Hach TU5300 sc Basic User Manual

Summary of TU5300 sc

  • Page 1

    Doc023.5 2.90477 tu5300 sc/tu5400 sc 03/2017, edition 4 basic user manual.

  • Page 3: Table of Contents

    Table of contents specifications on page 3 calibration on page 21 general information on page 4 verification on page 21 installation on page 10 maintenance on page 21 user navigation on page 20 troubleshooting on page 27 operation on page 20 additional information an expanded user manual is availabl...

  • Page 4: General Information

    Specification details measurement units tu5300 sc: ntu, fnu, te/f, ebc or ftu; tu5400 sc: ntu, mntu 2 , fnu, mfnu, te/f, ebc, ftu or mftu. Range 0 to 1000 ntu, fnu, te/f and ftu; 0 to 250 ebc method detection limit tu5300 sc: 0.002 fnu at 25 °c (77 °f) ( 0.0001 fnu at 25 °c (77 °f) ( response time t...

  • Page 5

    Safety information n o t i c e the manufacturer is not responsible for any damages due to misapplication or misuse of this product including, without limitation, direct, incidental and consequential damages, and disclaims such damages to the full extent permitted under applicable law. The user is so...

  • Page 6

    This symbol identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment. This symbol indicates radio waves. Class 1 laser product d a n g ...

  • Page 7

    W a r n i n g electromagnetic radiation hazard. Do not use the instrument in dangerous environments. N o t i c e this instrument is sensitive to electromagnetic and electromechanical interference. These interferences can have an effect on the analysis performance of this instrument. Do not put this ...

  • Page 8

    Residential area is likely to cause harmful interference, in which case the user will be required to correct the interference at their expense. The following techniques can be used to reduce interference problems: 1. Disconnect the equipment from its power source to verify that it is or is not the s...

  • Page 9

    Figure 1 product overview 1 programmable button 8 process head 2 status indicator light 5 9 overflow drain 3 rfid module indicator (optional) 10 process head (open) 4 cleaning lid screws (3x) 11 process head (closed) 5 cleaning lid 12 channels for cables 6 sample inlet 13 extension connector for acc...

  • Page 10: Installation

    Figure 2 product components 1 tu5300 sc or tu5400 sc 6 cleaning lid screws and washers for hot water applications 2 wall mount bracket (two tubing clips on bracket) 7 vial replacement tool 3 tubing clips 8 flow regulator 4 tubing clip screws, 2.2 x 6 mm 9 service bracket 5 mounting screws, 4 x 16 mm...

  • Page 11

    Figure 3 installation overview with no accessories 1 service bracket 5 sample inlet 2 process head 6 sample outlet 3 sc controller 7 tu5300 sc or tu5400 sc 4 flow regulator wall mount install the instrument on a wall in a vertical position. Install the instrument so that it is level. Install with th...

  • Page 12

    12 english.

  • Page 13

    Install directly on a wall as an alternative, refer to the illustrated steps that follow to install the instrument directly on a wall. The mounting hardware is supplied by the user. Remove the thin, plastic film from the mounting holes on the back of the instrument. Install the desiccant cartridge n...

  • Page 14

    Figure 4 examine the desiccant cartridge 1 install by date (mm.Yyyy = month and year) 2 indicator (blue = not expired, pink = expired) 14 english.

  • Page 15

    English 15.

  • Page 16

    Replace the cleaning lid screws n o t i c e do not overtighten the screws or breakage will occur. Hand tighten the screws. If the sample temperature is 40 to 60 °c (104 to 140 °f), the cleaning lid screws will become hot. To prevent burns, replace the standard cleaning lid screws with the cleaning l...

  • Page 17

    1. Install the latest software version on the sc controller before the instrument is connected to the sc controller. Get the latest software version from www.Hach.Com . Refer to the software installation instructions supplied in the box or supplied in the software download for the sc controller. 2. ...

  • Page 18

    • tubing 6 • tubing cutter 1. Plumb the instrument. Refer to the illustrated steps that follow and figure 6 . Note: to plumb the instrument with accessories, refer to the documentation supplied with the accessories. 2. If a service bracket is not installed near the instrument, do the steps that foll...

  • Page 19

    English 19.

  • Page 20: User Navigation

    Set the flow rate 1. Measure the flow with the flow regulator fully open. Make sure that the flow is in the middle of the flow specification. Refer to specifications on page 3. 2. Slowly close the flow regulator until the flow decreases by 20 to 30%. Note: the flow regulator causes back pressure in ...

  • Page 21: Calibration

    Calibration w a r n i n g chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets (msds/sds) for safety protocols. The instrument is factory calibrated and the...

  • Page 22

    N o t i c e stop the sample flow to the instrument and let the instrument become cool before maintenance is done. To set the output behavior during maintenance, push menu and select sensor setup>tu5x00 sc>diag/test>maintenance>output mode. Maintenance schedule table 2 shows the recommended schedule ...

  • Page 23

    1. Push menu. 2. Select sensor setup>tu5x00 sc>diag/test>maintenance>vial cleaning. 3. Complete the steps that show on the controller display. The date the vial was cleaned is automatically saved after the last screen shows. 4. If the optional automatic cleaning module is installed, push menu and se...

  • Page 24

    Clean the vial with the optional wiper n o t i c e carefully remove most of the water in the vial. Carefully put the vial wiper into the process vial so that no water spills out. Clean the process vial with the optional vial wiper as shown in the illustrated steps that follow. Clean the vial compart...

  • Page 25

    N o t i c e do not to touch or scratch the glass of the process vial. Contamination or scratches on the glass can cause measurement errors. Note: make sure that no particles fall into the vial compartment. 1. Push menu. 2. Select sensor setup>diag/test>maintenance>vial replacement. 3. Complete the s...

  • Page 26

    Replace the desiccant cartridge the controller display will show when a desiccant cartridge replacement is due. Refer to the documentation included in the desiccant cartridge bag to replace the desiccant cartridge. 26 english.

  • Page 27: Troubleshooting

    Replace the tubing replace the tubing when the tubing has a blockage or has damage. Turn the flow shutoff valve to stop flow to the instrument. Then refer to plumb the instrument on page 17 to replace the tubing. Troubleshooting refer to the expanded user manual on www.Hach.Com for troubleshooting i...

  • Page 28

    28 english.

  • Page 30

    Hach company world headquarters p.O. Box 389, loveland, co 80539-0389 u.S.A. Tel. (970) 669-3050 (800) 227-4224 (u.S.A. Only) fax (970) 669-2932 orders@hach.Com www.Hach.Com hach lange gmbh willstätterstraße 11 d-40549 düsseldorf, germany tel. +49 (0) 2 11 52 88-320 fax +49 (0) 2 11 52 88-210 info-d...