wires, and thermocouples j, k, t, e, s, r, n, c, l and x. Note - thermocouple x is a linear signal at 10uv/ºc. The instrument manages up to 4 alarms. Each ...

  • Page 9

    User’s manual s40-t user’s manual s40-t fema electrÓnica - page 9 password (pass) - select a number to act as password. This password will be requested when entering the configuration menu. To deactivate the password select “off”. Factory settings (fact) - factory default configuration. Se- lect “ye...

  • Page 10

    User’s manual s40-t fema electrÓnica - page 10 relay type 3 contacts (com, nc, no) maximum current 8a (resistive load) voltage 250 vac continuously instalable in option1 and/or option2 a b c d e common nc no k l m n f g h i j common nc no o p q r connections for option1 connections for option2 manuf...

  • Page 11

    User’s manual s40-t user’s manual s40-t fema electrÓnica - page 11 t h i s p a g e b l a n k.

  • Page 12: Other Products

    Large displays 60&100mm digit signal converters & isolators panel meters standard 96x48mm other products electronic instrumentation for industry www.Fema.Es fema electrÓnica, s.A. Altimira 14 - pol. Ind. Santiga e08210 barberà del vallès barcelona - spain tel. (+34) 93.729.6004 - www.Fema.Es fax (+3...