Francis Searchlights FR 230 User Information - Fault Finding
FR 230 Searchlight
6
Using the multi-meter take a
voltage reading;
Using the Ammeter, take an amps
reading from the live cable to the
lamp;
Multiply these values together to
give an approximate wattage.
For Example:
Using a 240v 1000w Tungsten
Halogen lamp,
Voltage reading : 240v
Amps reading: 4 amps
Therefore:
Wattage = 240 x4 = 960 watts.
Fault Finding
If a fault occurs the following
procedures should be adopted. Any
work to be carried out should be
conducted by a competent person. A
multimeter should be used for fault
finding purposes.
Fault
Lamp Fails to light
Causes
1) Power not supplied
2) Fuse blown
3) Failed lamp
Remedy
1) Check voltage at supply/battery. If
necessary recharge battery. If
supply is not present fault is at
customer supply. If power is
present, see remedy 2.
2) Check fuse for visual failure. If
none noticeable check fuse for
continuity. If fuse found to be
faulty, replace with new part. If
found to be working correctly see
remedy 3.
3) Firstly, check supply at lampholder
connecting block (within
searchlight barrel). If supply
present, disconnect unit from
power supply before removing
lamp. If noticeable damage to
filament is present, the lamp will
have failed. The lamp can also be
checked using a multimeter for
continuity of the filament. Replace
new lamp following all precautions
mentioned previously and lamp
manufacturers instructions.