Hach H125 Operating Instructions

Summary of H125

  • Page 1

    Continue as follows for two-point calibration tip: when using two-point calibration, always bracket the expected ph value of the sample. For example, if the expected ph of a sample is ph 5.0, use ph 7.00 and ph 4.01 buffer. 8. Apply one drop of either ph 4.01 or ph 10.01 buffer to the probe tip. 9. ...

  • Page 2

    Specifications minilab h125 catalog number h125 model minilab h125 meter pocket-sized waterproof ph meter sensor silicon chip ph sensor calibration 1,2, or 3 point buffer recognition automatic ph 4.0, ph 7.0, ph 10.0 temp. Compensation automatic reference replaceable kcl gel filled. Part no. Rf01-01...