Freedom KD2-00-KNEE2-KT Instructions For Use Manual

Manual is about: Plié 2.0 MPC Knee

Summary of KD2-00-KNEE2-KT

  • Page 1

    Plié® 2.0 mpc knee instructions for use (r-720-109 rev. E), english! ! ! ! ! 1 instructions for use.

  • Page 2: Table of Contents

    Table of contents product number: ks2-00-knee1-kt/kd2-00-knee2-kt appropriate use! 5 symbols legend! 6 safety information! 6 knee functions! 8 technical data! 8 plié® 2.0 mpc knee specifications! 8 battery and battery charger specifications! 8 environmental specifications! 8 storage! 8 water resista...

  • Page 3: Plié® 2.0 Mpc Knee

    Plié® 2.0 mpc knee instructions for use (r-720-109 rev. E), english! ! ! ! ! 3 h d c a b e f g plié® 2.0 mpc knee (images not to scale).

  • Page 4

    Plié® 2.0 mpc knee instructions for use (r-720-109 rev. E), english! ! ! ! ! 4 n i j k l m o recommended foot components p q r.

  • Page 5

    Pictured part description part number included w/knee quantity available separately a b c d e f g h i j k l m n o p q r plié mpc knee ks2-00-knee1-00 na plié® 2.0 mpc knee, threaded top kd2-00-knee2-00 na wireless usb adapter ks1-00-blue2-00 ✓ 1 ✓ battery case ✓ 1 software cd-rom ks2-00-sftw1-00 ✓ 1...

  • Page 6

    Symbols legend the following symbols are found on the plié® 2.0 mpc knee: • attention, refer to plié® 2.0 mpc knee instructions for use. • type bf applied part. • conformité européene (european conformity). • federal communications commission (usa conformity) . • separate collection for electrical a...

  • Page 7

    • if the user changes shoe heel height, the plié control threshold setting may require adjustment as the knee requires a minimum user toe loading (adjustable toe threshold) to release into swing phase. • during repetitive activities without swing phase, such as the use of exercise equipment, the use...

  • Page 8

    Knee functions the plié® 2.0 mpc knee is a single axis prosthetic knee joint system providing microprocessor control of both the swing and stance phases of gait. The microprocessor monitors an embedded load sensor and an angle sensor to precisely control the transitions between the stance and swing ...

  • Page 9

    Finishing options the plié® 2.0 mpc knee may be finished with a cosmetic/protective cover. However, the choice of cover material may affect normal knee function. A cover should be modified or not used if it detrimentally affects the knee performance. A discontinuous, two-piece cover is recommended a...

  • Page 10

    Plié 2.0 control the plié 2.0 control software is necessary for programming plié® 2.0 mpc knees (figure a). The plié 2.0 control consists of a main window and four auxiliary windows. The auxiliary windows may be accessed from the main window. The plié 2.0 control software is compatible with agility™...

  • Page 11

    7. Calibrate - select calibrate to establish the reference zero values for the embedded sensors. 8. Wireless sync - select wireless sync if the software and knee fail to sync (circles remain red). 9. Setting wizard - select the setting wizard to open the plié 2.0 setup wizard window (figure b). The ...

  • Page 12

    Plié 2.0 real time data display the plié 2.0 real time data display allows the viewing of live data as the user ambulates. The plié 2.0 real time data display provides a data rate of 4 hz (4 samples per second). Due to the limited data stream bandwidth (knee microprocessor data rate is 1000 hz), the...

  • Page 13

    Manually adjusting the additional parameters is not recommended. If the setup mode box is checked, the advanced stumble recovery is not active. The knee should only be in setup mode during set-up. Plié 2.0 setup wizard the plié 2.0 setup wizard optimizes all the computer settings for plié® 2.0 mpc k...

  • Page 14

    ! The white tabs on the bottom and left frames of the data window allow the data view to be modified. To ! Scroll the data, simultaneously place the cursor between the white tabs, hold the left mouse button, and ! Drag the mouse. To collapse or expand the data, simultaneously place the cursor on a w...

  • Page 15

    ! Spinal column (figure h). !During ambulation, an unaccommodated hip flexion contracture may cause an excessively asymmetrical !Gait pattern. Bench alignment the plié® 2.0 mpc knee was designed to be in a fully extended position during stance phase for level ground ambulation. Consequently, an inhe...

  • Page 16

    B. With equal weight on each limb, adjust the height of the prosthesis as necessary. C. Ensure the trochanter-knee-ankle (tka) alignment follows the recommendations of inherent stability. D. Ensure the transverse rotation of the knee and foot is appropriate. E. Instruct the user to sit in a chair. A...

  • Page 17

    D. Instruct the user to ambulate. If the knee does not release into swing, lower the toe threshold setting in the plié control window until the knee releases. E. Select next to continue. 3. Adjust manual hydraulic settings adjust the default manual hydraulic settings in the following order to ensure...

  • Page 18

    J. Select the save data icon. The slow walk recorded data file will be transferred wirelessly to the computer. K. Select the next icon to proceed to the set knee window. L. In the set knee window, select the apply to knee icon. The plié 2.0 setup wizard will transfer the optimal stumble recovery (le...

  • Page 19

    Regulatory compliance, disclosures, and classification wireless operation plié® 2.0 mpc knee bluetooth certification qpln ref. No.: q31606_wt11_sgs the plié® 2.0 mpc knee complies with the following: • fcc id ampck001 or fcc id qoqwt11. This device complies with part 15 of the fcc rules. Operation i...