Good Will Instrument GPT-9600 series User Manual - page 7
Insulation Resistance
[for GPT-9612/9603]
Output Voltage
50V/100V/250V/500V/1000V
Output-Voltage Accuracy
± (3.0% of setting + 1 count) with no load
Voltage Regulation
± (1.5% + 2 counts) [full load ->no load]
Voltmeter Accuracy
± (1.5% of reading + 2 counts)
Resistance Measurement Range
1M
Ω~ 2000MΩ
Test Voltage
Measure Range
Accuracy
50V / 100V / 250V
1~50MΩ
51~2000MΩ
± (5% of reading + 2MΩ)
± (10% of reading + 2MΩ)
500V / 1000V
1~500MΩ
501~2000MΩ
± (5% of reading + 2MΩ)
± (10% of reading + 2MΩ)
Output Impedance
600kΩ
Window Comparator Method
Yes
RAMP (Ramp-Up Time)
0.1s Fixed
TIMER (Test Time)
1s~180s
GND
OFF
General
Note
The specifications apply when the GPT-9600 series are powered on for at least 30
minutes at 15°C~35°C.
Display
240 x 48 Ice Blue dot matrix LCD
AUTO Mode*
AC-IR / IR-AC / DC-IR / IR-DC
Interface
Remote, Signal I/O
Power Source
AC 100-120 V / 220-240 V ±10%
Power Line Frequency 50/60Hz
Power Consumption
Max 400VA
Dimensions
322(W) x 148(H) x 385(D) mm
Weight
Approx. 9kg max.
* The available Auto mode depends on the selected model
Environment
Note
To operate over the specific temperature range may cause damage to the circuit.
And, do not use in a place where strong magnetic or electric field exists.
Warranty Environment 15°C to 35°C, Humidity 70% (no condensation)
≦
Operation Environment 0°C to 40°C, Humidity 70% (no condensation)
≦
Storage Environment
-10°C to 70°C, Humidity 85% (no condensation)
≦
Installation Location
Indoors at an amplitude of up to 2000m