8.4 Octopus 900 control unit / PC
A standard PC can be used as control unit for the Perimeter. The control unit soft-
ware runs on WINDOWS VISTA SP2, WINDOWS 7, WINDOWS 8 and WINDOWS
10
NOTE!
The minimum requirements for PCs can be found in the EyeSuite in-
structions for use.
9. Maintenance
WARNING!
•The housing components of the Perimeter appliance may only be re-
moved by suitably qualified service personnel.
•The ON/OFF switch does not isolate the Perimeter from the mains.
Before removing the housing components, ensure that the appliance
is unplugged from the mains power socket.
•This device must not be modified without the manufacturer's approval.
Installation and repairs may only be performed by trained specialists.
•Contact your HAAG-STREIT representative for installation, repairs
and modification work on the system. The contact details are avail-
able at www.haag-streit.com.
•Warranty claims can be made only if the instructions in these instruc-
tions for use have been complied with.
9.1 Repairs
We recommend having an authorised professional check the Octopus 900 every
two years to ensure long-term and trouble-free functioning. Further information and
the corresponding technical documentation for this are available from HAAG-STRE-
IT or your local representative.
9.2 Cleaning
Regular dusting of the device with a soft cloth is sufficient. More stubborn dirt can
be removed using a soft, lint-free cloth dampened with water or alcohol at maximum
70%.
NOTE!
Do not allow the appliance to become wet and do not use other sol-
vents of any kind.
A dust cover is included in the accessories of the Octopus 900. Cover the appliance
when the room is being cleaned or if it is not used for longer periods. Always re-
move the dust cover before switching on the power.
WARNING!
The appliance must not be switched on when covered (heat build-up,
fire hazard).
9.2.1 Cupola
The inner surface of the cupola is coated with a special paint finish designed to en-
sure optimum results in perimetric examinations. It is not necessary to clean this in-
ner surface in normal cases. Should dust be visible in the cupola, you can remove
it by gently wiping with a soft, dry and fluff-free cloth. A soft cloth dampened slightly
with mild soapsuds may only be used for local cleaning in emergencies, such as if
spots have appeared due to patients' sneezing.
9.2.2 Response button, chin and forehead rest, eye occluder
These components are all made of easy-to-clean plastic materials. Disinfect them
after every patient to keep them hygienically clean.
NOTE!
These application parts should be disinfected prior to every examina-
tion (e.g., with 70% isopropyl alcohol) in order to comply with general
hygiene requirements and prevent the transmission of infections.
9.2.3 Display, control panel
Fingerprints and dust can be removed using a soft, moist cloth.
9.2.4 Lightsources
In contrast to other perimetric devices, LEDs are used in the Octopus 900 as light
sources for background and stimulus. These have a service life of >30,000h. If any
of the LEDs still has to be replaced, please contact your dealer's customer service
department.